Clinical Trial of a Hydrophilic EMV Toric Lens RAO210T in the Correction of Aphakia and Corneal Astigmatism

Clinical Trial to Investigate the Safety and Effectiveness of a Hydrophilic EMV Toric Lens RAO210T in the Correction of Aphakia and Post-operative Corneal Astigmatism

Status
Completed
Phases
Unknown
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05985304
Enrollment
274
Registered
2023-08-14
Start date
2023-08-10
Completion date
2024-10-23
Last updated
2026-01-21

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Cataract, Corneal Astigmatism

Keywords

Toric Intraocular Lens

Brief summary

The objectives of the clinical investigation are to determine the safety and performance of the RayOne EMV Toric (Model RAO210T) following unilateral implantation and approximately 6 months (120 to 180 days) of post-operative assessment, as a randomized comparison to aspheric monofocal control for low cylinder (1.50 D). This patient population with cataracts and low corneal cylinder (+1.00 to +1.50 D) will have the natural crystalline lens removed and replaced with an intraocular lens in the capsular bag.

Detailed description

This study is a prospective, multicenter, randomized, active controlled, masked (assessor and subject) pivotal investigation for unilateral implantation of low cylinder (1.50 D) toric IOL. It compares the Rayner RAO210T toric IOL to the Rayner RAO600C aspheric monofocal IOL. Subjects who sign the ICF are considered enrolled in the study. After signing the ICF, subjects will be screened for eligibility. Inclusion and Exclusion Criteria must be applied prior to subject randomization. Subjects who meet all protocol-specified eligibility criteria will be assigned to the randomized controlled study arms if the study eye's estimated IOL cylinder power using Barrett Toric Calculator is 1.50 D. Randomized subjects receive either the Rayner RAO210T Toric IOL (1.50 D low cylinder) or the Rayner RAO600C aspheric monofocal IOL in the study eye. Up to 295 adult subjects will be enrolled (consented) assuming a 15% screen failure rate, then up to 250 subjects will be randomized, of which approximately 125 subjects will be randomized to receive low cylinder (1.50 D) RayOne EMV Toric (Model RAO210T) in one eye and approximately 125 subjects randomized to receive the RAO600C aspheric monofocal IOL in one eye, to complete 120-180 days follow-up for at least 100 subjects in each group. All subjects who meet inclusion criteria will be assigned to the randomized controlled study arms to receive either the RAO210T toric IOL (1.50 D low cylinder) or the RAO600C aspheric monofocal IOL in the study eye according to a 1:1 ratio. Randomization will be stratified by site. Up to 11 U.S. sites will be encouraged to enroll a minimum of 20 subjects. No site will enroll more than 25% of the subjects enrolled in the study. If a subject has significant cataract in both eyes, it is recommended that cataract surgery is performed in one eye before the subject is enrolled in the study. Once a subject has been enrolled, it is recommended that the fellow eye does not undergo cataract surgery (except for a YAG capsulotomy) throughout the duration of the study. At screening, if both eyes qualify for the study, the eye to undergo cataract surgery and IOL implantation is the eye with worse pre-operative BCDVA. If pre-operative BCDVA is the same for each eye, the right eye will be the study eye. Subjects will complete 6 study visits in approximately 9 months. Subject participation is calculated as the difference between the time of the pre-operative visit to completion of Visit 4 (120 to 180 days post-operative). Standard clinical trial methods will be used to minimize bias, such as the use of site personnel performing manifest refraction and visual acuity assessments masked to subject treatment assignment, masking of subjects in the randomized controlled investigation evaluating low cylinder power, standardized test procedures, common Investigator training and common inclusion and exclusion criteria. In order to minimize bias, measures will be taken to mask the site personnel performing post-operative manifest refraction and visual acuity assessments, as to the subject's treatment assignment until after the final database lock. Every attempt should be made to have the same masked site personnel perform the same masked post-operative assessments for an individual subject throughout the subject's study participation. Subjects will be masked to their IOL assignment in the randomized controlled investigation evaluating low cylinder power. All material which may indicate the subjects' assignment, e.g., packaging, documents, etc., will be removed from any areas where subjects and/or masked site personnel may see them. Unmasked personnel will further be instructed to scrupulously avoid conversation and communication with masked personnel, subjects and all other persons regarding subjects' assignments, outcomes, clinical courses, and all other information potentially relevant to the study and its conduct. Study Groups: * Toric (test) group - all subjects randomized to be implanted with the low cylinder (1.50 D) RayOne EMV Toric IOL (Model RAO210T) * Monofocal (control) group - all subjects randomized to be implanted with the RAO600C Monofocal IOL Sample Size Determination: Subjects will be enrolled with the goal of completing a total of 200 subjects through Visit 4 within the randomized groups, of whom 100 were implanted with the low cylinder RayOne EMV Toric IOL (Model RAO210T) and 100 were implanted with the RAO600C monofocal IOL. Effectiveness For effectiveness, 100 subjects (eyes) randomized to the Toric (test) group (lowest cylinder power +1.50D) and 100 subjects (eyes) randomized to the Monofocal (control) group completing Visit 4 yields over 98% statistical power to reject H0e in favor of H1e and conclude the Toric IOL has statistically significantly lower mean residual manifest cylinder at Visit 4 compared to the Monofocal IOL, using a one-sided t-test with an alpha level of 0.025 and assuming a true mean difference of 0.4 D and a standard deviation of 0.7 D in both groups. Safety For safety, ISO 11979-7 and ANSI Z80.30 specifies that a minimum of 100 subjects should complete a clinical evaluation of an IOL, where a parent IOL has been approved, to obtain appropriate specificity around adverse event and visual acuity rates. Handling of Missing Data: Missing data will be imputed for selected endpoints using the methods specified under the analysis descriptions for the endpoints. Where possible, multiple imputation will be used. Except where mentioned in Section 7, missing data will not be imputed. Multiplicity Considerations: The overall type I error rate for the toric effectiveness analyses will be controlled at 0.05. The study will be considered successful if all of the primary effectiveness and safety endpoints are met. Thus, no adjustments for multiplicity are necessary. All primary effectiveness and primary safety endpoints as described below are required to achieve successful outcomes in order to demonstrate study success.

