Birth, Preterm, Newborn
Conditions
Brief summary
The overall purpose is to directly compare pulse rate (PR) acquisition and oxygen saturation performance between two pulse oximeters (PO) in newborns.
Interventions
The Nellcor™ N-600x Pulse Oximetry System with N-600X Pulse Oximeter and Nellcor™ Sensors and Cables with OxiMax technology is indicated for prescription use only for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin and PR. The N-600x Pulse Oximeter is intended for use with neonatal, pediatric, and adult patients during both no motion and motion conditions and for patients who are either well or poorly perfused, in hospitals, hospital-type facilities, intra-hospital transport, and home environments.
The Masimo Rad-97™ and accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin, PR, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or acoustic respiratory rate (RRa). The Masimo Rad-97™ and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused. In addition, the Masimo Rad-97™ and accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Rad-97™ and accessories for functional oxygen saturation of arterial hemoglobin and PR to multi-parameter devices for the display on those devices.
Market-released ECG monitoring system
Sponsors
Study design
Eligibility
Inclusion criteria
* Parent or legal guardian is willing and able to provide informed consent prior to any study procedures being performed. * Subject is expected to be able to participate for the duration of the evaluation in the delivery room. In most instances, the length of the study will not exceed the anticipated duration of ECG monitoring. * Subject is ≥ 29 weeks gestational age. * Subject can accommodate multiple sensors.
Exclusion criteria
* Subjects that have an existing health condition, in which the investigator determines that safe or accurate oximetry measures may not be obtained at the areas where the PO sensor could be placed. For example, major congenital anomalies, limb anomalies or anyone with abrasions or surface lesions on the contact areas would be excluded. * Any subject whose treatment or condition would be compromised if they were to participate in this study and any patient whose use of a pulse oximeter and sensor would present a potential for serious risk to health, safety, or welfare. * Currently participating in any other study expected to confound study results in the opinion of the investigator.
Design outcomes
Primary
| Measure | Time frame |
|---|---|
| To compare time to stable signal of heart rate (HR) estimation between two pulse oximeters (PO). | Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours. |
| To compare overall accuracy of heart rate (HR) estimation between two pulse oximeters (PO) as compared to gold standard electrocardiography (ECG). | Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours. |
Secondary
| Measure | Time frame |
|---|---|
| To evaluate instances of false bradycardia. | Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours. |
| To evaluate instances of false bradycardia requiring emergency medical treatment. | Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours. |
| To evaluate variability in PR between the two PO sensors | Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours. |
| To compare the number of repositions and replacements required for the two PO sensors during monitoring. | Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours. |