Perimetric Testing With Virtual Reality

A Pilot Study on the Repeatability of a Novel Perimetric Test Administered Via a Virtual Reality Headset.

Status
Recruiting
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT06549010
Enrollment
60
Registered
2024-08-12
Start date
2024-07-03
Completion date
2024-10-31
Last updated
2024-08-12

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Glaucoma

Keywords

Glaucoma, Perimetric test, virtual reality

Brief summary

Perimetry is an essential component in the diagnosis and monitoring of glaucoma. Since the advent of Standard Automated Perimetry (SAP), one of the clinical standards has been the Humphrey Field Analyzer (HFA, Carl Zeiss AG, Oberkochen, Germany). Visual field (VF) testing provided by the HFA is standard-of-care in glaucoma clinics and other ophthalmology and optometry practices. However, the HFA is a large device that does not allow for examination outside the clinic and can be uncomfortable for patients with limited mobility. This study aims to evaluate a novel head-mounted perimeter against the HFA as an alternative method of VF testing. This new perimeter uses a Virtual Reality (VR) headset and a gamified version of VF testing to assess the visual function of healthy eyes and patients. The results of this study will potentially serve as pilot data for the design of a larger study that involves the full assessment of the VR headset and its VF test, based on various testing strategies.

Detailed description

Procedures: 1. Ophthalmic screening: (a) Visual acuity, slit lamp exam, rebound tonometry (all subjects) 2. OCT examination of the study eye with the Topcon Maestro2 Optical Coherence Tomography (OCT): (a) Standard Retinal Nerve Fiber Layer (RNFL), Optic Disc and Ganglion Cell Complex (GCC) scan patterns (i) Fundus photo is automatically obtained. (b) Abnormalities consistent with ophthalmic disease will be an exclusion for the healthy group of subjects. Reasonable and common artifacts that do not significantly impair the scan report will be accepted. (c) OCT abnormalities consistent with glaucoma are expected on the 30 glaucoma subjects. 3. Once eligible, each patient will perform four (4) visual field tests: (a) once with the HFA Swedish Interactive Threshold Algorithm (SITA) Standard 24-2 pattern (i) Reliability criteria: False Positives no higher than 15% (b) three tests with the VR headset: (i) two tests with ZEST strategy (ii) one test with gamified ZEST strategy (c) Counter-balancing will be used to determine the order of (a) and (b). In particular, a study participant will either start with the HFA visual field test and then proceed to the three tests with the VR headset; or, start with the three VR tests and finish with the HFA visual field test. The sequence of the visual field tests for each study ID will be provided by the sponsor.

Interventions

DIAGNOSTIC_TESTPerimetric test

This study aims to evaluate a novel head-mounted perimeter against the HFA as an alternative method of VF testing. This new perimeter uses a Virtual Reality (VR) headset and a gamified version of VF testing to assess the visual function of healthy eyes and patients.

Sponsors

Illinois College of Optometry
CollaboratorOTHER
Envision Health Technologies, Inc.
Lead SponsorINDUSTRY

Study design

Allocation
RANDOMIZED
Intervention model
FACTORIAL
Primary purpose
DIAGNOSTIC
Masking
NONE

Intervention model description

30 healthy subjects 30 glaucoma subjects

Eligibility

Sex/Gender
ALL
Age
25 Years to 70 Years
Healthy volunteers
Yes

Inclusion criteria

Inclusion /

Exclusion criteria

1. Age range: 25-70 years 2. For normal subjects: 1. Patient is free of any current or prior ophthalmic pathologies that would potentially cause visual field defects: e.g. retina pathology, glaucoma, cataracts, keratoconus 2. No ocular surgery other than cataract surgery within 6 months prior to the study visit 3. For glaucoma subjects: 1. Patient has a diagnosis of primary open-angle glaucoma 2. Patient is free of any other current or prior ophthalmic pathologies that would potentially cause visual field defects. 3. No other ocular surgery other than glaucoma or cataract surgery within 6 months prior to the study visit 4. No IOP lower than 10 mm/Hg or higher than 30 mm/Hg 4. Refractive Error: 1. Sphere = up to -6.00 D to and up +3.00 2. Cylinder = up to -3.00 is common 5. No past refractive eye surgery 6. No Multifocal or phakic (ICLs) IOL; no multifocal CLs 7. One eligible eye 8. Ability to sit for standard VF bowl exam. Ability to wear VR perimetry headset device.

Design outcomes

Primary

MeasureTime frameDescription
Visual field perimetry light sensitivity1 dayLocalized differential light sensitivity threshold measured in decibels (dB)
Visual field perimetry foveal sensitivity1 dayFoveal threshold in decibels (dB)
Visual field perimetry test duration1 dayTest duration measured in minutes

Secondary

MeasureTime frameDescription
Optical Coherence Tomography (OCT)1 dayOptical Coherence Tomography (OCT) thickness values may be used, such as Retinal nerve fiber layer (RNFL) and ganglion cell complex (GCC) thickness, measured in microns.