The Purpose of This Clinical Trial is to Evaluate the Accuracy of Blood Pressure Measurement of HiCardi M350 Compared to Blood Pressure Measured by Auscultatory Sphygmomanometer in Adult Volunteers (Including Those Meeting Blood Pressure Distribution Requirements) in Accordance With ISO 81060-2:2018

A Prospective, Single-center, Single-arm, Pivotal Clinical Trial to Evaluate the Accuracy of Blood Pressure Measurements From the HiCardi M350(Patient Monitor) in Comparison With Those Measured Using an Auscultatory Sphygmomanometer

Status
Recruiting
Phases
Unknown
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT07622498
Enrollment
100
Registered
2026-06-03
Start date
2026-06-01
Completion date
2026-08-31
Last updated
2026-06-03

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

No Restriction on Disease

Keywords

HiCardi M350, Blood Pressure

Brief summary

This prospective, single-center, single-arm pivotal clinical trial is designed to evaluate the accuracy of blood pressure measurements obtained using the HiCardi M350 patient monitor compared with measurements obtained using an auscultatory sphygmomanometer. The study will enroll adult volunteers aged 19 years or older, including participants needed to satisfy the blood pressure distribution requirements of ISO 81060-2:2018. The planned enrollment is 100 participants, considering the ISO 81060-2:2018 minimum requirement of 85 participants and an estimated dropout rate of approximately 15%. The study is also designed to obtain at least 255 valid paired blood pressure values. After written informed consent, participants will undergo screening assessments, including demographics, vital signs, medical and surgical history, prior and concomitant medications, pregnancy testing for women of childbearing potential, bilateral arm circumference measurement, eligibility assessment, and adverse event assessment. On the blood pressure test day, eligibility will be reconfirmed, an enrollment number will be assigned, HiCardi M350 calibration will be performed, and blood pressure measurements will be obtained using both HiCardi M350 and the reference auscultatory method. The primary effectiveness endpoint is the difference between systolic and diastolic blood pressure values measured by HiCardi M350 and the reference sphygmomanometer under resting conditions. Accuracy will be evaluated according to ISO 81060-2:2018 using Criteria 1 and Criteria 2. Safety will be evaluated based on adverse events, adverse device effects, serious adverse events, serious adverse device effects, unanticipated adverse device effects, and device deficiencies.

Interventions

DEVICEHiCardi M350

HiCardi M350 is a wearable patient monitoring device used to measure systolic and diastolic blood pressure. In this study, the device will be attached to each participant and calibrated according to the predefined procedure. Blood pressure values measured by HiCardi M350 will be compared with reference values obtained using an auscultatory sphygmomanometer and mechanical stethoscope by two independent observers.

Sponsors

Yonsei University
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
19 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

1. Adults aged 19 years or older. 2. Individuals who visit Severance Hospital and voluntarily provide written informed consent after receiving and understanding sufficient explanation of the study. 3. Individuals with hypertension-range blood pressure may be included to meet the blood pressure distribution requirements. Participants receiving antihypertensive medication may be included only if the type and dose of medication have remained stable for at least 4 weeks before screening. Individuals with dose changes, medication additions, or medication discontinuation within 4 weeks before screening will be excluded.

Exclusion criteria

1. Individuals in whom Korotkoff phase V cannot be clearly identified during auscultatory blood pressure measurement. 2. Individuals for whom blood pressure measurement is difficult or expected to be difficult. 3. Individuals with bilateral arm circumference greater than 42 cm. 4. Pregnant women. 5. Individuals currently participating in another clinical trial or who have participated in another clinical trial within 30 days before the screening date. 6. Individuals judged by the investigator to be inappropriate for participation in the clinical trial.

Design outcomes

Primary

MeasureTime frameDescription
Difference between systolic and diastolic blood pressure measurements obtained using HiCardi M350 and a reference auscultatory sphygmomanometerAfter completion of all participants' data collection (within approximately 7 months of enrollment period)The primary outcome is the difference between systolic and diastolic blood pressure values measured by HiCardi M350 and the reference auscultatory sphygmomanometer under resting conditions. Accuracy will be assessed separately for systolic and diastolic blood pressure according to ISO 81060-2:2018 Criteria 1 and Criteria 2.