Standard Post Market Clinical Follow-up (PMCF) Study WSA @ HZO

Standard PMCF Study WSA @ HZO

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT06700512
Enrollment
20
Registered
2024-11-22
Start date
2024-06-14
Completion date
2024-08-26
Last updated
2025-08-08

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hearing Loss, Sensorineural

Keywords

Hearing aid, Freiburger, Matrix Test, SSQ, IOI-HA

Brief summary

The purpose of the study is to follow-up on a released low-risk device as part of standard PMCF activities

Detailed description

This is a comparative study which is conducted monocentric at Hoerzentrum Oldenburg, Germany. The focus is to test a released hearing aid and compare this to the subject's own hearing aids with a minimum of 16 mild-to-moderate sensorineural hearing impaired adult subjects. The subjects, all of them native German speakers, have 2 appointments of 1-2 hours including a fitting procedure with real-ear measurements and some standard speech tests and undergo two 2-weeks home trials (with questionnaires).

Interventions

DEVICEHearing Aid Signia Pure C&G 7IX

Receiver in Canal (RIC) hearing aid, M Receiver

subjects' own hearing aids, any brand, any model

Sponsors

WSAUD A/S
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Hearing loss should fall within fitting range of study hearing aid * They should be experienced hearing aid users * Air-Bone-Gap should be less or equal to 20 dB Hearing Loss (HL). * Sensorineural HL, mild-to-moderate * HL corresponding to the fitting range of the hearing aid * Healthy (outer) ear * Older than 18 years * German is mother tongue * Able to understand the instructions * Willing to participate in laboratory tests and to wear the hearing aids at home for 2 weeks * Informed consent

Exclusion criteria

* Contraindication for hearing aid treatment * Fluctuating or rapidly progressing hearing loss * Central hearing problems * Limited mobility * Limited dexterity (in handling the hearing aid) * Known psychological problems

Design outcomes

Primary

MeasureTime frameDescription
Oldenburg Sentence TestEstimated Time Frame 2 weeks (Test conducted at second lab session).The Oldenburg Sentence Test (OLSA) is an audiometric test for determining the speech recognition threshold (SRT in dB). Sentences of the form name - verb - numeral - adjective - noun are used as speech material. The sequence of words is a random combination from an inventory of 50 words in total. The design of the test prevents memorization of the sentences, so the OLSA can be repeated as often as desired. For this study speech in noise recognition was measured. Noise level was 65 dB SPL played from the back, speech was presented from front. The SRT shows the signal-to-noise ratio needed for 80% speech recognition. Lower values mean better speech understanding (i.e. the speech can be quieter with respect to the noise and still be understood). There are no minimum or maximum values.

Secondary

MeasureTime frameDescription
Freiburger Monosyllabic Speech TestEstimated Time Frame 2 weeks (Test conducted at second lab session).To investigate speech intelligibility (SI) performance for speech at 65dB in a quiet listening situation, the Freiburger monosyllabic speech test is used which is still the gold standard for Hearing Care Professionals (HCPs) in Germany. Results are reported in Percent correct data. Minimum is 0%, maximum is 100% with higher values meaning better speech understanding.
Match-to-Target1 day (Fitting procedure with real-ear measurements is done at first lab session)Investigate how close the study hearing aids fitted and fine-tuned with the fitting software to NAL-NL2 Targets match the postulated targets, as measured in-situ with real-ear measurements. Absolute deviations were calculated at specific frequencies ranging from 250 Hz to 4 kHz. Unit of measure is dB SPL. Minimum Value is 0 (i.e. measured value and target value identical), no maximum value. Higher values mean greater deviation from target.
Speech, Spatial and Qualities of Hearing Questionnaire (SSQ)2 weeksA short form of the SSQ is used (SSQ 17, German version), to investigate speech perception, sound localisation, quality of hearing perceived in real-life. Questions are answered on a scale from 0 to 10, higher scores mean better outcomes. The responses to the individual questions are grouped for the subsections (speech, spatial, quality) by calculating the means.
International Outcome Inventory for Hearing Aids (IOI-HA)2 weeksInvestigate outcomes and effectiveness of hearing aid usage in real-life. Responses are given on a scale from 1 - 5. Here, the mean of questions 2 -7 is reported. Higher values represent better outcomes. Question 1 was on wearing time, which is not considered a rating of effectiveness and was therefore excluded from the mean and will be reported separately. Fpr wearing times higher values represent longer wearing time (maximum 5 equals more than 8 hours a day).
Satisfaction Questionnaire2 weeksInvestigate satisfaction in real-life for different aspects such as sound quality, naturalness of sound, conversation with one person and overall satisfaction. Each aspect was assessed on on a 7-point likert scale, going from 1 (very dissatisfied) to 7 (very satisfied) The overall satisfaction and sound quality overall were individual questions and are not the combination of different situations/aspects. These questions were also assessed on on a 7-point likert scale, going from 1 (very dissatisfied) to 7 (very satisfied). Higher values represent better outcomes.

