Skip to content

Project VerioVue Enhancements - Neonate Study

VerioVue Enhancements Regulatory Clinical Evaluation - Professional Comparison to Validated Method: Neonate Study

Status
Terminated
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT06052371
Enrollment
99
Registered
2023-09-25
Start date
2023-10-16
Completion date
2023-12-01
Last updated
2023-12-27

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes Mellitus

Keywords

Neonate

Brief summary

The goal of this performance evaluation is to compare the blood glucose results obtained using the VerioVue Blood Glucose Monitoring system (BGMS) with those obtained from a validated comparator method (iSTAT 1 Analyser) using blood obtained from neonates. The main question it aims to answer is: •How accurate is the VerioVue BGMS when compared to a product that has already been confirmed as accurate (iSTAT 1 Analyser) when hospital staff test blood taken from neonates on these instruments? Participants (neonates) will have a small amount of blood taken from a heel prick (performed by a HCP for medical purposes) or from an existing arterial line.

Detailed description

UK NHS Hospital sites (between 1 and 3 sites) will be used to collect a minimum of 200 evaluable blood samples from neonates meeting the inclusion criteria. Consent will be taken from the neonates parent prior to participation in the study. They will consent to one or two blood samples to be taken from their child. If the participant meets the enrolment criteria then demographics and prescription medication (including dietary supplements) information will be collected. A small volume of blood (max 150µl) will be collected from an existing arterial line or from a heel prick that is being performed for medical purposes. This blood will then be applied to six investigational One Touch VerioVue meters (with expanded haematocrit range of 20-65%) fitted with OneTouch Verio test strips and the blood glucose meter readings recorded. Three different lots of test strips will used in rotation across the study. Within 5 minutes of the last meter test, NHS study staff will use blood from the same sample and perform a test on the iSTAT 1 Analyser to obtain blood glucose, haematocrit and oxygen levels of the sample. No treatment decisions will be made based on the blood glucose results obtained for this study. The study documentation will be reviewed by LifeScan monitors for completeness and accuracy and the results entered into a validated database. The data will be extracted and statistically analysed to determine the accuracy of the blood glucose results obtained using the OneTouch VerioVue BGMS when compared to the blood glucose results obtained on the iSTAT 1 Analyser.

Interventions

DEVICEVerioVue (Enhancements) blood glucose monitoring system

VerioVue meter with expanded haematocrit range

Sponsors

LifeScan Scotland Ltd
Lead SponsorINDUSTRY

Study design

Observational model
CASE_ONLY
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
0 Days to 28 Days
Healthy volunteers
Yes

Inclusion criteria

* Age - Newborn babies, 28 days old or less with a corrected gestational age of at least 34 weeks and a weight of at least 1700g or more at the time of participation. * Informed Consent - Study participants parents must read the Parental Participant Information Sheet and sign the Parental Informed Consent Form * Study participants parent agrees to provide information related to demographics, prescription medication, dietary supplements, and results from other physician ordered testing e.g., Bilirubin (total, conjugated, and unconjugated) where available. * Study participants parent agrees to allow study staff access to medical records where necessary. * Study participants parents agree to all aspects of the study process, including where applicable arterial blood draw from an existing arterial line or a heel prick performed by an HCP for medical purposes.

Exclusion criteria

* Age- Newborn more than 28 days old. * Study participants with a gestational age of less than 34 weeks at the time of participation. * Study participants with a gestational weight of less than 1700g * Current positive test result for Covid-19.

Design outcomes

Primary

MeasureTime frameDescription
Performance evaluation: blood glucose monitoring system accuracy: BGMS vs reference instrument1 hourPerformance Evaluation to compare the VerioVue BGMS blood glucose results to results obtained using the iSTAT 1 Analyser, a validated comparator method.

Countries

United Kingdom

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026