Interventions

DEVICERayOne EMV Toric IOL

Implantation of intraocular lens.

DEVICERayOne Monofocal IOL

Implantation of intraocular lens.

PROCEDURECataract Surgery

Removal of natural crystalline lens due to cataracts

Sponsors

Rayner Intraocular Lenses Limited
Lead SponsorINDUSTRY

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
TREATMENT
Masking
TRIPLE (Subject, Caregiver, Outcomes Assessor)

Eligibility

Sex/Gender
ALL
Age
22 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1. Male or female, 22 years or older at the pre-operative visit who have cataract with best corrected distance visual acuity of 0.30 logMAR (20/40) or worse in at least one eye with or without a glare source present who are eligible for phacoemulsification cataract surgery 2. Subjects who are projected to have best corrected distance visual acuity 0.20 logMAR (20/30) or better after IOL implantation by potential acuity meter (PAM) or Investigator estimation 3. Clear intraocular media other than cataract 4. Contact lens wearers must demonstrate stability of biometry 5. Have the capability to understand and sign an IRB approved informed consent form and privacy authorization in accordance with local regulations 6. Female subjects must be 1-year postmenopausal, surgically sterilized, or, if of childbearing potential, have a negative urine pregnancy test at the Pre-operative Visit. Women of childbearing potential must use an acceptable form of contraception throughout the study. Acceptable methods include at least one of the following: intrauterine (intrauterine device), hormonal (oral, injection, patch, implant, ring), barrier with spermicide (condom, diaphragm), or abstinence. 7. Have Investigator selected IOL spherical equivalent power between +10.0 D to +25.0 D in 0.5 D steps and IOL cylinder power of +1.50 D 8. Have pre-existing corneal astigmatism of 1.00 D to 1.50 D as determined by keratometry 9. Dilated pupil size 5.5 mm or greater to allow visualization of the toric IOL axis markings post-operatively

Exclusion criteria

1. Previous intraocular, corneal, or retinal detachment surgery, including corneal transplant, LASIK / LASEK / PRK, SMILE, astigmatic keratotomy and limbal relaxing incisions in the planned operative eye. 2. Diagnosed degenerative visual disorders (e.g. macular degeneration, retinal detachment, proliferative diabetic retinopathy, or other retinal disorders) that are predicted to cause future acuity losses to a level of 0.20 logMAR (20/30) or worse 3. Significant anterior segment pathology that might increase intraoperative risk or compromise IOL stability (e.g. pseudoexfoliation syndrome, any iris pathology) 4. Subjects with conditions associated with increased risk of zonular rupture (that may affect post-operative centration or tilt of IOL) in the planned operative eye 5. Potentially occludable angle or ciliary body tumor, or other pathology that might increase risk to subject safety, based on gonioscopic observation 6. Subjects reasonably expected to require secondary ocular surgical intervention or laser treatment (other than YAG capsulotomy) 7. Subjects with clinically significant corneal pathology, potentially affecting corneal topography 8. Subjects with traumatic cataract in the planned operative eye 9. Participating in a concurrent drug or device clinical trial or who have participated in a drug or device trial within 30 days of the pre-operative visit 10. Subjects with any other serious ocular pathology (e.g. glaucoma, severe dry eye, history of intraocular inflammation, history of retinal surgery or retinal laser procedure) or underlying systemic medical condition (e.g., uncontrolled diabetes) or circumstance that, based on the Investigator's judgment, poses a concern for the subjects' safety or could confound the results of the study (History of cataract surgery with PC-IOL in one eye is allowed.) 11. Use of medications known to interfere with visual performance, pupil dilation, or iris structure within 30 days of the pre-operative visit, at the discretion of the Investigator 12. Pregnant or nursing females 13. Irregular astigmatism in the planned operative eye

