Pediatric Expansion Study of the Sentio System

A Prospective, Multicenter, Single-arm Clinical Investigation of the Safety and Performance of the Sentio System in a Pediatric Population

Status
Active, not recruiting
Phases
Unknown
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT06976086
Acronym
BC119
Enrollment
50
Registered
2025-05-16
Start date
2025-04-30
Completion date
2027-07-01
Last updated
2026-08-14

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hearing Loss, Conductive, Hearing Loss, Mixed, Single Sided Deafness

Keywords

Sentio system, Paediatric, Hearing loss, Children, Bone conduction, Transcutaneous

Brief summary

This study is performed to collect safety and performance data supporting the use of the Sentio system in children below 12 years of age. The system is currently available on the market for patients 12 years and above. It is a prospective study conducted at eight European hospitals, aiming to treat 50 patients with a hearing loss than would benefit from treatment with a bone-anchored hearing system (BAHS). The primary objective of the study is to demonstrate that the Sentio system improves the hearing on the implanted ear, compared with the unaided situation.

Detailed description

This study is performed to collect safety and performance data supporting the use of the Sentio system in children below 12 years of age. The system is currently available on the market for patients 12 years and above. It is a prospective, multicentre study sponsored by Oticon Medical AB, which will be conducted at eight hospitals across Europe (UK, Spain and the Netherlands). A total of 50 patients with a hearing loss than would benefit from treatment with a bone-anchored hearing system (BAHS) will be included. Participation is voluntary and the decision to participate must be well-founded and well informed. The Sentio system is a transcutaneous bone conduction hearing system (an implant and an external sound processor) for individuals with certain types of hearing losses. The implant is implanted into the bone behind the ear (mastoid and temporal bone area). After a period of skin healing, a sound processor which transmits sound waves to the implant can be magnetically attached. The implant transforms the sound waves to vibrations, which are then transmitted to the inner ear via the bone. The primary objective of the study is to demonstrate that the Sentio system improves the hearing on the implanted ear, compared with the unaided situation. The study includes eight study visits; one before surgery, the surgery visit, and six follow-up visits, with the last visit taking place 12 months after surgery.

Interventions

DEVICEActive transcutaneous bone conduction system

The Sentio system consists of the Sentio 1 Mini sound processor (SP) and Sentio Ti implant with fixation band and screws.

Sponsors

Oticon Medical
Lead SponsorINDUSTRY
Guy's and St Thomas' NHS Foundation Trust
CollaboratorOTHER
Birmingham Women's and Children's NHS Foundation Trust
CollaboratorOTHER
Great Ormond Street Hospital for Children NHS Foundation Trust
CollaboratorOTHER
University Medical Center Groningen
CollaboratorOTHER
Radboud University Medical Center
CollaboratorOTHER
Hospital Universitario La Fe
CollaboratorOTHER
Hospital Universitario Virgen Macarena
CollaboratorOTHER
Hospital Donostia
CollaboratorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
3 Years to 11 Years
Healthy volunteers
No

Inclusion criteria

1. Signed Informed Consent Form (depending on age of child, signed by parent or legal guardian and child) 2. Subject aged 3 to 11 years 3. Subject with: 3.1 conductive or mixed hearing losses with pure tone average (PTA) bone conduction (BC) threshold (measured at 0.5, 1, 2 and 3 kHz) of the indicated ear better than or equal to 45 dB HL. 3.2 OR subject who has a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear (i.e. SSD). The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should then be better than or equal to 20 dB HL (measured at 0.5, 1, 2 and 3 kHz) 3.3 OR subject indicated for an air-conduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS. 4. Subject and parent or legal guardian have the ability and willingness to comply with investigational procedures/ requirements, as determined by the investigator. 5. Subject with prior experience of amplified sound through a properly fitted hearing aid, a CROS device, or a non-surgical bone anchored solution (e.g. softband). 6. For subject with conductive or mixed hearing losses, ensure sufficient air bone gap (ABG) at the ear to be implanted. 7. Sufficient bone quality/quantity/depth or skull size for implantation of a Sentio Ti implant as assessed according to clinical practice.

