hart rate monitoring neonatal transition
Conditions
Interventions
None listed
Sponsors
Leids Universitair Medisch Centrum
Eligibility
Age
2 Years to 11 Years
Inclusion criteria
Inclusion criteria: Infants born at a gestational age at least 32 weeks who are evaluated on the resuscitation table during neonatal transition at birth. The estimated birth weight must be at least 1500 grams
Exclusion criteria
Exclusion criteria: Infants that have congenital abnormalities of the thorax, an estimated foetal weight of less than 1500 grams, or are withheld life support at birth
Design outcomes
Primary
| Measure | Time frame |
|---|---|
| Correlation of all pulse rate measurements of the Neobeat, ECG monitor and pulse oximetry device: All continuous data will be compared during the first 10 minutes after birth. The relationship between the three devices will be evaluated using Bland-Altman bias analysis; the difference between the measurements will be plotted against their average. The sensitivity, specificity, positive predictive values, and negative predictive values of measurements collected using the Neobeat, ECG and pulse oximetry for detecting HRECG | — |
Secondary
| Measure | Time frame |
|---|---|
| 1) Correlation between signals of the three monitoring devices over time. All continuous data will be compared during the first 10 minutes after birth. Data will be plotted over time and averaged for 30 second intervals and plotted for each different monitoring device. Data will be compared using an Anova for multiple continuous normally distributed values 2) Duration from application until a reliable signal is acquired. The time until the first reliable signal is acquired is averaged for all three devices and compared using an ANOVA for normally distributed values. The Friedmans* test will be used for non-normally distributed data. | — |
Countries
Netherlands
Outcome results
None listed