cardiac arrhythmia heart rhythm disorders
Conditions
Interventions
None listed
Sponsors
MicroPort CRM B.V.
Eligibility
Age
18 Years to 99 Years
Inclusion criteria
Inclusion criteria: Indication for ICD or CRT/D implant (defibrillator) according to ESC guidelines.
Exclusion criteria
Exclusion criteria: Tricuspid valvular disease or mechanical heart valve. Active myocarditis. Pregnant, or not taking/using contraceptives when of a fertile age.
Design outcomes
Primary
| Measure | Time frame |
|---|---|
| Lead safety after 3 months: absence of any lead-related complications. Electrical performance after 3 months: stable stimulation thresholds in the right ventricle (expected values: 1 Volt +/- 0,5 Volt and a pulse duration of 0,5 ms. | — |
Secondary
| Measure | Time frame |
|---|---|
| Document electrical performance of the right ventricular lead: sensing thershold in mV, impedance in Ohm, stimulation threshold in Volt x ms. Document good detection of VT and VF, as well as adequate delivery of therapy (anti-tachy pacing or defibrillation shock). Safety of the right ventricular lead. Handling assessment of the right ventricular lead. Successful implant percentage of the right ventricular lead. | — |
Countries
Netherlands
Outcome results