Diabetes Mellitus, Diabetes Mellitus, Type 1, Diabetes Mellitus, Type 2
Conditions
Brief summary
A prospective, multicenter evaluation of performance and safety of the Eversense Gemini System with flash glucose measurement feature. The purpose of this clinical investigation is to evaluate the accuracy of the Gemini System with new technological flash glucose monitoring (FGM) feature enhancements compared to reference glucose measurements and the Eversense 365 CGM System. The investigation will also evaluate safety of the Gemini System usage.
Interventions
Eversense 365 Continuous Glucose Monitoring System and Gemini Glucose Monitoring System
Sponsors
Study design
Eligibility
Inclusion criteria
Subjects meeting all of the following inclusion criteria will be included in this study: 1. Subjects ≥18 years of age 2. Clinically confirmed diagnosis of diabetes mellitus for ≥1 year 3. Subject has signed an informed consent form (ICF) and is willing to comply with protocol requirements
Exclusion criteria
Subjects meeting any of the following
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| System agreement (15/15%, 20/20%, 30/30%, and 40/40%) for paired sensor FGM and reference measurements post-insertion for reference glucose values from 40-400 mg/dL. | 365 days | Effectiveness Objective: To determine accuracy of the Eversense 365 Continuous Glucose Monitoring System and Gemini Glucose Monitoring System. The effectiveness measure will be system agreement (15/15%, 20/20%, 30/30%, and 40/40%) for paired sensor FGM and reference measurements post-insertion for reference glucose values from 40-400 mg/dL. Effectiveness measures will be evaluated descriptively and compared to the Eversense 365 CGM System. |
| Number of device-related or sensor insertion/removal procedure-related serious adverse events | 365 days | Safety Objective: To demonstrate safety of the Eversense 365 Continuous Glucose Monitoring System and Gemini Glucose Monitoring System. Incidence of device-related or sensor insertion/removal procedure-related serious adverse events in clinic and during home use from sensor insertion through removal and follow-up. |
Countries
United States