Keywords
pulse oximetry, pediatrics, Owlet
Brief summary
The purpose of the study is to collect information from the Owlet OSS 3.0-M1 Sensor in toddlers and preschoolers to evaluate accuracy.
Detailed description
The objective of this study is to collect evidence to verify the accuracy of the Owlet OSS 3.0-M1 Sensor to measure SpO2 and pulse rate in children 19-60 months in a simulated use environment as compared to an FDA cleared reference pulse oximeter.
Interventions
Owlet OSS 3.0-M1 Sensor
Application and monitoring with an FDA cleared reference pulse oximeter
Sponsors
Study design
Intervention model description
Prospective study
Eligibility
Inclusion criteria
1. Subject is between 19 months and ≤ 60 months of age. 2. Subject's current body weight is ≥ 6 lbs. 3. Subject/legal guardian is fluent in both written and spoken English. 4. Subject is willing and able to follow study requirements 5. Legal guardian has provided consent.
Exclusion criteria
1. Subject has preexisting cardiovascular or respiratory disease or conditions. 2. Subjects with abnormalities (e.g., limb deformities, open wounds, severe edema) at the planned application sites that would interfere with system measurements. 3. The device(s) are unable to fit properly without risk of injury (e.g., hook of VELCRO is against the skin, wrap is too tight and constricting). 4. Subject has skin irritation at the location of device placement. 5. Subject has known allergies to adhesive.
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Pulse rate accuracy | 2 hours | Average root mean square (ARMS) error between the pulse rate measured by the Owlet OSS 3.0-M1 vs. a reference pulse oximeter |
| SpO2 Accuracy | 2 hours | The average root mean square difference between the Owlet OSS 3.0-M1 Sensor vs. readints from an FDA cleared reference pulse oximeter. |