Countries

Germany

Participant flow

Participants by arm

ArmCount
Study Participants
within subject design: compare Pure C&G 7IX to subjects' own hearing aids. Baseline measures are the same for all study participants
20
Total20

Baseline characteristics

CharacteristicStudy Participants
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
19 Participants
Age, Categorical
Between 18 and 65 years
1 Participants
mild to moderate Hearing Impairment20 Participants
Race and Ethnicity Not Collected— Participants
Region of Enrollment
Germany
20 participants
Sex: Female, Male
Female
10 Participants
Sex: Female, Male
Male
10 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 200 / 20
other
Total, other adverse events
0 / 200 / 20
serious
Total, serious adverse events
0 / 200 / 20

Outcome results

Primary

Oldenburg Sentence Test

The Oldenburg Sentence Test (OLSA) is an audiometric test for determining the speech recognition threshold (SRT in dB). Sentences of the form name - verb - numeral - adjective - noun are used as speech material. The sequence of words is a random combination from an inventory of 50 words in total. The design of the test prevents memorization of the sentences, so the OLSA can be repeated as often as desired. For this study speech in noise recognition was measured. Noise level was 65 dB SPL played from the back, speech was presented from front. The SRT shows the signal-to-noise ratio needed for 80% speech recognition. Lower values mean better speech understanding (i.e. the speech can be quieter with respect to the noise and still be understood). There are no minimum or maximum values.

Time frame: Estimated Time Frame 2 weeks (Test conducted at second lab session).

ArmMeasureValue (MEAN)Dispersion
Results With Study DeviceOldenburg Sentence Test-6.6 dBStandard Deviation 4.5
Results UnaidedOldenburg Sentence Test-1.1 dBStandard Deviation 7.5
Results With Own DevicesOldenburg Sentence Test-3.5 dBStandard Deviation 5.1
Comparison: H0 (Objective 1) SRT (study HA) is either not significantly different from SRT without HAs or significantly worse than SRT without HAs.~H1(Objective 1) SRT(study HA) is significantly better than SRT without HAs. A mean difference in SRT of 1 dB measured with the Oldenburg sentence test can be assumed as a typical difference in speech intelligibility to show a clinically meaningful improvement.p-value: <0.001t-test, 1 sided
Comparison: H0 (Objective 2) SRT (study HA) is significantly worse than SRT(own HAs). H1(Objective 2) There is no difference between SRT(study HA) and SRT(own HAs) or SRT (study HA) is significantly better than SRT (own HAs) SRT= Speech Reception Threshold in dBp-value: <0.001t-test, 1 sided
Secondary

Freiburger Monosyllabic Speech Test

To investigate speech intelligibility (SI) performance for speech at 65dB in a quiet listening situation, the Freiburger monosyllabic speech test is used which is still the gold standard for Hearing Care Professionals (HCPs) in Germany. Results are reported in Percent correct data. Minimum is 0%, maximum is 100% with higher values meaning better speech understanding.

Time frame: Estimated Time Frame 2 weeks (Test conducted at second lab session).

ArmMeasureValue (MEAN)Dispersion
Results With Study DeviceFreiburger Monosyllabic Speech Test75.1 Percent CorrectStandard Deviation 12.3
Results UnaidedFreiburger Monosyllabic Speech Test31.5 Percent CorrectStandard Deviation 20.1
Results With Own DevicesFreiburger Monosyllabic Speech Test68.6 Percent CorrectStandard Deviation 12.7
Secondary

International Outcome Inventory for Hearing Aids (IOI-HA)

Investigate outcomes and effectiveness of hearing aid usage in real-life. Responses are given on a scale from 1 - 5. Here, the mean of questions 2 -7 is reported. Higher values represent better outcomes. Question 1 was on wearing time, which is not considered a rating of effectiveness and was therefore excluded from the mean and will be reported separately. Fpr wearing times higher values represent longer wearing time (maximum 5 equals more than 8 hours a day).

Time frame: 2 weeks

Population: The IOI-HA was developed to measure outcomes in audiological rehabilitation using Hearing Aids and is not designed for the unaided condition. For this reason only the two interventional arms are included for this outcome measure.

ArmMeasureGroupValue (MEAN)Dispersion
Results With Study DeviceInternational Outcome Inventory for Hearing Aids (IOI-HA)Mean effectiveness ratings3.6 score on a scaleStandard Deviation 1
Results With Study DeviceInternational Outcome Inventory for Hearing Aids (IOI-HA)Wearing Time5.0 score on a scaleStandard Deviation 0.2
Results UnaidedInternational Outcome Inventory for Hearing Aids (IOI-HA)Mean effectiveness ratings3.8 score on a scaleStandard Deviation 0.7
Results UnaidedInternational Outcome Inventory for Hearing Aids (IOI-HA)Wearing Time4.9 score on a scaleStandard Deviation 0.3
Secondary

Match-to-Target

Investigate how close the study hearing aids fitted and fine-tuned with the fitting software to NAL-NL2 Targets match the postulated targets, as measured in-situ with real-ear measurements. Absolute deviations were calculated at specific frequencies ranging from 250 Hz to 4 kHz. Unit of measure is dB SPL. Minimum Value is 0 (i.e. measured value and target value identical), no maximum value. Higher values mean greater deviation from target.

