Echocardiography With Bubble Test for Tip Location of Central Venous Catheters

Transthoracic Echocardiography With Bubble Test for Tip Location of Central Venous Catheters: a Push to Bubble Time Analysis

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT06243445
Enrollment
22
Registered
2024-02-06
Start date
2024-02-01
Completion date
2025-11-27
Last updated
2025-12-01

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Cardiac Surgery Patient

Keywords

central venous catheter tip location, transthoracic echocardiography, bubble test

Brief summary

The goal of this prospective observational study is to test the accuracy of the transthoracic echocardiography with bubble test to verify the location of the tip of the central venous catheter in patients undergoing cardiac surgery procedures. The main questions it aims to answer are: 1. Is there a difference in time elapsed between the injection of a saline solution (echogenic contrast) through a central venous catheter and the visualization of micro bubbles in right atrium (push to bubbles time) with transthoracic echocardiography, when the catheter tip is placed in four different positions identified by transesophageal echocardiography? 2. Is there a difference in time for the right atrium to be completely filled by bubbles after saline injection in a central venous catheter with the tip placed in four different positions identified by transesophageal echocardiography? 3. In what percentage of patients the acoustic window is good enough to evaluate push to bubbles time? 4. Is there a difference in push to bubbles time between different echocardiographic projections? 5. Is the bubbles flow laminar or turbulent in the four different positions? 6. Is there a relation between push to bubble time and heart rate, blood pressure, pulmonary artery pressure, central venous pressure and cardiac output measured through a pulmonary artery catheter? Participants will require simultaneously a central venous catheter and an intraoperative transesophageal echocardiography.

Detailed description

The placement of central vascular access devices (VADs), with cervical-thoracic or femoral insertion (such as CICC, Central Inserted Central Catheter, and FICC, Femoral Inserted Central Catheter), as well as peripheral insertion (PICC, Peripherally Inserted Central Catheter), is a common procedure for patients admitted to the Intensive Care Unit and for patients undergoing major surgery or prolonged infusion therapy. To reduce the incidence of complications related to improper positioning of the central venous catheter, the tip should be positioned in the lower part of the superior vena cava or in the upper part of the right atrium near the cavo-atrial junction, identified by the crista terminalis. Among the methods used to verify the correct tip placement, the intracavitary ECG method is strongly recommended by various international guidelines, and its use has been widely validated in terms of safety, efficacy, and accuracy. However, the intracavitary ECG method, based on the principle that the appearance of a high P-wave indicates the proximity of the tip to the cavo-atrial junction, has limitations when it comes to patients with non-sinus rhythm. Recent studies are validating the use of modified intracavitary ECG in patients with atrial fibrillation, but in other cases where a P-wave is not visible, the method is not applicable. Various methods for ultrasound-guided placement and confirmation of the final tip position have been developed, also because the chest X-ray performed at the patient's bedside has proven to be a less accurate method due to the lack of definite findings of the cavo-atrial junction. Transesophageal echocardiography (TEE), while being the method of choice as it allows direct visualization of the cavo-atrial junction and, therefore, the real position of both the guidewire and the tip of the catheter at the end of placement, remains an invasive method that is difficult to perform outside of cardiac operating rooms. Transthoracic echocardiography may allow for the direct visualization of the tip of the VAD inside the heart chambers or indirectly assessing the tip's position by injecting echogenic contrast. Echogenic contrast refers to a rapid bolus of 5-20 ml of 0.9% saline solution, with or without the addition of 1 ml of air, vigorously shaken to create micro bubbles that can be easily visualized during the echocardiographic exam. The bolus is injected into the VAD once it has been positioned, and the micro bubbles are visualized in the right atrium, using transthoracic echocardiography. In nearly all studies, the VAD is considered in the correct position if the micro bubbles appear in the atrium within 2 seconds of injection (push to bubble time); otherwise, it is not considered correctly positioned if the bubbles appear after 2 seconds, but in one study, the adequate time was considered to be less than 500 ms. In almost all echocardiography studies, however, the reference method used to evaluate and confirm the VAD tip position has been chest radiography, which has been proven to be a less accurate method; moreover, literature shows significant variations in the indicative times for the visualization of micro bubbles that suggest incorrect position (range 0.5-2 seconds). In this context, it could be useful to determine accurately if different push to bubble times are indicative for different positions of the VAD tip. For this purpose, the echogenic contrast (10 ml of 0,9% saline solution with the addition of 1 ml of air vigorously shaken to create micro bubbles) will be injected in the distal lumen of a VAD, whose tip is positioned in four different points, identified with using transesophageal echocardiography and the time from the contrast injection to the visualization of the bubbles in right atrium will be recorded using transthoracic echocardiography. The four positions in which the catheter tip will be placed are: right atrium, cavo-atrial junction (CAJ), at 4 centimeters proximal to the CAJ in superior vena cava and at 8 centimeters to the CAJ. The catheter tip positions in right atrium and at CAJ will be directly evaluated using 2D and 3D transesophageal echocardiography, the other positions will be obtained moving the catheter back 4 and 8 centimeters from the CAJ. The transthoracic echocardiography will be carried out always by the same cardiologist trained in echocardiography unaware of the purpose of the study. The time measurements will be carried out always by the same cardiac anaesthesiologist, trained in echocardiography, unaware of the different catheter tip positions.

