This Study Aims to Assess the Performances of Optical Blood Pressure Monitoring Device, tBPC Product With Reference to ISO 81060-3:2022

A Photoplethysmography-Based Continuous Automated Non-Invasive Sphygmomanometer: Performance Evaluation in Stability and Blood Pressure Change Tracking

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT06815614
Enrollment
40
Registered
2025-02-07
Start date
2023-11-29
Completion date
2024-11-26
Last updated
2025-02-07

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Blood Pressure, Blood Pressure Variability

Keywords

ISO 81060-3:2022, Photoplethysmography

Brief summary

In this study, according to the latest released ISO 81060-3:2022 standard requirements, non-invasive blood pressure measurement by cuff at the start and the end time points will be used as oCareTM BP100 start-to-end blood pressure calibration. The invasive arterial catheter (A-line) blood pressure measurement will be used as a reference invasive blood pressure monitor (BPM) equipment. The purpose of the study is to evaluate the stability of the device under evaluation (DUE) and its performance in blood pressure change tracking.

Detailed description

1. The primary goal (Stability Test): to validate the stability of the repeated measurements of DUE, and the corrected standard deviation of each analysis period shall not exceed 6 mmHg. 2. The secondary goal (Blood Pressure Change Tracking): to validate the consistency of DUE vs. the reference BPM equipment in blood pressure change tracking during the change evaluation interval.

Interventions

DEVICEA Photoplethysmography-Based Continuous Automated Non-Invasive Sphygmomanometer

The investigation medical device under evaluation is oCareTM BP 100, a non-invasive continuous blood pressure monitoring system developed by Taiwan Biophotonic Co. (tBPC). The main function is to use a fingerclip sensor to receive PPG signals and collect continuous blood pressure values and uninterruptedly within the initialization time. In this study, according to the latest released ISO 81060-3:2022 standard requirements, non-invasive blood pressure measurement by cuff at the start and the end time points will be used as oCareTM BP 100 start-to-end blood pressure calibration. The A-line invasive arterial catheter blood pressure measurement will be used as a reference invasive BPM equipment. The purpose of the study is to confirm the stability of the investigational medical device under evaluation and its performance of blood pressure change tracking.

Sponsors

Taiwan Biophotonic Corporation
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
22 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1. Subject must be over 22 years old on the date of informed consent. 2. Subject who has undergone radial artery cannulation (A-line) (for medical purposes, specifically for arterial pressure monitoring. 3. Subject or his/her legal representative understand the written informed consent and willing to participate as evidenced by signing the informed consent.

Exclusion criteria

1. Pregnancy 2. Patients with an intra-aortic balloon pump 3. Patients with arm or finger injuries 4. Patients who have undergone vascular surgery for upper extremities or contraindications for local treatment upon upper extremities 5. Patients with A-line placement in the brachial artery 6. Participants are considered not suitable for inclusion after evaluation by the researchers (such as poor compliant mental status or health conditions, etc.) 7. Participants or his/her legal representative refuse to sign the informed consent form

Design outcomes

Primary

MeasureTime frame
Confirm the stability of the repeated measurements of SUT by calculating the corrected standard deviation of each analysis period4 hours
Confirm the consistency of SUT vs. the reference BPM equipment by calculating the blood pressure change during the change evaluation interval4 hours