Skip to content

Evaluation of Bilirubin Measurements in Newborns From Smartphone Digital Images in a Population in Botswana

Evaluation of Bilirubin Measurements in Newborns From Smartphone Digital Images in a Population in Botswana

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT06372093
Enrollment
172
Registered
2024-04-17
Start date
2023-03-16
Completion date
2024-02-26
Last updated
2024-04-17

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Jaundice, Neonatal

Brief summary

The general objective of this study is to evaluate the accuracy of a novel smartphone application for jaundice screening (Picterus Jaundice Pro) in a population with high melanin content in the skin. The specific objectives for this study are: i. To assess the correlation between bilirubin level measurements obtained by Picterus Jaundice Pro with Total Serum Bilirubin (TSB), and TcB, in newborns with high melanin content in the skin. ii. To determine the accuracy of Picterus Jaundice Pro in newborns with high melanin content in the skin.

Detailed description

The goal for this project is to demonstrate that the Picterus Jaundice Pro screening method can be used to identify children with severe neonatal jaundice independent of skin color. Thus, providing a cheaper, patient-friendly, and more readily available method of neonatal jaundice detection. The long-term goal is that with the implementation of the app, the identification of newborns with severe jaundice will be improved and, thus, referred at an earlier stage for diagnostics and treatment. The study will be a cross-sectional study with quantitative methods of data collection. The study population will comprise 150 newborns from the Neonatal Unit and the maternity ward at Princess Marina Hospital. Parents to newborns with and without signs of jaundice will be asked to participate. Following informed consent, background data such as birth weight, age on examination, gestational age, and type of feeding will be obtained. Gestational age will be based on ultrasound determination and the last normal menstruation period (LNMP). The skin type of the infant will be classified according to the newborn Neomar's scale score. Transcutaneous bilirubin measurements will be performed over the sternum of the infant. A Dräger Jaundice Meter JM-105 will be used in this study. Skin reflectance will be measured using a portable Konica Minolta spectrophotometer CM-700d. A validated smartphone with Picterus Jaundice Pro will be used to collect digital images of skin of the infant chest together with the Picterus calibration card. After all the images are obtained, a unique ID will be displayed on the smartphone. This ID will be recorded on the case report forms and later used to pair clinical data and digital images. A blood sample to determine TSB will be obtained within 60 minutes of obtaining the images and processed at the Department of clinical biochemistry in the hospital laboratory. After completion of the study the bilirubin measurement from Picterus Jaundice Pro will be determined and compared to the TSB and TCB measurements using the Pearson correlation coefficient. Sub-analysis for the different skin colors will be performed. Systematic over- or under-estimation of bilirubin levels will be evaluated using Bland-Altman plots. Sensitivity and specificity analysis will be calculated for different cut-off values (ROC analysis).

Interventions

Picterus Jaundice Pro is a smartphone based screening tool for neonatal jaundice that takes images of the skin of the newborn chest where the Picterus Calibration Card is placed.

Sponsors

University of Copenhagen
CollaboratorOTHER
University of Botswana
CollaboratorOTHER
Picterus AS
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
SCREENING
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
1 Days to 14 Days
Healthy volunteers
Yes

Inclusion criteria

* Gestational age \> 37 weeks * Birth weight \> 2000 g and \< 4500 g * Age 1-14 days * Infants requiring a blood sample for clinically suspected jaundice/screening

Exclusion criteria

* Infants transferred to the pediatric ward for advanced treatment. * Infants with a skin rash or other disease that affects the skin where measurements are performed. * Infants that receive or have received phototherapy in the last 24 hours. * Infants with an inborn disease.

Design outcomes

Primary

MeasureTime frameDescription
Enable high qualitative estimation of bilirubin levels in the blood of newborns in a population in Botswana using Picterus JP.1 yearEvaluate of smartphone-based screening tool (Picterus JP) for neonatal jaundice in newborns with high melanin content.

Secondary

MeasureTime frameDescription
Correlation of bilirubin levels obtained by Picterus JP with Tsb and TcB1-2 hoursCorrelate estimates of bilirubin levels obtained by Picterus JP with TSB and TcB in newborns with high melanin content in the skin.