Interventions

DIAGNOSTIC_TESTTransthoracic echocardiography for central venous catheter tip location

Measure of time between injection of echographic contrast in the distal lumen of a central venous catheter, inserted via supra/infra clavicular right veins, and visualization, using transthoracic echocardiography, of micro bubbles in right atrium with the catheter tip positioned under transesophageal ultrasound control in four different positions: right atrium, cavo-atrial junction (CAJ), at 4 centimeters proximal to CAJ, in superior vena cava and at 8 centimeters proximal to CAJ

Sponsors

Fondazione Policlinico Universitario Agostino Gemelli IRCCS
Lead SponsorOTHER

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* age \> 18 years * written informed consent from the patient * intra-operative need for a central venous catheterization via the supra/infra clavicular right veins (centrally inserted central catheter, CICC) * indication to intra-operative Trans Esophageal Ultrasound (TEE) based on American Society of Anesthesiologists (ASA)

Exclusion criteria

* contraindications to TEE based on American Society of Anesthesiologists (ASA) recommendations (esophageal or gastric diseases or previous surgery) * emergency procedures * central VAD already placed * need for a CICC via supra/infra clavicular left veins * need for a femorally inserted central catheter (FICC) * need for a peripherally inserted central catheter (PICC) * tricuspid valve regurgitation * hemodynamic instability/ presence of an intra-aortic ballon pump (IABP) * absence of septal atrial defect