Design outcomes

Primary

MeasureTime frameDescription
Mean Magnitude of Residual Manifest Cylinder120 to 180 days post-operativelyMean magnitude of residual manifest cylinder (as measured by manifest refractive cylinder).
Adjusted Mean Difference Magnitude of Residual Manifest Cylinder120 to 180 days post-operativelyThe adjusted mean difference of residual manifest cylinder (as measured by manifest refractive cylinder) between Toric and Monofocal groups will be given together with a 95% confidence interval and associated one-sided p-value. This endpoint will be considered successful if the one-sided p-value ≤ 0.025.
Axis Misalignment (Frequency and Percentage 10 Degrees)120 to 180 days post-operativelyFrequency and percentage of RayOne EMV Toric (Model RAO210T) IOLs with IOL axis misalignment at Visit 4 (as determined by photographic method) less than 10 degrees.
Axis Misalignment (Frequency and Percentage 20 Degrees)120 to 180 days post-operativelyFrequency and percentage of RayOne EMV Toric (Model RAO210T) IOLs with IOL axis misalignment at Visit 4 (as determined by photographic method) less than 20 degrees.
Stability of Toric IOL Axis Orientation (Frequency and Percentage)30 to 60 days and 120 to 180 days post-operativelyStability of toric IOL axis orientation, expressed as frequency and percentage of RayOne EMV Toric (Model RAO210T) IOLs that rotate ≤ 5 degrees postoperatively between 30 to 60 days (Visit 3) and 120 to 180 days (Visit 4)
Rates of IOL Adverse Events Per ISO 11979-7120 to 180 days post-operativelyRates of IOL adverse events through 120 to 180 days post-operatively (Visit 4) compared to the ISO Safety and Performance Endpoint (SPE) rates as described in ISO 11979-7. Rates of cumulative and persistent adverse events from ISO 11979-7 (2018) Table E.2 will be reported by treatment group (Toric test group, Monofocal control group) and AE type. Rates of adverse events will be compared to the SPE rate from Table E.2. The one-sided p-value from the exact binomial test of each hypothesis will also be provided for each adverse event. If none of these hypothesis test results are statistically significant, this endpoint will be considered successful for the toric IOL.
Rates of All Other Adverse Events Not Included in ISO 11979-7120 to 180 days post-operativelyThe eyes with at least one adverse event not included in the cumulative and persistent adverse events from ISO 11979-7 (2018) Table E.2 through 120 to 180 days post-operatively (Visit 4) will be presented by treatment group (Toric test group, Monofocal control group).
Rates of Secondary Surgical Interventions for IOL Repositioning Due to IOL Misalignment120 to 180 days post-operativelyThe percentage of eyes implanted with the Toric test group IOL requiring a secondary surgical intervention for IOL repositioning due to IOL misalignment through 120 to 180 days post-operatively (Visit 4), as well as a two-sided 95% exact binomial confidence interval for the percentage, will be presented.
Rate of BCDVA of 0.30 logMAR or Better Compared to the ISO SPE Rates120 to 180 days post-operativelyThe percentage of eyes that achieve a BCDVA at 4 meters of 0.30 logMAR or better at 120 to 180 days post-operatively (Visit 4) will be presented by treatment group (Toric test group, Monofocal control group) for the Safety Set and Best Case Set. If neither of these hypothesis test results are statistically significant, this endpoint will be considered successful for the toric IOL.

Secondary

MeasureTime frameDescription
Residual Manifest Cylinder Stratified by Preoperative Keratometric Cylinder120 to 180 days post-operativelyResidual manifest refractive cylinder by subgroups of 0.25 D preoperative keratometric cylinder at 120 to 180 days post-operatively
Percent Change in Absolute Cylinder Stratified by Preoperative Keratometric Cylinder120 to 180 days post-operativelyPercent reduction in absolute cylinder (as measured by residual manifest cylinder / preoperative keratometric cylinder), at 120 to 180 days post-operatively. Positive numbers represent an decrease in cylinder and negative numbers represent an increase in cylinder.

