Skip to content

DAISY Uterine Drain Device Evaluation

DAISY Uterine Drain: Device Evaluation With Standard Wall Suction

Status
Completed
Phases
Unknown
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT06219538
Enrollment
10
Registered
2024-01-23
Start date
2024-05-22
Completion date
2025-04-16
Last updated
2026-05-01

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Cesarean Section Complications, Post Partum Hemorrhage, Uterine Atony With Hemorrhage, Uterine Bleeding

Keywords

uterine drain, PPH

Brief summary

The goal of this study is to obtain user feedback while placing and observing the DAISY uterine drain with wall suction. This study defines the obstetrical surgeons as "users" and the patients in whom the drain is placed as "participants." Participants are pregnant women who are undergoing cesarean delivery (CD), who have not entered active labor, who have consented to drain placement and who have met all the inclusion/exclusion criteria. Users are staff or fellow obstetrical surgeons who will use the drain and provide the evaluation.

Detailed description

Postpartum hemorrhage (PPH) is the leading cause of maternal mortality worldwide. (D'Alton 2020). It is estimated that 14 million cases of PPH occur each year worldwide (Rath, 2011). In the United States, PPH causes up to 12% of maternal deaths and is the leading cause of death occurring on the day of delivery (ACOG 2017; Butwick 2019; Evensen 2020; Hawkins 2020). According to the American College of Obstetricians and Gynecologists (ACOG), between 54% and 93% of PPH-associated mortality may be preventable while 70-80% of PPH is due to abnormal uterine tone (atony) (ACOG 2017). Atony involves a deficient response to physiological signals that promote uterine contraction and blood vessel compression after delivery. Contraction of the myometrium compresses the blood vessels supplying the placental bed and thereby causes mechanical hemostasis. Uterine atony following CD is a relatively common problem in the non-laboring patient because contraction of the myometrium is almost entirely absent. Restoring uterine tone can be addressed through 1) medication, 2) surgery, or 3) mechanical means. Post-partum unrecognized uterine hemorrhage due to a closed or narrow cervix that allows blood to collect in a patient undergoing a CD could lead to complications such as dangerously low blood pressure. The DAISY drain is intended to provide a channel through the cervix for fluid drainage after pelvic surgery. The DAISY device consists of a soft silicone drainage tube (proximal end) attached to a semi-flexible catheter, inserted through the hysterotomy created at the time of the CD. The DAISY drain is placed intraoperatively after delivery of the placenta, stabilization of the patient, and cleansing of the uterine cavity but prior to closure of the hysterotomy. The drain is placed through the hysterotomy and threaded through the cervix and vaginal canal, out the introitus, so the distal end is accessible outside of the body. In this study, the drain is placed on continuous suction according to the IFU for the first two postoperative hours or more if required. At the end of the 2-hour period, the uterus is evaluated by manual palpation and abdominal ultrasound. The abdominal ultrasound is performed to document device placement and to assess uterine involution, approximate uterine size and uterine cavity size, presence of intrauterine clots or retained products of conception. The trained user removes the drain after at least two hours post-surgery based on clinical management. Suction is discontinued before removal. One method of determining if the drain is no longer needed is by checking and documenting uterine tone. If uterine tone has returned to normal as assessed by palpation, the DAISY drain may be removed.

Interventions

DEVICEDAISY Uterine Drain

Insertion of the DAISY Uterine Drain and connection to wall suction in women who have undergone a Caesarean Delivery,

Sponsors

Raydiant Oximetry, Inc.
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DEVICE_FEASIBILITY
Masking
NONE

Intervention model description

The participants consist of women undergoing planned CD for reasons unrelated to this study.

Eligibility

Sex/Gender
FEMALE
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* The participants consist of women undergoing planned CD for reasons unrelated to this study.

Exclusion criteria

* Known intrauterine or cervical pathology that would interfere with device placement and/or use * Ongoing intrauterine pregnancy * Untreated uterine rupture * Unresolved uterine inversion * Current cervical cancer * Current purulent infection of vagina, cervix, or uterus * Retained products of conception * Arterial bleeding requiring surgical or angiographic embolization * Indication for hysterectomy * Lack of study consent or unable to provide informed consent, * Any condition (temporary or permanent) in which the investigator or user deems the patient unsuitable for the study procedures

Design outcomes

Primary

MeasureTime frameDescription
Device EvaluationDay of surgery (preoperative) and one day post procedureThis is a prospective, single-arm, device evaluation study of device use by five (5) obstetrical surgeons placing the drain with low-level suction in up to ten (10) women undergoing elective, non-emergent cesarean delivery. Hemoglobin was measured at two time points: baseline (preoperative) and 1 day post operative. The mean change in hemoglobin concentration between baseline (preoperative) values obtained on the day of surgery and postoperative day 1 was analyzed and compared to historical means in the US in 31 hospitals with 5470 participants.

Countries

United States

Contacts

PRINCIPAL_INVESTIGATORTetsuya Kawakita, MD

Eastern Virginia Medical School

Participant flow

Recruitment details

Pregnant subjects who were planning a Cesarean Delivery (CD) and had one or more risks for post-partum hemorrhage where enrolled at the care center.

Baseline characteristics

Characteristic
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
10 Participants
Sex/Gender, Customized10 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 10
other
Total, other adverse events
0 / 10
serious
Total, serious adverse events
0 / 10

Outcome results

None listed