Design outcomes

Primary

MeasureTime frameDescription
Difference in push to bubbles time recorded with the catheter tip placed in four different positions by three different ultrasound views; measure in position 1: right atrium, using subcostal 4-chamber viewImmediately after the induction of anesthesia, placement of transesophageal probe and insertion of a central venous catheter. The video recorded will be analyzed immediately after the end of the procedureAfter the insertion of a central venous catheter in the supraclavicular right area by ultrasound guidance and direct Seldinger technique, under transesophageal ultrasound control, the catheter tip will be placed in right atrium, 2 cm below the CAJ; the echographic contrast will be injected through the proximal end of the catheter connected, via an Arrow-Johans adapter (Teleflex) and a dedicated ECG cable (Teleflex), to the transthoracic ultrasound machine. The video with the micro bubbles injected will be recorded by a cardiologist expert in echocardiography, unaware of the purpose of the study, using a subcostal 4-chamber view. Later a cardiac anesthesiologist, unaware of the purpose of the position of the catheter, will measure the time elapsed between the first positive wave created by the contrast injection on the intracavitary ECG recorded by the transthoracic ultrasound machine and the time in which the bubbles will appear in the right atrium.
Difference in push to bubbles time recorded with the catheter tip placed in four different positions by three different ultrasound views; measure in position 1: right atrium, using subcostal bicaval viewImmediately after the video recorded with the catheter tip placed in right atrium using subcostal 4-chamber viewWith the procedure described above, the push to bubble time will be measured with the catheter tip placed in right atrium using subcostal bicaval view
Difference in push to bubbles time recorded with the catheter tip placed in four different positions by three different ultrasound views; measure in position 1: right atrium, using apical 4-chamber viewImmediately after the video recorded with the catheter tip placed in right atrium using subcostal bicaval viewWith the procedure described above, the push to bubble time will be measured with the catheter tip placed in right atrium using apical 4-chamber view
Difference in push to bubbles time recorded with the catheter tip placed in four different positions by three different ultrasound views; measure in position 2: cavo-atrial junction (CAJ), using subcostal 4-chamber viewImmediately after the video recorded with the catheter tip placed in right atrium using apical 4-chamber viewWith the procedure described above, the push to bubble time will be measured with the catheter tip placed at the CAJ, using subcostal 4-chamber view
Difference in push to bubbles time recorded with the catheter tip placed in four different positions by three different ultrasound views; measure in position 2: cavo-atrial junction (CAJ), using subcostal bicaval viewImmediately after the video recorded with the catheter tip placed at the CAJ using subcostal 4-chamber viewWith the procedure described above, the push to bubble time will be measured with the catheter tip placed at the CAJ, using subcostal bicaval view
Difference in push to bubbles time recorded with the catheter tip placed in four different positions by three different ultrasound views; measure in position 2: cavo-atrial junction (CAJ), using apical 4-chamber viewImmediately after the video recorded with the catheter tip placed at the CAJ using subcostal bicaval viewWith the procedure described above, the push to bubble time will be measured with the catheter tip placed at the CAJ, using apical 4-chamber view
Difference in push to bubbles time recorded with the catheter tip placed in four different positions by three different ultrasound views; measure in position 3: at 4 centimeters proximal to CAJ in superior vena cava, using subcostal 4-chamber viewImmediately after the video recorded with the catheter tip placed at the CAJ using apical 4-chamber viewWith the procedure described above, the push to bubble time will be measured with the catheter tip placed at , 4 centimeters proximal to CAJ in superior vena cava, using subcostal 4-chamber view
Difference in push to bubbles time recorded with the catheter tip placed in four different positions by three different ultrasound views; measure in position 3: at 4 centimeters proximal to CAJ in superior vena cava, using subcostal bicaval viewImmediately after the video recorded with the catheter tip at 4 centimeters proximal to CAJ, in superior vena cava, using subcostal 4-chamber viewWith the procedure described above, the push to bubble time will be measured with the catheter tip placed at 4 centimeters proximal to CAJ in superior vena cava, using subcostal bicaval view
Difference in push to bubbles time recorded with the catheter tip placed in four different positions by three different ultrasound views; measure in position 3: at 4 centimeters proximal to CAJ in superior vena cava, using apical 4-chamber viewImmediately after the video recorded with the catheter tip at 4 centimeters proximal to CAJ, in superior vena cava, using subcostal bicaval viewWith the procedure described above, the push to bubble time will be measured with the catheter tip placed at 4 centimeters proximal to CAJ in superior vena cava, using apical 4-chamber view
Difference in push to bubbles time recorded with the catheter tip placed in four different positions by three different ultrasound views; measure in position 4: at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewImmediately after the video recorded with the catheter tip at 4 centimeters proximal to the CAJ, in superior vena cava, using apical 4-chamber viewWith the procedure described above, the push to bubble time will be measured with the catheter tip placed at 8 centimeters proximal to the CAJ, using subcostal 4-chamber view
Difference in push to bubbles time recorded with the catheter tip placed in four different positions by three different ultrasound views; measure in position 4: at 8 centimeters proximal to CAJ, using subcostal bicaval viewImmediately after the video recorded with the catheter tip at 8 centimeters, using subcostal 4-chamber viewWith the procedure described above, the push to bubble time will be measured with the catheter tip placed at 8 centimeters proximal to the CAJ, using subcostal bicaval view
Difference in push to bubbles time recorded with the catheter tip placed in four different positions by three different ultrasound views; measure in position 4: at 8 centimeters proximal to CAJ, using apical 4-chamber viewImmediately after the video recorded with the catheter tip at 8 centimeters, using subcostal bicaval viewWith the procedure described above, the push to bubble time will be measured with the catheter tip placed at 8 centimeters proximal to the CAJ, using apical 4-chamber view