Countries

United States

Contacts

STUDY_CHAIRMark Packer, MD

Participant flow

Pre-assignment details

Of the 274 enrolled subjects, 32 subjects were screen failures who did not meet the study-specific inclusion/exclusion criteria (32/274, 11.7%), five subjects were withdrawn before randomization (5/274, 1.7%)

Baseline characteristics

Characteristic
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
97 Participants
Age, Categorical
Between 18 and 65 years
36 Participants
Age, Continuous69.4 years
STANDARD_DEVIATION 7
Anterior Chamber Depth3.247 mm
STANDARD_DEVIATION 0.381
Axial Length24.120 mm
STANDARD_DEVIATION 1.117
Best-corrected Distance Visual Acuity (BCDVA) with Glare0.863 logMAR
STANDARD_DEVIATION 0.551
Best-corrected Distance Visual Acuity (BCDVA) Without Glare0.227 logMAR
STANDARD_DEVIATION 0.236
Intraocular Lens (IOL) Power Spherical Equivalent (SE)18.10 Diopter
STANDARD_DEVIATION 3.29
Keratometric Cylinder1.247 Diopter
STANDARD_DEVIATION 0.136
LOCS III: Cortical Opacity
C1
29 Participants
LOCS III: Cortical Opacity
C2
20 Participants
LOCS III: Cortical Opacity
C3
18 Participants
LOCS III: Cortical Opacity
C4
7 Participants
LOCS III: Cortical Opacity
C5
3 Participants
LOCS III: Cortical Opacity
NA
44 Participants
LOCS III: Nuclear Color
NA
2 Participants
LOCS III: Nuclear Color
NC1
14 Participants
LOCS III: Nuclear Color
NC2
35 Participants
LOCS III: Nuclear Color
NC3
50 Participants
LOCS III: Nuclear Color
NC4
11 Participants
LOCS III: Nuclear Color
NC5
10 Participants
LOCS III: Nuclear Color
NC6
1 Participants
LOCS III: Nuclear Opalescence
N/A
21 Participants
LOCS III: Nuclear Opalescence
NO1
11 Participants
LOCS III: Nuclear Opalescence
NO2
38 Participants
LOCS III: Nuclear Opalescence
NO3
87 Participants
LOCS III: Nuclear Opalescence
NO4
13 Participants
LOCS III: Nuclear Opalescence
NO5
4 Participants
LOCS III: Nuclear Opalescence
NO6
0 Participants
LOCS III: Posterior Subcapsular
NA
139 Participants
LOCS III: Posterior Subcapsular
P1
54 Participants
LOCS III: Posterior Subcapsular
P2
12 Participants
LOCS III: Posterior Subcapsular
P3
4 Participants
LOCS III: Posterior Subcapsular
P4
7 Participants
LOCS III: Posterior Subcapsular
P5
0 Participants
Manifest Refraction Cylinder-1.038 Diopter
STANDARD_DEVIATION 0.592
Manifest Refraction Optical Infinity Spherical Equivalent (MROISE)-1.1396 Diopter
STANDARD_DEVIATION 2.7307
Manifest Refraction Spherical Equivalent (MRSE) at 4m-0.8896 Diopter
STANDARD_DEVIATION 2.7307
Preoperative Keratometric Cylinder Subgroup
1.0 to <1.25 D
63 Participants
Preoperative Keratometric Cylinder Subgroup
1.25 to 1.5 D
99 Participants
Race/Ethnicity, Customized
American Indian or Alaska Native
2 Participants
Race/Ethnicity, Customized
Asian
0 Participants
Race/Ethnicity, Customized
Black or African American
4 Participants
Race/Ethnicity, Customized
Hispanic or Latino
3 Participants
Race/Ethnicity, Customized
Multiple
0 Participants
Race/Ethnicity, Customized
Native Hawaiian or Other Pacific Islander
0 Participants
Race/Ethnicity, Customized
Not Hispanic or Latino
115 Participants
Race/Ethnicity, Customized
White
115 Participants
Region of Enrollment
United States
120 participants
Sex: Female, Male
Female
136 Participants
Sex: Female, Male
Male
52 Participants
Study Eye
Left Eye (OS)
103 Participants
Study Eye
Right Eye (OD)
62 Participants
Target Postoperative (PO) Residual Astigmatism0.191 Diopter
STANDARD_DEVIATION 0.115
Target Residual Spherical Equivalent (SE)-0.012 Diopter
STANDARD_DEVIATION 0.099
Uncorrected Distance Visual Acuity (UCDVA)0.556 logMAR
STANDARD_DEVIATION 0.363

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 1170 / 116
other
Total, other adverse events
49 / 11740 / 116
serious
Total, serious adverse events
1 / 1170 / 116

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026