Time frame: 1 day (Fitting procedure with real-ear measurements is done at first lab session)

ArmMeasureGroupValue (MEAN)Dispersion
Results With Study DeviceMatch-to-TargetDeviation from Target @ 250 Hz3.3 dBStandard Deviation 4
Results With Study DeviceMatch-to-TargetDeviation from Target @ 500 Hz1.6 dBStandard Deviation 1.7
Results With Study DeviceMatch-to-TargetDeviation from Target @ 1000 Hz1.2 dBStandard Deviation 1.4
Results With Study DeviceMatch-to-TargetDeviation from Target @ 2000 Hz1.6 dBStandard Deviation 1.9
Results With Study DeviceMatch-to-TargetDeviation from Target @ 4000 Hz1.5 dBStandard Deviation 1.7
Secondary

Satisfaction Questionnaire

Investigate satisfaction in real-life for different aspects such as sound quality, naturalness of sound, conversation with one person and overall satisfaction. Each aspect was assessed on on a 7-point likert scale, going from 1 (very dissatisfied) to 7 (very satisfied) The overall satisfaction and sound quality overall were individual questions and are not the combination of different situations/aspects. These questions were also assessed on on a 7-point likert scale, going from 1 (very dissatisfied) to 7 (very satisfied). Higher values represent better outcomes.

Time frame: 2 weeks

Population: The satisfaction questionnaire was developed to measure satisfaction with hearing aids and not designed for the unaided condition. For this reason data was only collected for the two interventional arms.

ArmMeasureGroupValue (MEAN)Dispersion
Results With Study DeviceSatisfaction QuestionnaireOverall Satisfaction5.1 score on a scaleStandard Deviation 1.5
Results With Study DeviceSatisfaction QuestionnaireConversation with one person5.4 score on a scaleStandard Deviation 1.6
Results With Study DeviceSatisfaction QuestionnaireNaturalness5.4 score on a scaleStandard Deviation 1.4
Results With Study DeviceSatisfaction QuestionnaireComfort with loud sounds5.0 score on a scaleStandard Deviation 1.3
Results With Study DeviceSatisfaction QuestionnaireSound quality overall5.4 score on a scaleStandard Deviation 1.3
Results UnaidedSatisfaction QuestionnaireComfort with loud sounds4.5 score on a scaleStandard Deviation 1.4
Results UnaidedSatisfaction QuestionnaireSound quality overall5.5 score on a scaleStandard Deviation 1.2
Results UnaidedSatisfaction QuestionnaireOverall Satisfaction5.1 score on a scaleStandard Deviation 1.5
Results UnaidedSatisfaction QuestionnaireNaturalness5.2 score on a scaleStandard Deviation 1.5
Results UnaidedSatisfaction QuestionnaireConversation with one person5.6 score on a scaleStandard Deviation 1.4
Secondary

Speech, Spatial and Qualities of Hearing Questionnaire (SSQ)

A short form of the SSQ is used (SSQ 17, German version), to investigate speech perception, sound localisation, quality of hearing perceived in real-life. Questions are answered on a scale from 0 to 10, higher scores mean better outcomes. The responses to the individual questions are grouped for the subsections (speech, spatial, quality) by calculating the means.

Time frame: 2 weeks

Population: The aim of the SSQ questionnaire was to compare the performance/benefit of the study hearing aids to the subjects' own devices. As the unaided ratings would have been the same for both arms, they would have cancelled out by comparing the two interventions and thus it was considered as not necessary to collect the unaided responses.

ArmMeasureGroupValue (MEAN)Dispersion
Results With Study DeviceSpeech, Spatial and Qualities of Hearing Questionnaire (SSQ)Speech Perception6.2 score on a scaleStandard Deviation 2.3
Results With Study DeviceSpeech, Spatial and Qualities of Hearing Questionnaire (SSQ)Sound Localization6.9 score on a scaleStandard Deviation 2.4
Results With Study DeviceSpeech, Spatial and Qualities of Hearing Questionnaire (SSQ)Quality of Hearing7.9 score on a scaleStandard Deviation 1.3
Results UnaidedSpeech, Spatial and Qualities of Hearing Questionnaire (SSQ)Speech Perception5.8 score on a scaleStandard Deviation 1.9
Results UnaidedSpeech, Spatial and Qualities of Hearing Questionnaire (SSQ)Sound Localization6.2 score on a scaleStandard Deviation 1.6
Results UnaidedSpeech, Spatial and Qualities of Hearing Questionnaire (SSQ)Quality of Hearing7.3 score on a scaleStandard Deviation 1.7