Non-invasive Point-Of-Care Hemoglobin Testing In An Obstetric Population

Non-invasive Point-Of-Care Hemoglobin Testing In An Obstetric Population

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT06521840
Enrollment
124
Registered
2024-07-26
Start date
2024-12-01
Completion date
2025-09-01
Last updated
2026-03-09

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Pregnancy Anemia

Brief summary

Anemia is common during pregnancy due to increased metabolic rate and normal physiologic changes associated with pregnancy. Anemia during pregnancy has been associated with increased of adverse outcomes during delivery and the postpartum period. Currently, it is recommended to screen for anemia at three pre-specified points during routine prenatal care, typically 3 months apart. This screening method may miss patients who develop anemia between these intervals and lead to a delay in diagnosis and treatment. MasimoTM has developed a non-invasive device that can detect anemia without a blood draw. This device has not been extensively studied in pregnant patients. The aim of this study is to determine the agreement in hemoglobin readings between the non-invasive device and the standard of care blood draws in pregnant patients.

Detailed description

This is a prospective observational study of a non-invasive hemoglobin measurement device in a pregnant patient population. The device is FDA approved for use in adults, children, and neonates. The current standard of care for anemia screening during pregnancy involves performing invasive venipuncture at two separate visits throughout the course of the pregnancy. The purpose of this research study is to compare the device performance with the gold standard hemoglobin measurement (venipuncture) when patients receive routine hemoglobin measurement during the course of their prenatal care. The research team will approach pregnant patients who are presenting for prenatal care at Beth Israel Deaconess Medical Center, including the Labor and Delivery Triage Unit. The research team also will approach patients during an antepartum or postpartum admission. The research team will use the Masimo Rad-67 handheld co-oximeter which has a reusable finger-tip sensor that enables spot-check monitoring measurements of both oxygen saturation and non-invasive hemoglobin measurement. Study personnel will receive training on proper device use and cleaning by the local Masimo support team. At the time of a blood draw, the research team will place the pulse oximeter on the participant's finger. The devices requires 1-3 minutes for an accurate measurement. After the measurement is complete, the participant will be asked 2 questions on their satisfaction and comfortability with the device and 1 question regarding their skin tone using a validated skin tone scale.

Interventions

DEVICEMasimo RAD 67

This is an observational study of a non-invasive hemoglobin measurement device in a pregnant patient population. The current standard of care for anemia screening during pregnancy involves performing invasive venipuncture at two separate visits throughout the course of the pregnancy. The purpose of this research study is to compare the device performance with the gold standard hemoglobin measurement (venipuncture). The research team will use the Masimo Rad-67 handheld co-oximeter which has a reusable finger-tip sensor that enables spot-check monitoring measurements of both oxygen saturation and non-invasive hemoglobin measurement. At the time of a blood draw, the research team will place the pulse oximeter on the participant's finger. The devices requires 1-3 minutes for an accurate measurement. After the measurement is complete, the participant will be asked 2 questions on their satisfaction and comfortability with the device and 1 question regarding their skin tone.

Sponsors

Beth Israel Deaconess Medical Center
Lead SponsorOTHER

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
FEMALE
Age
18 Years to 60 Years
Healthy volunteers
No

Inclusion criteria

* Pregnant or postpartum; * Age ≥18 years old; * Undergoing phlebotomy for clinical purpose

Exclusion criteria

\- None

Design outcomes

Primary

MeasureTime frameDescription
Hemoglobin (g /dL) Accuracy of Masimo RAD 67At time of phlebotomyPerformance will be determined by calculating mean difference and repeatability analysis through the comparison of the hemoglobin values obtained from Masimo RAD67 to the hemoglobin values obtained from a reference phlebotomy sample collected at the same time.

Secondary

MeasureTime frameDescription
Hemoglobin (g /dL) Accuracy of Masimo RAD 67 stratified by skin toneAt time of phlebotomyPerformance will be determined by calculating mean difference and repeatability analysis through the comparison of the hemoglobin values obtained from Masimo RAD67 to the hemoglobin values obtained from a reference phlebotomy sample collected at the same time stratified by skin color using a validated scale. Monk, Ellis. "The monk skin tone scale." (2019))
Hemoglobin (g /dL) Accuracy of Masimo RAD 67 stratified by gestational ageAt time of phlebotomyPerformance will be determined by calculating mean difference and repeatability analysis through the comparison of the hemoglobin values obtained from Masimo RAD67 to the hemoglobin values obtained from a reference phlebotomy sample collected at the same time stratified by gestational age.

Countries

United States

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Mar 10, 2026