Secondary

MeasureTime frameDescription
Percentage in which push to bubbles time is evaluable using subcostal 4-chamber viewImmediately after every patient recruited will be evaluated with ultrasound in subcostal 4-chamber viewTo measure the percentage of the patients in which acoustic window is good enough to evaluate push to bubbles time using subcostal 4-chamber view.
Percentage in which push to bubbles time is evaluable using subcostal bicaval viewImmediately after every patient recruited will be evaluated with ultrasound in subcostal bicaval viewTo measure the percentage of the patients in which acoustic window is good enough to evaluate push to bubbles time using subcostal bicaval view.
Percentage in which push to bubbles time is evaluable using apical 4-chamber viewImmediately after every patient recruited will be evaluated with ultrasound in apical 4-chamber viewTo measure the percentage of the patients in which acoustic window is good enough to evaluate push to bubbles time using 4-chamber view.
Difference between push to bubbles time averages measured using the three different ultrasound views with the catheter tip in right atriumImmediately after push to bubbles time will be evaluated in every patient recruited using the three different ultrasound views with the catheter tip in right atriumTo determine if there is a significative difference between push to bubbles time averages measured using the three different ultrasound views: subcostal 4-chamber view, subcostal bicaval view and apical 4-chamber view with the catheter tip in right atrium
Difference between push to bubbles time averages measured using the three different ultrasound views with the catheter tip at the CAJImmediately after push to bubbles time will be evaluated in every patient recruited using the three different ultrasound views with the catheter tip at the CAJTo determine if there is a significative difference between push to bubbles time averages measured using the three different ultrasound views: subcostal 4-chamber view, subcostal bicaval view and apical 4-chamber view with the catheter tip at the CAJ
Difference between push to bubbles time averages measured using the three different ultrasound views with the catheter tip at 4 centimeters proximal to CAJ, in superior vena cavaImmediately after push to bubbles time will be evaluated in every patient recruited using the three different ultrasound views with the catheter tip at 4 centimeters proximal to CAJTo determine if there is a significative difference between push to bubbles time averages measured using the three different ultrasound views: subcostal 4-chamber view, subcostal bicaval view and apical 4-chamber view with the catheter tip at 4 centimeters proximal to CAJ, in superior vena cava
Difference between push to bubbles time averages measured using the three different ultrasound views with the catheter tip at 8 centimeters proximal to CAJImmediately after push to bubbles time will be evaluated in every patient recruited using the three different ultrasound views with the catheter tip at 8 centimeters proximal to CAJTo determine if there is a significative difference between push to bubbles time averages measured using the three different ultrasound views: subcostal 4-chamber view, subcostal bicaval view and apical 4-chamber view with the catheter tip at 8 centimeters proximal to CAJ
Evaluation of the quality of bubbles flow when the catheter tip is in the right atrium, using subcostal 4-chamber viewImmediately after Immediately after the evaluation of time for the right atrium to be completely filled by micro bubbles with the catheter tip placed in right atrium using subcostal 4-chamber viewTo evaluate the percentage in which the bubbles flow is laminar or turbulent when the catheter tip is in the right atrium, using subcostal 4-chamber view
Evaluation of the quality of bubbles flow when the catheter tip is in the right atrium, using subcostal bicaval viewImmediately after the evaluation of time for the right atrium to be completely filled by micro bubbles with the catheter tip placed in right atrium using subcostal bicaval viewTo evaluate the percentage in which the bubbles flow is laminar or turbulent when the catheter tip is in the right atrium, using subcostal bicaval view
Evaluation of the quality of bubbles flow when the catheter tip is in the right atrium, using apical 4-chamber viewImmediately after the evaluation of time for the right atrium to be completely filled by micro bubbles with the catheter tip placed in right atrium using apical 4-chamber viewTo evaluate the percentage in which the bubbles flow is laminar or turbulent when the catheter tip is in the right atrium, using apical 4-chamber view
Evaluation of the quality of bubbles flow when the catheter tip is at the CAJ using subcostal 4-chamber viewImmediately after the evaluation of time for the right atrium to be completely filled by micro bubbles with the catheter tip placed in right atrium using subcostal 4-chamber viewTo evaluate the percentage in which the bubbles flow is laminar or turbulent when the catheter tip is at the CAJ, using subcostal 4-chamber view
Evaluation of the quality of bubbles flow when the catheter tip is at the CAJ, using subcostal bicaval viewImmediately after the evaluation of time for the right atrium to be completely filled by micro bubbles with the catheter tip placed at the CAJ using subcostal bicaval viewTo evaluate the percentage in which the bubbles flow is laminar or turbulent when the catheter tip is at the CAJ, using subcostal bicaval view
Evaluation of the quality of bubbles flow when the catheter tip is at 4 centimeters proximal to CAJ, using subcostal 4-chamber viewImmediately after the evaluation of time for the right atrium to be completely filled by micro bubbles with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber viewTo evaluate the percentage in which the bubbles flow is laminar or turbulent when the catheter tip is at 4 centimeters proximal to CAJ, using subcostal 4-chamber view
