Anterior Segment Optical Coherence Tomography Microstent Positioning

Post-operative Anterior Segment OCT Evaluation of Trabecular Microstent Positioning and the Impact on Intraocular Pressure

Status
Active, not recruiting
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT06666751
Enrollment
116
Registered
2024-10-31
Start date
2024-04-09
Completion date
2027-05-30
Last updated
2026-02-27

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Glaucoma, Intraocular Pressure, Stent Dislodgement, Stent Migration

Brief summary

The goal of this observational study is to learn about the long-term results of glaucoma stent surgery in men and women 18 or older who have had cataract and microstent surgery in the past 2 years at the Moran Eye Center. The main questions the study aims to answer are: 1. How often are microstents mispositioned? 2. What effect does microstent positioning have on intraocular pressure after surgery? Participants will have a one-time clinic visit for an eye exam and Optical Coherence Tomography (OCT) . From the eye exam and OCT, Researchers will assess and categorize the position of the microstents and assess the Schlemm's canal dilation. Researchers will review the participants' medical records to acquire 2-year data for the following: 1. Surgeon type and level of experience (resident, fellow, attending ophthalmologist) 2. Proportions of proper gonioscopic placement intraoperatively for each surgeon group 3. Effectiveness of surgery, based on change in intraocular pressure (IOP) from baseline medicated IOP, number of IOP-lowering medications the patient was able to discontinue, and rate of need for additional glaucoma surgery.

Interventions

PROCEDURECataract plus Microstent surgery

Both the Hydrus microstent and iStent Inject W are used in combination with cataract surgery, in order to help lower intraocular pressure for patients with open-angle glaucoma. The microstent is a small flexible device that is inserted into the eye during cataract surgery in order to improve the outflow of fluid from the eye.

Sponsors

University of Utah
Lead SponsorOTHER
Alcon Research
CollaboratorINDUSTRY

Study design

Observational model
COHORT
Time perspective
RETROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to 100 Years
Healthy volunteers
No

Inclusion criteria

* Any adult glaucoma patient (age 18+) who underwent Hydrus microstent or iStent Inject placement at the Moran Eye Center.

Exclusion criteria

* Any patient who subsequently had to have their stent removed for any reason, any patient with a history of trauma to the anterior segment or reconstruction altering typical anatomy.

Design outcomes

Primary

MeasureTime frameDescription
Difference in percentage of mispositioned stents between Hydrus microstent and iStent groupsTwo years post stent placementPositioning of each participant's stent(s) will be evaluated by anterior segment OCT at the one-time clinical study visit. iStent Inject W mispositions are defined as: microstent protrusion (entire conical head not within canal), stent completely buried in the trabeculae, or not seen (i.e. 0 or only 1 stent). If the iStent position is located but the examiner is unsure if it is positioned correctly, it will be counted as misposition. If the iStent is not seen, an in office gonioscopy will be performed to confirm results. Visualization of three stents is not considered a misposition. Hydrus mispositions are defined as: distal tip not in canal, at least 1 window not in canal, or trabecular microstent not within transition zone. The between group difference of mispositioned stents will be calculated in percentage.

Secondary

MeasureTime frameDescription
Dilation of Schlemm's canalTwo years post stent placementThe Schlemm's canal is a circular lymphatic-like vessel in the eye that maintains fluid homeostasis by draining aqueous humor from the eye into the systemic circulation. If the canal is blocked, the pressure in the eye can increase, resulting in glaucoma. Dilation will be visualized on OCT and graded on a numerical scale 1-3 (1=dilated, 2=some dilation, 3=no dilation) at the one-time study visit.
Change in IOP from baselineTwo years post stent placementIOP is measured in millimeters of mercury (mmHg). A retrospective chart review will be conducted to acquire 2-year data to determine the change in each participant's IOP from baseline. The between group difference of change in IOP will be calculated in percentage.
Reduction in number of IOP lowering medicationsTwo years post stent placementA retrospective chart review will be conducted to determine how many IOP lowering medications each patient is using at 2 years post-stent placement compared to the number of IOP lowering medications they were using prior to stent placement. The between group difference of reduction in number of medications will be calculated in percentage.
Percentage of patients needing additional glaucoma surgeryTwo years post stent placementA retrospective chart review will be conducted to determine the percentage of patients who had additional glaucoma surgery by the two-year post stent placement timepoint. The between group difference in number of patients needing additional glaucoma surgery will be calculated in percentage.
Sub-analysis of primary outcome by surgeon typeTwo years post stent placementThere will be a sub-analysis of the percentage of mispositions (primary endpoint) by level of surgeon experience. A retrospective chart review of the intraoperative note will determine if the primary surgeon was a resident or non-resident (attending or fellow).
Proportion of stent cases with proper placement via anterior segment OCT versus proportion of cases with proper placement via intraoperative gonioscopyTwo years post stent placementThe intraoperative notes at the time of stent placement will include confirmation of proper stent placement via gonioscopy. At the time of the study visit, stent placement will be assessed via anterior segment OCT. The number of correctly positioned stent placements will be reported in percentages.