Otosclerosis
Conditions
Keywords
stapes, titanium, computed tomography
Brief summary
Actual CT scanners overestimate stapes piston size, and do not represent a valuable technique for their follow-up, especially in case of complication. Ultra-high resolution has not yet been evaluated in this setting.
Detailed description
Modern temporal bone imaging systems, using 0.5 mm slice thickness, are reputed not yet able to depict the stapes piston position with a sufficient degree of accuracy, as high-resolution CT measurements consistently overestimated intravestibular piston dimensions (width by 44%, length by 6.25%) and vestibular intrusion (by 39%) because of metallic artifacts. . The exact limits of intravestibular depth are also unclear. The safe intravestibular prosthesis depth varies widely between studies and has been reported to range from 0.2 to 2.5 mm, so 39% may represent up to 0.9 mm of misestimation. Ultra-high resolution CT (UHR-CT) has improved spatial resolution to 0.12 mm by using ultra-small detector elements (0.25 x 0.25 mm), allowing better visualization of temporal bone structures compared to high-resolution temporal bone CT (HR-CT). However, this technique has not been described for stapes prosthesis measurement.
Interventions
stapes removal and replacement by a prosthesis
imaging study
Sponsors
Study design
Eligibility
Inclusion criteria
\- stapes prosthesis Ultra-high resolution CT available
Exclusion criteria
no materiovigilance report available
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| length | february 2020 - novembre 2023 | length of the prosthesis on CT |
Countries
France