Skip to content

Diagnostic Feasibility of 100 Hz Tetanic Stimulation

Diagnostic Feasibility of 100 Hz Tetanic Stimulation in the Perioperative Setting in Patients With Normal and Abnormal ENG: a Pilot Study

Status
Recruiting
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT06754995
Enrollment
40
Registered
2025-01-01
Start date
2025-01-06
Completion date
2027-12-31
Last updated
2025-03-27

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Polyneuropathies

Keywords

electromyography, electroneurography, tetanic stimulation, residual neuromuscular block, neuromuscular monitoring

Brief summary

The goal of this observational study is to investigate the applicability of 100 Hz tetanic stimulation with an electromyography device in the exclusion of postoperative residual neuromuscular block. The aim is to determine whether fatigue develops with 100 Hz tetanic stimulation in anaesthetized, non-relaxed patients with normal or abnormal baseline electroneurographic (ENG) findings.

Detailed description

The aim of the present study is to investigate the applicability of 100 Hz tetanic stimulation with an electromyography (EMG) device. The aim is to determine whether fatigue develops with 100 Hz tetanic stimulation in anaesthetized, non-relaxed patients with normal or abnormal baseline electroneurographic (ENG) findings. Electric responses of the musculus digiti minimi elicited by supramaximal stimulation of the ulnar nerve at the carpal tunnel are monitored. Alpine Biomed ENG-EMG device with Keypoint software version 5.11 is used for the examinations. The exposed arm is fixed on the armboard, the skin is degreased and then the electrodes for monitoring are mounted lege artis. After the anesthesia has been induced, ENG of the ulnar nerve is examined. Based on the results of the ENG test, patients are divided into separate groups (normal or abnormal ENG). During the ENG test, the supramaximal current is determined. Afterwards, repetitive tetanic stimulation of the defined frequency (60 and 100 Hz) and duration (5 sec) is applied to the two test groups at the current according to the test plan. The clinical utilities of the study results are: * Determination of the safety margin of neuromuscular junction, which is not provided by routine monitoring. * To investigate how comorbid status (especially chronic conditions affecting peripheral nerves such as polyneuropathy) influences the feasibility of tetanic stimulation in the diagnostics of residual neuromuscular block and safety margin.

Interventions

None listed

Sponsors

University of Debrecen
Lead SponsorOTHER

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to 65 Years
Healthy volunteers
Yes

Inclusion criteria

* ASA 1-3; * BMI 18.5-25 (normal body weight); * patient is in supine position with one arm accessible.

Exclusion criteria

* diseases with impaired neuromuscular function (myopathies, severe liver and kidney failure); * drugs affecting neuromuscular function (magnesium, aminoglycosides); * pregnancy (pregnancy tests are carried out in women of childbearing age to rule out pregnancy); * breast-feeding; * acute surgery

Design outcomes

Primary

MeasureTime frameDescription
100 Hz tetanic fade response of abductor digiti minimi muscle to ulnar nerve stimulation in patients with normal and abnormal ENGFrom enrollment until end of anesthesia asessed up to 1 hour. No further ENG measurements are performed in the postoperative period.Electroneurography response of the abductor digiti minimi muscle to 100 Hz tetanic stimulation (tetanic fade) is examined to find out if there is difference in patients with or without polyneuropathy. Tetanic fade is the decrease of the twitch heights of the compound muscle action potential due to depletion of cetylcholin stores in the nerve terminal during tetanic stimulation.

Secondary

MeasureTime frameDescription
Diagnostic feasibility of electromyography in the examination of safety marginFrom enrollment until end of anesthesia assessed up to 1 hour. No further ENG measurements are performed in the postoperative period.To see whether tetanic stimulation with electromyography can be used as a routine tool to examine the safety margin and residual neuromuscular block. It is expected that there will be no fade in muscle response to 60 Hz tetanic stimulation in either groups. It is suspected that there will be no tetanic fade response to 100 Hz tetanic stimulation in patinets with normal ENG, however patinets with abnormal baseline ENG will produce tetanic fade. The incidence and extent of tetanic fade response will be examined.

Other

MeasureTime frameDescription
60 Hz tetanic fade response of abductor digiti minimi muscle to ulnar nerve stimulation in patients with normal and abnormal ENGFrom enrollment until end of anesthesia assessed up to 1 hour. No further ENG measurements are performed in the postoperative period.Electroneurography response of the abductor digiti minimi muscle to 60 Hz tetanic stimulation (tetanic fade) is examined to find out if there is difference in patients with or without polyneuropathy. Tetanic fade is the decrease of the twitch heights of the compound muscle action potential due to depletion of cetylcholin stores in the nerve terminal during tetanic stimulation.

Countries

Hungary

Contacts

Primary ContactAdrienn Pongrácz, MD PhD
pongadri@gmail.com+36302390132
Backup ContactRéka Nemes, MD PhD
reka.nemes.11@gmail.com+36309668722

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026