Evaluation of the quality of bubbles flow when the catheter tip is at 4 centimeters proximal to CAJ, using subcostal bicaval viewImmediately after the evaluation of time for the right atrium to be completely filled by micro bubbles with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal bicaval viewTo evaluate the percentage in which the bubbles flow is laminar or turbulent when the catheter tip is at 4 centimeters proximal to CAJ, using subcostal bicaval view
Evaluation of the quality of bubbles flow when the catheter tip is at 4 centimeters proximal to CAJ, using subcostal apical 4-chamber viewImmediately after the evaluation of time for the right atrium to be completely filled by micro bubbles with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal apical 4-chamber viewTo evaluate the percentage in which the bubbles flow is laminar or turbulent when the catheter tip is at 4 centimeters proximal to CAJ, using subcostal apical 4-chamber view
Evaluation of the quality of bubbles flow when the catheter tip is at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewImmediately after the evaluation of time for the right atrium to be completely filled by micro bubbles with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewTo evaluate the percentage in which the bubbles flow is laminar or turbulent when the catheter tip is at 8 centimeters proximal to CAJ, using subcostal 4-chamber view
Evaluation of the quality of bubbles flow when the catheter tip is at 8 centimeters proximal to CAJ, using subcostal bicaval viewImmediately after the evaluation of time for the right atrium to be completely filled by micro bubbles with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal bicaval viewTo evaluate the percentage in which the bubbles flow is laminar or turbulent when the catheter tip is at 8 centimeters proximal to CAJ, using subcostal bicaval view
Evaluation of the quality of bubbles flow when the catheter tip is at 8 centimeters proximal to CAJ, using apical 4-chamber viewImmediately after the evaluation of time for the right atrium to be completely filled by micro bubbles with the catheter tip placed at 8 centimeters proximal to CAJ, using apical 4-chamber viewTo evaluate the percentage in which the bubbles flow is laminar or turbulent when the catheter tip is at 8 centimeters proximal to CAJ, using apical 4-chamber view
Correlation between heart rate and push to bubbles time measured with the catheter tip placed in right atrium using subcostal 4-chamber viewHeart rate will be measured immediately before the ultrasound video recording with the catheter tip placed in right atrium using subcostal 4-chamber viewTo correlate the heart rate measured immediately before the ultrasound video recording with the catheter tip placed in right atrium using subcostal 4-chamber view
Correlation between heart rate and push to bubbles time measured with the catheter tip placed at CAJ using subcostal 4-chamber viewHeart rate will be measured immediately before the ultrasound video recording with the catheter tip placed at CAJ using subcostal 4-chamber viewTo correlate the heart rate measured immediately before the ultrasound video recording with the catheter tip placed at CAJ using subcostal 4-chamber view
Correlation between heart rate and push to bubbles time measured with the catheter tip placed at 4 centimeters proximal to CAJ, in superior vena cava, using subcostal 4-chamber viewHeart rate wil be measured immediately before the ultrasound video recording with the catheter tip placed at 4 centimeters proximal to CAJ, in superior vena cava using subcostal 4-chamber viewTo correlate the heart rate measured immediately before the ultrasound video recording with the catheter tip placed at 4 centimeters proximal to CAJ, in superior vena cava using subcostal 4-chamber view
Correlation between heart rate and push to bubbles time measured with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewHeart rate will be measured immediately before the ultrasound video recording with the catheter tip placed at at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewTo correlate the heart rate measured immediately before the ultrasound video recording with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber view
Correlation between systolic arterial blood pressure and push to bubbles time measured with the catheter tip placed in right atrium, using subcostal 4-chamber viewSystolic arterial blood pressure will be measured immediately before the ultrasound video recording the contrast injection with the catheter tip placed in right atrium, using subcostal 4-chamber viewTo correlate the systolic arterial blood pressure measured immediately before the ultrasound video recording with the catheter tip placed in right atrium, using subcostal 4-chamber view
Correlation between systolic arterial blood pressure and push to bubbles time measured with the catheter tip placed at CAJ, using subcostal 4-chamber viewSystolic arterial blood arterial pressure will be measured immediately before the ultrasound video recording with the catheter tip placed at CAJ, using subcostal 4-chamber viewTo correlate the systolic arterial blood pressure measured immediately before the ultrasound video recording with the catheter tip placed at CAJ, using subcostal 4-chamber view
Correlation between systolic arterial blood pressure and push to bubbles time measured with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber viewSystolic arterial blood pressure is measured immediately before the ultrasound video recording the contrast injection with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber viewTo correlate the systolic arterial blood pressure recorded immediately before the ultrasound video with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber view
Correlation between systolic arterial blood pressure and push to bubbles time measured with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewSystolic arterial blood pressure is measured immediately before the ultrasound video with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewTo correlate the systolic arterial blood pressure measured immediately before the ultrasound video with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber view
