Evaluation of the Accuracy and Safety of A Novel Real-Time Continuous Glucose Monitoring System

Evaluation of the Accuracy and Safety of A Novel Real-Time Continuous Glucose Monitoring System

Status
Recruiting
Phases
Unknown
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT07092761
Enrollment
82
Registered
2025-07-30
Start date
2025-04-01
Completion date
2026-09-01
Last updated
2026-05-29

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Type1diabetes

Brief summary

The study is to evaluate the accuracy and safety of a novel real-time CGM system among adult patients with type 1 diabetes mellitus with respect to YSI reference venous plasma sample measurements.

Detailed description

The study is to evaluate the accuracy and safety of a novel real-time CGM system among adult patients with type 1 diabetes mellitus with respect to YSI reference venous plasma sample measurements. No less than 42 patients will be enrolled. Each patient will wear two sensors. Interstitial glucose readings from each sensor will be obtained immediatedly following each venous blood sample test. Each participant will make 4 visits to the clinical study site, including the screening visit.

Interventions

DEVICEGS3 CGM

Wearing CGM for a consecutive 14 days

Sponsors

Henan University of Science and Technology
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Age ≥ 18 years * Diagonsed with T1DM or T2DM * Venous blood sampling access can be established in the forearm * Capable of independently reading instructions and complying with the clinical trial requirements * Willing to sign the Informed Consent Form (ICF)

Exclusion criteria

* Severe hypoglycemia within the past 6 month * Heart failure or hemiplegic sequelae due to prior cerebrovascular disease * Severe skin conditions at the sensor wear site * Extensive systemic skin disorders * Coagulation disorders confirmed by the investigator * Anemia or abnormal hematocrit * Blood donation within the past 6 months * Pregnancy (defined as positive urine test in women ≤55 years), lactation, or plans for pregnancy within ≤30 days * Current or recent (≤1 month) participation in other clinical trials * Planned MRI/CT scans during sensor wear * Allergy to medical adhesives or alcohol * Conditions impairing comprehension of informed consent or study procedures * Other exclusionary conditions per investigator's discretion

Design outcomes

Primary

MeasureTime frameDescription
System performance14 daysThe CGM system performance will be assessed by comparing to the Yellow Springs Instrument comparator venous plasma measurements
System-related Safety14 daysThe CGM system will be characterized by adverse device effects experienced by the study participants

Countries

China

Contacts

CONTACTPeng Liu
peng.liu87@outlook.com0379-64811993
PRINCIPAL_INVESTIGATORHongwei Jiang

Henan University of Science and Technology

Outcome results

None listed