Vital Sign Evaluation
Conditions
Brief summary
Evaluating the Accuracy of a Web-Based Application - Vital Sign Measurement Platform. This single-arm cohort, open-label study is designed to evaluate the accuracy of a web-based application designed for measurements of vital signs including respiratory rate (RR). The purpose of the study is to conduct accuracy validation when compared to a FDA-cleared/approved vital sign monitoring device (reference device).
Interventions
Informed Vital Core App will be initiated by the study staff to take a measurement, which takes 60 seconds to calculate and report a breathing rate value.
Sponsors
Study design
Eligibility
Inclusion criteria
* Participants must have the ability to understand and provide written informed consent or have legally authorized representative consent to participate. * Participant must be ≥22 years of age * Participant must be willing and able to comply with study procedures and duration * Participants or legally authorized representative must be able to read or write in English
Exclusion criteria
* Participants who refuse or are unable to sign an informed written consent for study * Participants evaluated by the Investigator and Clinical Staff and found to be medically unsuitable or have self-reported health conditions that are currently unstable as identified in the Health Assessment Form and Health Screening * Compromised circulation, injury, or physical malformation the Region of Interest (ROI) which would limit the ability to test ROI needed for the study. (Note: Certain malformations may still allow participants to participate if the condition is noted and would not affect the particular areas utilized.) * Participants with severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, or other medical sensors (self-reported) * Other known health condition, should be considered upon disclosure in health assessment form
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Respiratory Rate Accuracy | 60 seconds | The endpoint of interest is accuracy as measured by the Accuracy root-mean-square (Arms) difference between the Device Under Test (DUT) and the Reference (Ref) Respiratory Rate as counted using the End Tidal Carbon Dioxide scored waveform by counting the respiratory peaks per minute for all stable respiratory periods. |
Countries
United States
Contacts
Element Materials Technology
Baseline characteristics
| Characteristic | — |
|---|---|
| Age, Categorical <=18 years | 0 Participants |
| Age, Categorical >=65 years | 31 Participants |
| Age, Categorical Between 18 and 65 years | 34 Participants |
| Age, Continuous | 49 years |
| Ethnicity (NIH/OMB) Hispanic or Latino | 15 Participants |
| Ethnicity (NIH/OMB) Not Hispanic or Latino | 50 Participants |
| Ethnicity (NIH/OMB) Unknown or Not Reported | 0 Participants |
| Race (NIH/OMB) American Indian or Alaska Native | 2 Participants |
| Race (NIH/OMB) Asian | 5 Participants |
| Race (NIH/OMB) Black or African American | 11 Participants |
| Race (NIH/OMB) More than one race | 8 Participants |
| Race (NIH/OMB) Native Hawaiian or Other Pacific Islander | 1 Participants |
| Race (NIH/OMB) Unknown or Not Reported | 0 Participants |
| Race (NIH/OMB) White | 38 Participants |
| Region of Enrollment United States | 65 participants |
| Sex: Female, Male Female | 32 Participants |
| Sex: Female, Male Male | 33 Participants |
Adverse events
| Event type | EG000 affected / at risk |
|---|---|
| deaths Total, all-cause mortality | 0 / 65 |
| other Total, other adverse events | 0 / 65 |
| serious Total, serious adverse events | 0 / 65 |