Correlation between mean arterial blood pressure and push to bubbles time measured with the catheter tip placed in right atrium, using subcostal 4-chamber viewMean arterial blood pressure is measured immediately before the ultrasound video recording with the catheter tip in right atrium, using subcostal 4-chamber viewTo correlate the mean arterial blood measured immediately before the ultrasound video recording with the catheter tip placed in right atrium, using subcostal 4-chamber view
Correlation between mean arterial blood pressure and push to bubbles time measured with the catheter tip placed at CAJ, using subcostal 4-chamber viewMean arterial blood pressure is measured immediately before the ultrasound video recording with the catheter tip at CAJ, using subcostal 4-chamber viewTo correlate the mean arterial blood measured immediately before the ultrasound video recording with the catheter tip placed at CAJ, using subcostal 4-chamber view
Correlation between mean arterial blood pressure and push to bubbles time measured with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber viewMean arterial blood pressure is measured immediately before the ultrasound video recording with the catheter tip at 4 centimeters proximal to CAJ, using subcostal 4-chamber viewTo correlate the mean arterial blood measured immediately before the ultrasound video recording with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber view
Correlation between mean arterial blood pressure and push to bubbles time measured with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewMean arterial blood pressure is measured immediately before the ultrasound video recording with the catheter tip at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewTo correlate the mean arterial blood measured immediately before the ultrasound video recording with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber view
Correlation between diastolic arterial blood pressure and push to bubbles time measured with the catheter tip placed in right atrium, using subcostal 4-chamber viewDiastolic arterial blood pressure is measured immediately before the ultrasound video recording with the catheter tip in right atrium, using subcostal 4-chamber viewTo correlate the diastolic arterial blood measured immediately before the ultrasound video recording with the catheter tip placed in right atrium using subcostal 4-chamber view
Correlation between diastolic arterial blood pressure and push to bubbles time measured with the catheter tip placed at CAJ, using subcostal 4-chamber viewDiastolic arterial blood pressure is measured immediately before the ultrasound video recording with the catheter tip at CAJ, using subcostal 4-chamber viewTo correlate the diastolic arterial blood measured immediately before the ultrasound video recording with the catheter tip placed at CAJ using subcostal 4-chamber view
Correlation between diastolic arterial blood pressure and push to bubbles time measured with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber viewDiastolic arterial blood pressure is measured immediately before the ultrasound video recording with the catheter tip at 4 centimeters proximal to CAJ, using subcostal 4-chamber viewTo correlate the diastolic arterial blood measured immediately before the ultrasound video recording with the catheter tip placed at 4 centimeters proximal to CAJ using subcostal 4-chamber view
Correlation between diastolic arterial blood pressure and push to bubbles time measured with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewDiastolic arterial blood pressure is measured immediately before the ultrasound video recording with the catheter tip at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewTo correlate the diastolic arterial blood measured immediately before the ultrasound video recording with the catheter tip placed at 8 centimeters proximal to CAJ using subcostal 4-chamber view
Difference in time for the right atrium to be completely filled by micro bubbles; measure with the catheter tip placed in position 1: right atrium, using subcostal 4-chamber viewImmediately after measuring push to time in the video recorded with the catheter tip in right atrium, using subcostal 4-chamber viewThe time needed for the right atrium to be completely filled by micro bubbles will be measured analyzing the video recorded with the catheter tip in right atrium using subcostal bicaval view
Correlation between systolic pulmonary artery pressure and push to bubbles time measured with the catheter tip placed at CAJ, using subcostal 4-chamber viewSystolic pulmonary artery pressure is measured immediately before the ultrasound video recording with the catheter tip at CAJ, using subcostal 4-chamber viewTo correlate the systolic pulmonary artery pressure measured immediately before the ultrasound video recording with the catheter tip placed at CAJ, using subcostal 4-chamber view
Correlation between systolic pulmonary artery pressure and push to bubbles time measured with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber viewSystolic pulmonary artery pressure is measured immediately before the ultrasound video recording with the catheter tip at 4 centimeters proximal to CAJ, using subcostal 4-chamber viewTo correlate the systolic pulmonary artery pressure measured immediately before the ultrasound video recording with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber view
Correlation between systolic pulmonary artery pressure and push to bubbles time measured with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewSystolic pulmonary artery pressure is measured immediately before the ultrasound video recording with the catheter tip at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewTo correlate the systolic pulmonary artery pressure measured immediately before the ultrasound video recording with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber view