Exclusion criteria

1. Medical condition that contraindicates implant surgery or anesthesia as judged by the investigator. 2. Untreated ongoing middle ear infection at the time of surgery. 3. Known or suspected contact allergy to silicone or other material used in the Sentio system. 4. Known condition that could jeopardize wound healing and skin condition, e.g. uncontrolled diabetes over time, as judged by the investigator. 5. Known skin or scalp conditions that may preclude attachment or interfere with the usage of the sound processor. 6. Evidence of conditions that would prevent good speech recognition potential as determined by good clinical judgment. 7. Any other known condition that the investigator determines could interfere with compliance or investigation assessments. 8. Use of ototoxic drugs that could be harmful to the hearing, as judged by the investigator. 9. Subject that has received radiotherapy in the area of implantation or is planned for such radiotherapy or similar during the investigation period. 10. For bilateral asymmetric candidates, subject already treated with a bone-anchored hearing solution on the side with the best bone conduction thresholds. 11. Known chronic or non-revisable vestibular or balance disorder. 12. Known abnormally progressive hearing loss. 13. Currently participating, or participated within the last 30 days, in another clinical investigation involving an investigational drug or device that could impact the safety or effectiveness of the Sentio system as determined by the investigator. 14. Use of active implantable or body worn device that for medical reasons cannot be removed or discontinued, such as CSF shunts, implantable cardiac pacemakers, defibrillators, or neurostimulators. 15. Known need for frequent MRI investigations for follow-up of other diseases. 16. Any subject that according to the Declaration of Helsinki is deemed unsuitable for investigation enrolment.

Design outcomes

Primary

MeasureTime frameDescription
Demonstrate that the Sentio system improves hearing on the implanted ear.3 months post-surgeryFunctional gain with the Sentio system, i.e. the difference between pre-operative unaided and post-operative aided sound field thresholds on the implanted ear. The functional gain is calculated as the average of frequencies 500, 1000, 2000 and 4000 Hz (PTA4).

Secondary

MeasureTime frameDescription
Evaluation of the safety profile of the Sentio system in terms of the occurrence of adverse events and serious adverse events related to the device.3 months post-surgeryTabulated adverse events and serious adverse events related to the device reported from surgery throughout the investigation reported at 3 months post-surgery.
Functional gain12 months post-surgeryFunctional gain is used to assess the performance of the Sentio system. The functional gain is a calculation and is defined as the difference between unaided and aided sound field thresholds. The average functional gain is calculated as the average of frequencies 500, 1000, 2000 and 4000 Hz (PTA4).
Effective gain12 months post-surgeryEffective gain is used to assess the degree the Sentio system compensates for the hearing loss on the implanted ear in patients with mixed/conductive hearing losses. Effective gain is a calculation and is defined as the aided threshold relative to the BC (unmasked bone conduction) threshold and is calculated as the difference in dB between aided sound field thresholds and BC in situ hearing thresholds. The average effective gain (PTA4) is calculated as the average of frequencies 500, 1000, 2000 and 4000 Hz.
Assessment of speech recognition improvement3 months post-surgeryDifference in speech recognition score (%) between pre-operative unaided condition and Sentio-aided condition, measured in quiet on the implanted ear.
Assessment of listening-related fatigue6 months post-surgeryChange in Vanderbilt Fatigue Scale (VFS) questionnaire score between pre-operative unaided condition at baseline and aided condition post-surgery. The Vandebilt Fatigue Score questionnaire as a valid measure of listening-related fatigue in children.
Assessment of quality of life12 months post-surgeryGlasgow Children Benefit Inventory (GCBI) questionnaire score addressing the consequences of a specific intervention on various aspects of the child's day-to-day life. The score goes from -100 (max negative effect) to +100 (max positive effect).
Usage hours12 months post-surgerySound processor usage hours per day.
Duration of surgery3 months post-surgeryAverage length of surgery measured in minutes.
IPS (Inflammation, Pain, Skin numbness) scores12 months post-surgeryDistribution of IPS scores (Inflammation score 0-4; Pain score 0-2; Skin numbness score 0-2) assigned by investigator. The IPS score is presented as \[Ix Px Sx\] with x being the individual score for each parameter. A higher score corresponds to a poorer outcome.

Countries

Netherlands, Spain, United Kingdom