Video Laryngeal Mask SaCoVLM Versus LMA Supreme and Ambu AuraGain in Children

Oropharyngeal Leak Pressure of the SaCoVLM Video Laryngeal Mask Compared With LMA Supreme and Ambu AuraGain in Paediatric Patients Undergoing General Anaesthesia: A Multicentre Randomised Controlled Trial

Status
Not yet recruiting
Phases
Unknown
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT07891832
Acronym
SACOVLM-PEDI
Enrollment
135
Registered
2026-10-23
Start date
2026-11-01
Completion date
2027-07-31
Last updated
2026-10-23

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Airway Management, Anesthesia, General, Pediatric Anesthesia

Keywords

supraglottic airway device, video laryngeal mask, SaCoVLM, LMA Supreme, Ambu AuraGain, oropharyngeal leak pressure, children

Brief summary

This randomised controlled trial compares three supraglottic airway devices (laryngeal masks) in children with a body weight of 3-30 kg undergoing elective surgery under general anaesthesia: the SaCoVLM video laryngeal mask (UE Medical), which has an integrated camera allowing visual confirmation and adjustment of mask position, the LMA Supreme (Teleflex) and the Ambu AuraGain (Ambu). Children are randomly assigned in a 1:1:1 ratio to one of the three devices. The main question is whether the SaCoVLM provides a higher oropharyngeal leak pressure (a measure of how well the mask seals the airway) than the LMA Supreme and the Ambu AuraGain. Secondary questions include first-attempt insertion success, the need for repositioning or reinsertion of the mask during surgery and airway-related complications.

Detailed description

Blind insertion of supraglottic airway devices is associated with a first-attempt success rate of about 85% and frequent suboptimal positioning, which may require repositioning or reinsertion and can lead to desaturation, laryngospasm or bronchospasm. The SaCoVLM combines a second-generation laryngeal mask with a video camera that visualises the laryngeal inlet and allows the position to be corrected without reinsertion. Design: prospective, parallel-group, randomised controlled superiority trial with three arms (SaCoVLM, LMA Supreme, Ambu AuraGain), conducted in accordance with the Declaration of Helsinki and Good Clinical Practice. Procedures: After confirmation of eligibility, written informed consent of the parent or legal guardian and pre-anaesthetic assessment, participants are randomised 1:1:1 using a computer-generated sequence with variable block sizes (Study Randomizer). Anaesthesia is induced with propofol and sufentanil at the discretion of the attending anaesthesiologist; neuromuscular blocking agents are not administered before device insertion and OLP measurement. Device size follows manufacturer weight-based recommendations; the dorsal surface is lubricated with water-based gel. In the SaCoVLM group, the integrated video display is used to visualise the laryngeal inlet and to verify and, if necessary, adjust the device position. A maximum of three insertion attempts is permitted; failure leads to conversion to tracheal intubation or a different supraglottic airway device. Cuff pressure is set to 40-60 cmH2O using a manual manometer. OLP is measured within 5 minutes after insertion once stable ventilation is achieved (manual mode, ventilation paused, fresh gas flow 10 L/min, APL valve set to 50 cmH2O; the airway pressure at which the manometer reading becomes stable is recorded). In the SaCoVLM group, video image quality (5-point Likert scale) and percentage of glottic opening (POGO) are recorded immediately after OLP measurement. Device function, need for repositioning or reinsertion and complications are recorded during surgery; blood staining is assessed after removal at the end of emergence. Blinding: the anaesthesiologists inserting the device are not blinded; statistical analysis is performed by investigators blinded to group allocation. Statistics: The primary hypothesis is tested by two pre-specified pairwise comparisons (SaCoVLM vs LMA Supreme and SaCoVLM vs Ambu AuraGain) using independent-samples t-tests (or Mann-Whitney U test), with Bonferroni-adjusted two-sided alpha of 0.025 per comparison. LMA Supreme vs Ambu AuraGain is reported descriptively. Secondary endpoints are exploratory (chi-square/Fisher exact test, t-test/Mann-Whitney U test). The primary analysis follows the intention-to-treat principle; a per-protocol analysis is a sensitivity analysis. Sample size: assuming mean OLP of 25 cmH2O (SD 5) for LMA Supreme and 29 cmH2O (SD 5) for SaCoVLM (minimal clinically important difference 4 cmH2O, Cohen's d 0.8), alpha 0.025 and 90% power, 41 evaluable patients per group are required; allowing 10% non-evaluable data, 45 patients per group (135 in total) will be enrolled.

