Oxygen Saturation Measurement
Conditions
Keywords
pulse oximetry, Hypoxia
Brief summary
Evaluation of Philips FAST technology with Nellcor sensors
Detailed description
Evaluate saturation accuracy claims for the Philips picoSAT X SpO2 board with FAST algorithm, paired with Nellcor™ sensors in a diverse subject population, over a specified saturation range
Interventions
Noninvasive pulse oximeter
Sponsors
Study design
Eligibility
Inclusion criteria
1. Subjects ≥ 18 and \< 51 years of age 2. Subject is willing and able to comply with study procedures and duration 3. Subject is willing to sign an informed consent 4. Subject weighs \>40kg 5. Subject is a non-smoker or has not smoked within 36 hours prior to the study 6. Cleared same day health assessment form and health screening
Exclusion criteria
1. Subject is considered as being morbidly obese (defined as BMI \>39.5) 2. Compromised circulation, injury, or physical malformation of fingers, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the sites utilized) 3. Participants of childbearing potential who are pregnant, who are trying to get pregnant, or who have a urine test positive for pregnancy on the day of the study 4. Subjects with COHb levels \>3% as assessed by CO-Oximetry during the procedure 5. tHb \< 10 g/dl as assessed by CO-Oximetry during the procedure 6. MetHb ≥ 2% as assessed by CO-Oximetry during the procedure 7. Subjects with known respiratory conditions such as: 1. uncontrolled / severe asthma 2. flu or influenza type infection 3. pneumonia / bronchitis 4. shortness of breath / respiratory distress 5. unresolved respiratory or lung surgery 6. emphysema, COPD, lung disease 7. recent COVID (last 2 months) 8. Subjects with known heart or cardiovascular conditions such as: 1. hypertension: systolic pressure \>140mmHg, or diastolic pressure \>90mmHg on 3 consecutive readings 2. previous cardiovascular surgery 3. chest pain (angina) 4. heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia (reviewed during health screen) 5. previous heart attack 6. blocked artery 7. unexplained shortness of breath 8. congestive heart failure (CHF) 9. history of stroke 10. transient ischemic attack 11. carotid artery disease 12. myocardial ischemia 13. myocardial infarction 14. cardiomyopathy 15. implantable active medical device such as pacemaker or automatic defibrillator 9. Self-reported health conditions as identified in the Health Assessment Form 1. diabetes 2. uncontrolled thyroid disease 3. kidney disease / chronic renal impairment 4. history of seizures (except childhood febrile seizures) 5. epilepsy 6. history of unexplained syncope 7. recent history of frequent migraine headaches 8. recent symptomatic head injury, within the last 2 months 9. cancer requiring chemotherapy, radiation, or current treatment 10. subjects with known clotting disorders 11. history of bleeding disorders or personal history of prolonged bleeding from injury 12. history of blood clots 13. hemophilia 14. sickle cell trait or disease 15. current use of blood thinner: prescription or daily use of aspirin 16. subjects with Severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors 17. participants with severe allergy to iodine (only applicable if iodine is used) 18. subjects with prior or known severe allergies to lidocaine (or similar pharmacological agents, e.g. Novocain) or heparin 19. arterial cannulation within the last 30 days prior to study date, (this may exclude only one radial artery site, left or right) 20. history of clinically significant complications from previous arterial cannulation 21. Failure of Allen's test 22. unwillingness or inability to remove colored nail polish or colored artificial nails other than clear from test digits 23. other known health condition, upon disclosure in Health Assessment form at PI's discretion
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Evaluation of Saturation Accuracy | 2 - 3 hours | Performance will be determined by calculating the accuracy in terms of root mean square difference (ARMS) comparing the noninvasive oxygen saturation measurement (SpO2) to the arterial oxygen saturation (SaO2) value obtained from a reference blood sample. |
Countries
United States