A Prospective, Nonrandomized, Noninterventional Study to Compare Nexfin CO-trek Cardiac Output with Thermodilution Cardiac Output

A Prospective, Nonrandomized, Noninterventional Study to Compare Nexfin CO-trek Cardiac Output with Thermodilution Cardiac Output - Nexfin/TD Clinical Study

Status
Active, not recruiting
Phases
Unknown
Study type
Observational
Source
NL-OMON
Registry ID
NL-OMON38805
Enrollment
40
Registered
2013-08-30
Start date
2013-12-11
Completion date
Unknown
Last updated
2024-04-29

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

specifieke aandoeningen zijn niet van belang voor het onderzoek

Interventions

None listed

Sponsors

Edwards Lifesciences BMEYE
Lead Sponsor

Eligibility

Age
18 Years to 99 Years

Inclusion criteria

Inclusion criteria: Subjects will be included if they meet the following criteria: 1. Subjects must be at least 18 years of age 2. Subjects must give signed written informed consent 3. Subjects* height and weight must be accurately obtained prior to study start.

Exclusion criteria

Exclusion criteria: Subjects will be excluded if any of these items exist: 1. Aortic or tricuspid valve regurgitation 2. Aortic stenosis or aneurysms 3. History of uncontrolled cardiac arrhythmia 4. Any peripheral vascular disease or conditions such as Raynaud*s disease or Buerger*s disease 5. Insufficient perfusion of the digits 6. Inability to place the finger cuff appropriately due to subject anatomy or condition 7. Known pregnancy 8. Patients being treated with an intra-aortic balloon pump 9. Patient is currently participating in an investigational drug or another device study that clinically interferes with the study endpoints

Design outcomes

Primary

MeasureTime frame
Noninvasive monitoring of CO with the ccNexfin System is comparable to TD as determined by a bias less than 0.6 L/min.

Secondary

MeasureTime frame
Comparability of both methods as determined by Bland-Altman analysis (Bias and Percentage Error). Precision of Nexfin CO-trek versus TD will be determined. The Pearson correlation coefficient for the CO pairs of both methods will be assessed. In patients where Trendelenburg / reverse Trendelenburg positions are clinically required, changes in CO due to these interventions on hemodynamics will be measured to demonstrate concordance of devices in case of repeated measurement. Fluid administration, use of vasoactive and inotropic drugs will be recorded when available to the Investigator.

Countries

Netherlands

Outcome results

None listed

Source: NL-OMON (via WHO ICTRP)