Correlation between diastolic pulmonary artery pressure and push to bubbles time measured with the catheter tip placed in right atrium, using subcostal 4-chamber viewDiastolic pulmonary artery pressure is measured immediately before the ultrasound video recording with the catheter tip placed in right atrium, using subcostal 4-chamber viewTo correlate the diastolic pulmonary artery pressure measured immediately before the ultrasound video recording with the catheter tip placed in right atrium, using subcostal 4-chamber view
Correlation between diastolic pulmonary artery pressure and push to bubbles time measured with the catheter tip placed at CAJ, using subcostal 4-chamber viewDiastolic pulmonary artery pressure is measured immediately before the ultrasound video recording with the catheter tip placed at CAJ, using subcostal 4-chamber viewTo correlate the diastolic pulmonary artery pressure measured immediately before the ultrasound video recording with the catheter tip placed at CAJ, using subcostal 4-chamber view
Correlation between diastolic pulmonary artery pressure and push to bubbles time measured with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber viewDiastolic pulmonary artery pressure is measured immediately before the ultrasound video recording with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber viewTo correlate the diastolic pulmonary artery pressure measured immediately before the ultrasound video recording with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber view
Correlation between diastolic pulmonary artery pressure and push to bubbles time measured with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewDiastolic pulmonary artery pressure is measured immediately before the ultrasound video recording with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewTo correlate the diastolic pulmonary artery pressure measured immediately before the ultrasound video recording with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber view
Correlation between central venous pressure and push to bubbles time measured with the catheter tip placed in right atrium, using subcostal 4-chamber viewCentral venous pressure is measured immediately before the ultrasound video recording with the catheter tip placed in right atrium, using subcostal 4-chamber viewTo correlate the central venous pressure measured immediately before the ultrasound video recording with the catheter placed in right atrium, using subcostal 4-chamber view
Correlation between central venous pressure and push to bubbles time measured with the catheter tip placed at CAJ, using subcostal 4-chamber viewCentral venous pressure is measured immediately before the ultrasound video recording with the catheter tip placed at CAJ, using subcostal 4-chamber viewTo correlate the central venous pressure measured immediately before the ultrasound video recording with the catheter tip placed at CAJ, using subcostal 4-chamber view
Correlation between central venous pressure and push to bubbles time measured with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber viewCentral venous pressure is measured immediately before the ultrasound video recording with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber viewTo correlate the central venous pressure measured immediately before the ultrasound video recording with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber view
Correlation between central venous pressure and push to bubbles time measured with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewCentral venous pressure is measured immediately before the ultrasound video recording with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewTo correlate the central venous pressure measured immediately before the ultrasound video recording with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber view
Correlation between cardiac output and push to bubbles time measured with the catheter tip placed in right atrium, using subcostal 4-chamber viewCardiac output is measured immediately before the ultrasound video recording with the catheter tip placed in right atrium, using subcostal 4-chamber viewTo correlate the cardiac output measured immediately before the ultrasound video recording with the catheter tip placed in right atrium, using subcostal 4-chamber view
Correlation between cardiac output and push to bubbles time measured with the catheter tip placed at CAJ, using subcostal 4-chamber viewCardiac output is measured immediately before the ultrasound video recording with the catheter tip placed at CAJ, using subcostal 4-chamber viewTo correlate the cardiac output measured immediately before the ultrasound video recording with the catheter tip placed at CAJ, using subcostal 4-chamber view
Correlation between cardiac output and push to bubbles time measured with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber viewCardiac output is measured immediately before the ultrasound video recording with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber viewTo correlate the cardiac output measured immediately before the ultrasound video recording with the catheter tip placed at 4 centimeters proximal to CAJ, using subcostal 4-chamber view
Correlation between cardiac output and push to bubbles time measured with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewCardiac output is measured immediately before the ultrasound video recording with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewTo correlate the cardiac output measured immediately before the ultrasound video recording with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber view
Correlation between systolic pulmonary artery pressure and push to bubbles time measured with the catheter tip placed in right atrium, using subcostal 4-chamber viewSystolic pulmonary artery pressure is measured immediately before the ultrasound video recording with the catheter tip in right atrium, using subcostal 4-chamber viewTo correlate the systolic pulmonary artery pressure measured immediately before the ultrasound video recording with the catheter tip placed in right atrium, using subcostal 4-chamber view
