Clinical Validation of the Blood Pressure Measuring Device Withings BPM Pro 2 (WIHYP-GP)

Clinical Validation Of The Brachial Blood Pressure Measuring Device Withings BPM Pro 2 According To "The Universal Protocol For The Validation Of Blood Pressure Measuring Devices By The Association For The Advancement Of Medical Instrumentation / European Society Of Hypertension / International Organization For Standardization (AAMI/ESH/ISO) (And Its Amendment 1 (2020) And 2 (2024))" In The General Population.

Status
Completed
Phases
Unknown
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT06957847
Acronym
WIHYP-GP
Enrollment
89
Registered
2025-05-05
Start date
2025-04-28
Completion date
2025-11-18
Last updated
2026-02-18

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hypertension

Brief summary

The aim of the study is to assess the accuracy of the automatic oscillometric BP measuring device at the brachial level, the WITHINGS BPM Pro 2, in the general population

Interventions

DEVICEBlood pressure measurement

The patient will be measured alternatively with a reference mercury sphygmomanometer (four times) and the device under study (3 times)

Sponsors

Withings
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DEVICE_FEASIBILITY
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* Patient older than 18 years; * Patient who signed the informed consent form; * Patient followed-up at site (in-patient or out-patient); * Patient with arm circumference between 22 cm and 42 cm.

Exclusion criteria

* Patient unable to give a consent or understand properly protocol information; * Patient suffering from arrhythmia; * Patient with poor quality of Korotkov sounds; * Patient for whom K5 sounds are absent; * Patient wearing an implantable electric medical device (pacemaker,…); * Patient with both upper arms suffering from open wound and/or damaged skin.

Design outcomes

Primary

MeasureTime frameDescription
Blood pressure measurement performance evaluation per measureThe time frame is three months.The blood pressure (systolic and diastolic blood pressure) measurement performance will be established using the mean value of the differences and the standard deviation between the device under study and the reference for each measure.
Blood pressure measurement performance evaluation by subjectThe time frame is three months.The blood pressure measurement performance by subject will be established using the mean value of the differences and the standard deviation between the averaged per subject value of blood pressure (systolic and diastolic blood pressure) device under study and its reference.

Secondary

MeasureTime frameDescription
Safety use of the deviceThe time frame is three months.The rate of adverse effects.