Standalone Observational Study Assessing the Performance of an AI/ML Tech-based SaMD on Chest LDCT Images (REALITY)

Multinational, Multicenter, Retrospective Study to Evaluate an AI/ML Technology-Based End-to-End CADe/CADx SaMD, Which Allows Detection, Localization and Characterization of Pulmonary Nodules (REALITY)

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT06576232
Acronym
REALITY
Enrollment
1147
Registered
2024-08-28
Start date
2022-09-21
Completion date
2024-08-21
Last updated
2024-08-28

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

High Risk Cancer

Brief summary

This is a Multinational, Multicenter, retrospective study for the evaluation of the standalone efficacy and safety of an Artificial Intelligence/Machine Learning (AI/ML) technology-based end-to-end Computer assisted Detection/Computer Assisted Diagnosis (CADe/CADx) Software as a Medical Device (SaMD) developed to detect, localize and characterize malignant, and suspicious for lung cancer nodules on Low Dose Computed Tomography (LDCT) scans taken as part of a Lung Cancer Screening (LCS) program. LDCT Digital Imaging and Communications in Medicine (DICOM) images of patients who underwent lung cancer screening were selected and included into the study. Selected scans will then be analyzed by the CADe/CADx SaMD and compared to radiologist generated reference standards including lesions localization and lesion cancer diagnosis. Figures of merit at patient level and lesion level detection and diagnostic efficacy will be calculated as well as sub-class analysis to ensure algorithm performance generalizability.

Interventions

End-to-end processing of chest LDCT DICOM images by an AI/ML tech-based SaMD to detect, localize, and characterize (assign a malignancy score) each detected pulmonary nodule. The output of the device is a DICOM File (Median LCS result report) summarizing results per patient.

Sponsors

Median Technologies
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
RETROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
50 Years to 80 Years
Healthy volunteers
Yes

Inclusion criteria

* ≥50-80 Years of age; * Current or ex-smoker (\>=20 pack years); * Patient screened and surveilled for lung cancer screening following lung cancer screening guidelines (equivalent to United States Preventive Services Task Force (USPSTF) 2021 Criteria); * Received LDCT due to inclusion in high-risk category for lung cancer.

Exclusion criteria

* Prior lung resection; * Pacemaker or other indwelling metallic medical devices in the thorax that interfere with CT acquisition; * Patients/images used during AI model development; * Patients with only hilar and/or mediastinal cancer(s); * Patients with only ground glass cancer(s); * Patients with nodules, solid or part-solid \>30mm (masses); * Patients that are not accompanied with the required clinical information; * Patients with imaging with any of the following: missing slices, slice thickness \>3mm; * Partial cover of the lung.

Design outcomes

Primary

MeasureTime frameDescription
AUROC (Area under ROC curve) at patient level12 monthsAUROC that measures Median LCS performance at patient level is strictly superior to 0.8. Support for Primary Endpoint: Derived from the patient level AUROC at the product fixed operating point : Sensitivity, Specificity, PPV, NPV.

Secondary

MeasureTime frameDescription
Specificity > 70% when Sensitivity=70%12 months
AUC of LROC > 0.7512 monthsIn contrast to the receiver operating characteristic (ROC) assessment paradigm, localization ROC (LROC) analysis provides a means to jointly assess the accuracy of localization and detection in an observational study.
Detection sensitivity>0.8 with average FP rate per scan<112 months
ICC>0.8 for average diameter12 monthsIntraclass Correlation Coefficient (ICC), is a descriptive statistic that can be used when quantitative measurements are made on units that are organized into groups. It describes how strongly units in the same group resemble each other.
Sensitivity > 70% when Specificity=70%12 months
ICC>0.8 for short axis diameter12 months
ICC>0.75 for Volume12 months
DICE Coefficient >0.712 months
ICC>0.8 for long axis diameter12 months

Countries

Spain, United States

Outcome results

None listed