Interventions

Airway management with the SaCoVLM video laryngeal mask (UE Medical, China), a second-generation supraglottic airway device with an integrated camera. The video display is used to visualise the laryngeal inlet and to verify and, if necessary, adjust the device position. Size per manufacturer weight-based recommendation; cuff pressure 40-60 cmH2O.

Airway management with the LMA Supreme (Teleflex, USA), a second-generation supraglottic airway device inserted using the standard blind technique. Size per manufacturer weight-based recommendation; cuff pressure 40-60 cmH2O.

Airway management with the Ambu AuraGain (Ambu A/S, Denmark), a second-generation supraglottic airway device inserted using the standard blind technique. Size per manufacturer weight-based recommendation; cuff pressure 40-60 cmH2O.

Sponsors

Masarykova Nemocnice v Usti nad Labem, Krajska Zdravotni a.s.
Lead SponsorOTHER
Faculty of Health Studies, J. E. Purkyně University in Ústí nad Labem
CollaboratorUNKNOWN
Charles University, Czech Republic
CollaboratorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
SUPPORTIVE_CARE
Masking
NONE

Masking description

The device cannot be masked for the anaesthesiologist inserting it. Statistical analysis is performed by investigators blinded to group allocation.

Eligibility

Sex/Gender
ALL
Age
1 Months to 10 Years
Healthy volunteers
No

Inclusion criteria

* Paediatric patients scheduled for elective surgery under general anaesthesia with a supraglottic airway device * Body weight 3-30 kg (supraglottic airway device sizes 1-2.5) * Written informed consent of the parent or legal guardian

Exclusion criteria

* Body weight \< 3 kg or \> 30 kg * Anticipated difficult airway * History of difficult airway * Indication for airway management by a method other than a supraglottic airway device * Emergency surgery * Absence of written informed consent of the parent or legal guardian

Design outcomes

Primary

MeasureTime frameDescription
Oropharyngeal leak pressure (OLP)From device insertion up to 5 minutes after insertionAirway pressure (cmH2O) at which the circuit manometer reading becomes stable, measured with the breathing circuit in manual mode, ventilation paused, fresh gas flow 10 L/min and the APL valve set to 50 cmH2O, after stable ventilation is achieved.

Secondary

MeasureTime frameDescription
First-attempt insertion success rateFrom the start of the first insertion attempt until successful insertion (maximum 3 attempts), up to 10 minutesProportion of patients in whom the device is successfully inserted with adequate ventilation at the first attempt.
Need for device repositioning or reinsertion during surgeryFrom device insertion until device removal at the end of anaesthesia, up to 5 hoursProportion of patients requiring repositioning, reinsertion of the device or other airway intervention after successful insertion.
Incidence of perioperative airway-related complicationsFrom device insertion until device removal at the end of anaesthesia, up to 5 hoursProportion of patients with at least one of: desaturation (SpO2 \< 92%), laryngospasm, bronchospasm, aspiration; each complication also reported separately.
Blood staining of the device on removalAt device removal at the end of emergence from anaesthesia, up to 5 hours after insertionProportion of patients with visible blood on the device or in secretions after device removal.
Video image quality (SaCoVLM group only)Immediately after OLP measurement, up to 5 minutes after device insertionQuality of the video image of the integrated display rated by the inserting anaesthesiologist on a 5-point Likert scale (1 = worst, 5 = best; image quality only).
Percentage of glottic opening (POGO) (SaCoVLM group only)Immediately after OLP measurement, up to 5 minutes after device insertionPercentage of the glottic opening visible on the integrated video display (0-100%).

Countries

Czechia

Contacts

CONTACTMichal Kalina, MD, PhD, MBA
michal.kalina@kzcr.eu+420477111111
PRINCIPAL_INVESTIGATORMichal Kalina, MD, PhD, MBA

Masaryk Hospital Usti nad Labem, Krajska zdravotni a.s.