Validating Dozee for Contactless Monitoring of Heart, Breathing, and Movemen

Clinical Validation of the Dozee System for Contactless Monitoring of Heart Rate, Respiration Rate, Motion, and Bed-Exits

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT07223931
Enrollment
91
Registered
2025-11-03
Start date
2025-09-16
Completion date
2025-12-18
Last updated
2026-07-27

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

No Specific Conditions, No Specific Medical Conditions or Disease States

Brief summary

The Dozee VS system (K221555) is a contactless device that tracks heart rate and breathing rate using a sensor sheet placed under the mattress. It captures body vibrations and converts them into vital signs, which can be monitored remotely in real time. The system also supports add-ons like a blood pressure cuff, pulse oximeter, and thermometer, offering a complete monitoring solution. This study aims to validate the updated AI-powered heart rate and improved breathing rate algorithms across different sensor sheet types to confirm clinical accuracy

Detailed description

The Dozee VS system (K221555), developed by Turtle Shell Technologies, is a non-invasive, contactless device that continuously monitors heart rate (HR) and breathing rate (RR). A thin sensor sheet placed under the mattress detects tiny body vibrations from the heart and lungs, which are processed by smart algorithms and displayed on a secure online dashboard for real-time monitoring. Unlike traditional methods such as ECGs, nasal tubes, or thermistors-which are invasive, require skilled placement, and are uncomfortable for long-term use-Dozee is simple, comfortable, and doesn't rely on patient compliance like wearables do. The system can also connect to add-on devices, including a blood pressure cuff, pulse oximeter, and thermometer, giving a full picture of a patient's vital signs and helping spot early signs of deterioration. Hardware upgrades now include dual-sensor sheets that improve accuracy across different mattresses and patient types. The new HR algorithm uses AI/ML for better accuracy, while the improved RR algorithm performs more reliably even with movement or weak signals. Study Goal: This study aims to validate the updated algorithms (Gen 2) for HR and RR across the three sensor sheet types (DS, S1, and S1-I) to ensure they meet clinical accuracy standards

Interventions

DEVICEDevice: Telemetry monitoring

All participants will be on standard telemetry monitoring for Heart Rate, Respiration Rate and movements using the gold standard (Electrocardiography and Capnography respectively).

Sponsors

Turtle Shell Technologies Pvt. Ltd.
Lead SponsorINDUSTRY
Eastside Research Associates
CollaboratorUNKNOWN

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to 100 Years
Healthy volunteers
Yes

Inclusion criteria

Subjects will be included if they meet the following inclusion criteria: 1. Adults Male or female. (Above 18 years of age) 2. Provide written informed consent. 3. Weight between 40kg/88.18lbs and 120kg/264.55lbs 4. Agree to not eat during the testing period 5. Able to move and exit the bed independently and safely.

Exclusion criteria

Subjects will be excluded if they have any of the following inclusion criteria: 1. Are connected to a device that may interfere with the device monitoring in this study. 2. Are receiving any bedside care which may be incompatible with the study procedures. 3. A likely need to receive or undergo a procedure during the testing period. 4. Cannot accept a nasal cannula, or have an ECG leads placed on the chest. 5. Have a significant medical condition in the judgement of the investigator, which may compromise the study testing procedures.

Design outcomes

Primary

MeasureTime frameDescription
Mean Absolute Error (MAE) of Dozee's Heart Rate a from the Gold Standard ECG6 monthsHeart Rate collected at 1 Min intervals from gold standard and Dozee. Mean Absolute Error (MAE) of the Dozee Heart Rate will be calculated.
Accuracy (%) of Dozee in identfying subject movement6 monthsTime stamps of subject movement during the study will be recorded manullay. Time stamps of Dozee's movement will be recorded. Accuracy of the Dozee device in detecting movement compared to manual observation will be computed in terms of percentage.
Accuracy(%) of of Dozee's bed-exits6 monthsTime stamps of bed-exits will be recorded manually during the study will be recorded. Time stamps of Dozee's bed-exit will be computed. Accuracy of the Dozee device in detecting bed-exit compared to manual observation as gold standard will be computed in terms of percentage.
Mean Absolute Error (MAE) of Dozee's Respiratory Rate from the Gold Standard Capnography6 MonthsRespiratory Rate collected at 1 Min intervals from gold standard and Dozee. Mean Absolute Error (MAE) of the Dozee Respiratory Rate will be calculated.

Secondary

MeasureTime frameDescription
Range of values (minimum, maximum, and spread) of Heart Rate in beats per min measured by the gold standard and the Dozee6 monthsThe outcome will be assessed by comparing the range of values (minimum, maximum, and spread) of Heart Rate ( beats per minute) measured by the gold standard and the Dozee over the study period.
Range of values (minimum, maximum, and spread) of Respiratory Rate in breaths per min measured by the gold standard and the Dozee6 MonthsThe outcome will be assessed by comparing the range of values (minimum, maximum, and spread) of Respiratory Rate ( breaths per minute) measured by the gold standard and the Dozee over the study period.
Bland- Altman Level of Agreement between Dozee Heart Rate and Respiratory Rate with the Gold Standard6 monthsLevel of Agreement between Dozee's HR and RR with reference device computed by Bland Altman.
Detection rate (%) of the Dozee's Heart Rate and Respiratory Rate6 monthsThe time-stamps of Dozee's Heart Rate and Respiratory Rate and the gold standard Heart Rate and Respiratory Rate will be recorded. DR is the proportion of gold standard points for which the DS successfully provides a corresponding measured value, relative to the total number of gold standard points.
Level of Agreement between Dozee Heart Rate and Respiratory Rate with the Gold Standard6 monthsLevel of Agreement between Dozee Heart Rate and Respiratory Rate with the Gold Standard by using Deming Regression with Pearson Co-efficient

Countries

United States

Contacts

PRINCIPAL_INVESTIGATORSapna Chilka, MD

ERA Health Research