Brief summary

The aim of the study is to validate the Mean Arterial Pressure (MAP) parameter of Masimo INVSENSOR00073 against the gold standard blood pressure measurements from the arterial line while subject undergo a blood pressure variation procedure.

Interventions

DEVICEMasimo INVSENSOR00073

Noninvasive blood pressure monitor

Sponsors

Masimo Corporation
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 40 Years
Healthy volunteers
Yes

Inclusion criteria

Influenced by study design: * Subject is 18 to 40 years of age. * Subject weighs a minimum of 110 lbs. * Baseline heart rate ≥ 45 bpm and ≤ 90 bpm. * Baseline blood pressure: * Systolic blood pressure ≤ 135 mmHg and ≥ 100 mmHg. * Diastolic blood pressure ≤ 95 mmHg and ≥ 55 mmHg. * Valid systolic blood pressure (SBP) auscultatory measurements for lateral difference is ≤ 15 mmHg. * Valid diastolic blood pressure (DBP) auscultatory measurements for lateral difference is ≤ 10 mmHg. * Hemoglobin value ≥ 11 g/dL. * CO value ≤ 3.0% FCOHb. * Subject is comfortable to read and communicate in English\*. \* This is to ensure the subject can provide informed consent (as study materials are currently available in English only) and can comply with study procedures.

Exclusion criteria

Influenced by device warning labels: * Subject is pregnant or breastfeeding. * Subject is experiencing dysrhythmia or arrhythmia. Influenced by study design/environment: * Subject is concurrently participating in another research study. * Subjects not suitable for the investigation at the discretion of the clinical team including but not limited to the items below. * Subjects who do not pass the health assessment for safe participation in the study procedures. * Inability to insert or difficulty with inserting an intravenous line in the subject's hand or arm and/or an arterial line in the radial artery of the subject's wrist.

Design outcomes

Primary

MeasureTime frameDescription
Mean Arterial Pressure (MAP) Accuracy and Precision of Masimo INVSENSOR000731-5 hours.Accuracy and precision of mean arterial pressure (MAP) measured by Masimo INVSENSOR00073 in comparison to radial arterial line.

Countries

United States

Participant flow

Participants by arm

ArmCount
Test Subject
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Masimo INVSENSOR00073 sensor. Masimo INVSENSOR00073: Noninvasive blood pressure monitor
25
Total25

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyAdverse Event4
Overall StudyDisqualified72

Baseline characteristics

CharacteristicTest Subject
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
25 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
5 Participants
Race (NIH/OMB)
Black or African American
5 Participants
Race (NIH/OMB)
More than one race
2 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
6 Participants
Race (NIH/OMB)
White
7 Participants
Region of Enrollment
United States
25 participants
Sex: Female, Male
Female
5 Participants
Sex: Female, Male
Male
20 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 101
other
Total, other adverse events
10 / 101
serious
Total, serious adverse events
0 / 101

Outcome results

Primary

Mean Arterial Pressure (MAP) Accuracy and Precision of Masimo INVSENSOR00073

Accuracy and precision of mean arterial pressure (MAP) measured by Masimo INVSENSOR00073 in comparison to radial arterial line.

Time frame: 1-5 hours.

Population: 9 subjects not analyzed due to incorrect software version

ArmMeasureValue (MEAN)Dispersion
Masimo INVSENSOR00073Mean Arterial Pressure (MAP) Accuracy and Precision of Masimo INVSENSOR000731.38 Millimeters of mercury (mmHg)Standard Deviation 4.98

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026