Difference in time for the right atrium to be completely filled by micro bubbles; measure with the catheter tip placed in position 1: right atrium, using subcostal bicaval viewImmediately after measuring push to time in the video recorded with the catheter tip placed in right atrium using subcostal bicaval viewThe time needed for the right atrium to be completely filled by micro bubbles will be measured analyzing the video recorded with the catheter tip in right atrium using subcostal bicaval view
Difference in time for the right atrium to be completely filled by micro bubbles; measure with the catheter tip placed in position 1: right atrium, using apical 4-chamber viewImmediately after measuring push to time in the video recorded with the catheter tip placed in right atrium, using apical 4-chamber viewThe time needed for the right atrium to be completely filled by micro bubbles will be measured analyzing the video recorded with the catheter tip in right atrium using apical 4-chamber view
Difference in time for the right atrium to be completely filled by micro bubbles; measure with the catheter tip placed in position 2: cavo-atrial junction (CAJ), using subcostal 4-chamber viewImmediately after measuring push to time in the video recorded with the catheter tip placed at cavo-atrial junction (CAJ), using subcostal 4-chamber viewThe time needed for the right atrium to be completely filled by micro bubbles will be measured analyzing the video recorded with the catheter tip placed at the CAJ, using subcostal 4-chamber view
Difference in time for the right atrium to be completely filled by micro bubbles; measure with the catheter tip placed in position 2: cavo-atrial junction (CAJ), using subcostal bicaval viewImmediately after measuring push to time in the video recorded with the catheter tip placed at cavo-atrial junction (CAJ), using subcostal bicavalThe time needed for the right atrium to be completely filled by micro bubbles will be measured analyzing the video recorded with the catheter tip placed at the CAJ, using subcostal bicaval view
Difference in time for the right atrium to be completely filled by micro bubbles; measure with the catheter tip placed in position 2: cavo-atrial junction (CAJ), using apical 4-chamber viewImmediately after measuring push to time in the video recorded with the catheter tip placed at cavo-atrial junction (CAJ), using apical 4-chamber viewThe time needed for the right atrium to be completely filled by micro bubbles will be measured analyzing the video recorded with the catheter tip placed at the CAJ, using apical 4-chamber view
Difference in time for the right atrium to be completely filled by micro bubbles; measure with the catheter tip placed in position 3: 4 centimeters proximal to CAJ in superior vena cava, using subcostal 4-chamber viewImmediately after measuring push to time in the video recorded with the catheter tip placed at 4 centimeters proximal to CAJ in superior vena cava, using subcostal 4-chamber viewThe time needed for the right atrium to be completely filled by micro bubbles will be measured analyzing the video recorded with the catheter tip placed at 4 centimeters proximal to CAJ in superior vena cava, using subcostal 4-chamber view
Difference in time for the right atrium to be completely filled by micro bubbles; measure with the catheter tip placed in position 3: 4 centimeters proximal to CAJ in superior vena cava, using subcostal bicaval viewImmediately after measuring push to time in the video recorded with the catheter tip placed at 4 centimeters proximal to CAJ in superior vena cava, using subcostal bicaval viewThe time needed for the right atrium to be completely filled by micro bubbles will be measured analyzing the video recorded with the catheter tip placed at 4 centimeters proximal to CAJ using subcostal bicaval view
Difference in time for the right atrium to be completely filled by micro bubbles; measure with the catheter tip placed in position 3: 4 centimeters proximal to CAJ in superior vena cava, using apical 4-chamber viewImmediately after measuring push to time in the video recorded with the catheter tip placed at 4 centimeters proximal to CAJ in superior vena cava, using apical 4-chamber viewThe time needed for the right atrium to be completely filled by micro bubbles will be measured analyzing the video recorded with the catheter tip placed at 4 centimeters proximal to CAJ in superior vena cava, using apical 4-chamber view
Difference in time for the right atrium to be completely filled by micro bubbles; measure with the catheter tip placed in position 4: 8 centimeters proximal to CAJ, using subcostal 4-chamber viewImmediately after measuring push to time in the video recorded with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal 4-chamber viewThe time needed for the right atrium to be completely filled by micro bubbles will be measured analyzing the video recorded with the catheter tip placed at 8 centimeters proximal to the CAJ, using subcostal 4-chamber view
Difference in time for the right atrium to be completely filled by micro bubbles; measure with the catheter tip placed in position 4: 8 centimeters proximal to CAJ, using subcostal bicaval viewImmediately after measuring push to time in the video recorded with the catheter tip placed at 8 centimeters proximal to CAJ, using subcostal bicaval viewThe time needed for the right atrium to be completely filled by micro bubbles will be measured analyzing the video recorded with the catheter tip placed at 8 centimeters proximal to the CAJ, using subcostal bicaval view
Difference in time for the right atrium to be completely filled by micro bubbles; measure with the catheter tip placed in position 4: 8 centimeters proximal to CAJ, using apical 4-chamber viewImmediately after measuring push to time in the video recorded with the catheter tip placed at 8 centimeters proximal to CAJ, using apical 4-chamber viewThe time needed for the right atrium to be completely filled by micro bubbles will be measured analyzing the video recorded with the catheter tip placed at 8 centimeters proximal to the CAJ, using apical 4-chamber view

Countries

Italy

Outcome results

None listed