Cardio-neural Pacing to Modulate AV Conduction in Persistent AF Patients - a Feasibility Study

Cardio-neural Pacing to Modulate AV Conduction in Persistent AF Patients - a Feasibility Study

Status
Not yet recruiting
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT07007000
Enrollment
10
Registered
2025-06-05
Start date
2025-07-31
Completion date
2027-03-31
Last updated
2025-06-05

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Cardio-neural Pacing

Brief summary

Atrial fibrillation is the commonest arrhythmia with a lifetime risk of one in 3-5. In patients with late stage of persistent atrial fibrillation, rate control is usually preferred over rhythm control. Besides medication therapy, atrioventricular nodal ablation is sometimes required with placement of a pacemaker afterwards. The AV node is being innervated by parasympathetic fibres that modulate its conduction. Ablation in these parasympathetic innervations around the coronary sinus (at the CS ostium or posterior to it) has been shown to be promising for treating vagal mediated syncope. Stimulation of these fibres with high frequency pacing could achieve chronic heart rate suppression in animal model. Intermittent pacing in this area has also been shown to be successful in reducing ventricular rate in atrial fibrillation to prevent inappropriate ICD shock. As a proof-of-concept case, we attempted pacing cardio-neural fibres in one of our patients. Pacing 30Hz at 10mA, 2ms pacing in ostial or postero-septal coronary sinus both resulted in a dose dependent prolongation of VV cycle length during atrial fibrillation. Patient did not complain of discomfort during such pacing. We postulate that pacing these fibers can achieve rate control and avoid the need for rate control medication or AV node ablation in some of these patients. This study aims to evaluate safety and efficacy of temporary cardio-neural pacing (CNP), and collect fluoroscopic images and electroanatomical mapping data on cardio-neural pacing sites.

Detailed description

This is a prospective cohort study. 10 patients with persistent atrial fibrillation with pacing indication will be recruited. They are all scheduled to undergo permanent transvenous pacemakers according to clinical need. Informed consent will be signed before procedure. Arterial line will be inserted for monitoring of blood pressure. After routine ventricular lead implantation, a coronary sinus sheath (attain command or deflectable) will be delivered to RA septum, posterior to CS ostium, at the expected location of parasympathetic ganglion plexus. Pace mapping will be performed with a pacing lead (such as Select Secure 3830) at 30Hz, variable amplitude (20, 10, 5V) at 1ms pulse width. An electrophysiology catheter may be used where necessary for pace-mapping the response. At the site where lowest output can generate 30% prolongation of ventricular CL, the lead is fixed for 2-5mm depth. The output is tested again to achieve 50% prolongation of ventricular CL (at 20, 10, 5, 2, 1V). Fluoroscopic image will be collected with contrast injection at the sheath. Repositioning of lead can be performed if the rate suppression cannot be achieved with 3V@0.4ms. Eventually, the implanted lead will serve as an atrial sensing and pacing lead and will not be removed at the end of procedure. Patient will be connected to a dual chamber pacemaker with device programming per usual clinical care. Patients will be followed up for 3 months.

Interventions

After routine ventricular lead implantation, a coronary sinus sheath will be delivered to RA septum, posterior to CS ostium, at the expected location of parasympathetic ganglion plexus. Pace mapping will be performed with a pacing lead at 30Hz, variable amplitude (20, 10, 5V) at 1ms pulse width. An electrophysiology catheter may be used where necessary for pace-mapping the response. At the site where lowest output can generate 30% prolongation of ventricular CL, the lead is fixed for 2-5mm depth. The output is tested again to achieve 50% prolongation of ventricular CL (at 20, 10, 5, 2, 1V). Fluoroscopic image will be collected with contrast injection at the sheath. Repositioning of lead can be performed if the rate suppression cannot be achieved with 3V@0.4ms. Implanted lead will serve as an atrial sensing and pacing lead and will not be removed at the end of procedure. Patient will be connected to a dual to a dual chamber pacemaker with device programming per usual clinical care.

Sponsors

Chinese University of Hong Kong
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Patients with persistent atrial fibrillation who are scheduled for pacemaker implantation per guideline indication

Exclusion criteria

* Patients who cannot provide informed consent * Patients \< 18 years old * Pregnant patients * Illiterate patients

Design outcomes

Primary

MeasureTime frameDescription
Percentage prolongation of CLIn implant procedureMeasure percentage prolongation of CL with lead fixation

Secondary

MeasureTime frameDescription
ThresholdIn implant procedureDetermine the lowest threshold to achieve prolongation
DiscomfortIn implant procedureAny discomfort from patient during cardio-neural pacing
Average arterial line blood pressureIn implant procedureEvaluate average arterial line blood pressure during cardio-neural pacing, compared with that before the pacing.
Fluoroscopic imagingIn implant procedureCollect fluoroscopic imaging to identify location and range of anatomical target area