Skip to content

Validation of Digital Heart Rhythm Devices in the Detection of Atrial Fibrillation

Validation of Digital Heart Rhythm Devices in the Detection of Atrial Fibrillation

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT06023290
Acronym
VALIDATION
Enrollment
126
Registered
2023-09-05
Start date
2023-07-03
Completion date
2023-11-17
Last updated
2023-12-01

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Atrial Fibrillation

Brief summary

The VALIDATION study aims to evaluate the performance of four state-of-the-art digital heart rhythm devices (Preventicus®, FibriCheck®, Apple Watch®, 6L Kardia Mobile®) for the detection of atrial fibrillation.

Interventions

DIAGNOSTIC_TESTHeart rhythm measurements with Preventicus®

The sequence of the digital heart rhythm measurements is randomly assigned for each patient. The reference heart rhythm is determined by a simultaneous recording with a 12-lead ECG as gold standard.

The sequence of the digital heart rhythm measurements is randomly assigned for each patient. The reference heart rhythm is determined by a simultaneous recording with a 12-lead ECG as gold standard.

DIAGNOSTIC_TESTHeart rhythm measurements with Apple Watch®

The sequence of the digital heart rhythm measurements is randomly assigned for each patient. The reference heart rhythm is determined by a simultaneous recording with a 12-lead ECG as gold standard.

DIAGNOSTIC_TESTHeart rhythm measurements with 6L Kardia Mobile®

The sequence of the digital heart rhythm measurements is randomly assigned for each patient. The reference heart rhythm is determined by a simultaneous recording with a 12-lead ECG as gold standard.

Sponsors

Ziekenhuis Oost-Limburg
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* The patient is an ambulatory cardiology patient. * The patient is at least 18 years old. * The patient is sinus or in atrial fibrillation. * The patient is able to perform the study procedures together with the study assistant.

Exclusion criteria

* Patients with a pacemaker * Inclusion in another clinical trial that may influence the results of this trial * The patient is unable to comprehend the informed consent.

Design outcomes

Primary

MeasureTime frameDescription
Cramer's VDuring the heart rhythm measurementsThe association of between the classification by the device of interest versus the 12-lead ECG, expressed as Cramer's V.
SensitivityDuring the heart rhythm measurementsProportion of true positives tests with the device of interest out of all patients with AF on the 12-lead ECG.
SpecificityDuring the heart rhythm measurementsProportion of true negative tests with the device of interest out of all patients in sinus rhythm on the 12-lead ECG.
AccuracyDuring the heart rhythm measurementsProportion of correct classifications with the device of interest. Correct classification is provided by the 12-lead ECG.

Secondary

MeasureTime frameDescription
Positive predictive valueDuring the heart rhythm measurementsProportion of true positive tests out of all positive tests.
Negative predictive valueDuring the heart rhythm measurementsProportion of true negative tests out of all negative tests.
Patient preference score (for the use of the diagnostic devices)During the heart rhythm measurementsThis score is a (1-5) ordinal result based on the VALIDATION study questionnaire.

Other

MeasureTime frameDescription
Sensitivity, specificity and Cramer's V using only measurements of sufficient quality for analysisDuring the heart rhythm measurementsThe proportion of true positive tests, true negative tests and the association of between the classification of the test devices and 12-lead ECG.

Countries

Belgium

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026