A Study of Non-invasive Measurement of Blood Glucose and Blood Pressure

Non-invasive Measurement of Blood Glucose, Blood Pressure and Extrapolation of QRS Complex Using a Wrist Wearable Photoplethysmography Sensor

Status
Terminated
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05904769
Enrollment
87
Registered
2023-06-15
Start date
2023-05-31
Completion date
2024-07-31
Last updated
2025-10-08

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes

Brief summary

The purpose of this research is to compare blood glucose, blood pressure and electrocardiogram (ECG) readings on the LIFELEAF Smartwatch compared to standard methods of measurement.

Interventions

DEVICELIFELEAF Smartwatch

Noninvasive wearable smartwatch/monitor that utilizes photoplethysmography (PPG) which is a set of measurements acquired from the light flashed from the back of the watch and measuring the intensity of the reflected light from the blood in the blood vessels just below the skin at the wrist. Provides a non-invasive method for measuring blood glucose as well as alternative method for measuring blood pressure, heart rate and rhythm, etc.

Sponsors

Mayo Clinic
Lead SponsorOTHER

Study design

Allocation
NON_RANDOMIZED
Intervention model
PARALLEL
Primary purpose
HEALTH_SERVICES_RESEARCH
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Diabetic patients on insulin undergoing four times daily blood sugar checks as part of standard of care. * Patients undergoing continuous monitoring for heart rate, ECG and blood pressure as part of their routine care. * Patients undergoing invasive and noninvasive electrophysiologic procedures.

Exclusion criteria

* Patients unable to provide informed consent. * Patients without arms as these are needed to wear the wrist-based sensor.

Design outcomes

Primary

MeasureTime frameDescription
Systolic Blood PressureBaselineDifference in measurement of Systolic Blood Pressure using the wrist-based PPG sensor versus invasive measurement of blood pressure measured in millimeters of mercury (mmHg).
Diastolic Blood PressureBaselineDifference in measurement of Diastolic Blood Pressure using the wrist-based PPG sensor versus invasive measurement of blood pressure measured in millimeters of mercury (mmHg).
Blood Glucose LevelBaselineThe difference in measurement of blood glucose levels using the wrist-based PPG sensor versus standard blood-based measurement measured in milligrams per deciliter (mg/dL) in subjects who had standard of care blood glucose monitoring done. Only subjects who are hospitalized and diabetic per standard of care receive standard of care blood-based glucose monitoring.

Countries

United States

Participant flow

Pre-assignment details

Zero subjects were enrolled in the hospitalized diabetic arm as the study was terminated early due to a lack of staffing support.

Participants by arm

ArmCount
Hospitalized Diabetic Group
Subjects hospitalized and diabetic on insulin, undergoing four times daily blood sugar checks, wore the LIFELEAF Smartwatch for the duration of their hospital stay. LIFELEAF Smartwatch: Noninvasive wearable smartwatch/monitor that utilizes photoplethysmography (PPG) which is a set of measurements acquired from the light flashed from the back of the watch and measuring the intensity of the reflected light from the blood in the blood vessels just below the skin at the wrist. Provides a non-invasive method for measuring blood glucose as well as alternative method for measuring blood pressure, heart rate and rhythm, etc.
0
Electrophysiologic (EP) Group
Subjects undergoing an EP procedure wore the LIFELEAF Smartwatch for the duration of the procedure. LIFELEAF Smartwatch: Noninvasive wearable smartwatch/monitor that utilizes photoplethysmography (PPG) which is a set of measurements acquired from the light flashed from the back of the watch and measuring the intensity of the reflected light from the blood in the blood vessels just below the skin at the wrist. Provides a non-invasive method for measuring blood glucose as well as alternative method for measuring blood pressure, heart rate and rhythm, etc.
87
Total87

Withdrawals & dropouts

PeriodReasonFG000FG001
Overall StudyPhysician Decision06

Baseline characteristics

CharacteristicTotalElectrophysiologic (EP) Group
Age, Continuous62.3 years
STANDARD_DEVIATION 10.24
62.3 years
STANDARD_DEVIATION 10.24
Race and Ethnicity Not Collected0 Participants
Region of Enrollment
United States
87 participants87 participants
Sex: Female, Male
Female
27 Participants27 Participants
Sex: Female, Male
Male
60 Participants60 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 00 / 87
other
Total, other adverse events
0 / 00 / 87
serious
Total, serious adverse events
0 / 00 / 87

Outcome results

Primary

Blood Glucose Level

The difference in measurement of blood glucose levels using the wrist-based PPG sensor versus standard blood-based measurement measured in milligrams per deciliter (mg/dL) in subjects who had standard of care blood glucose monitoring done. Only subjects who are hospitalized and diabetic per standard of care receive standard of care blood-based glucose monitoring.

Time frame: Baseline

Population: 0 subjects enrolled in the hospitalized diabetic arm as study was terminated early (lack of staffing support). Data not collected for this outcome measure in the EP group as it didn't apply to the EP arm. Standard blood-based glucose measurements were only measured on subjects that were diabetic + hospitalized. EP subjects were not hospitalized; due to this standard blood glucose was not monitored in this group therefore a comparison couldn't be made to the PPG sensor blood glucose measurements.

Primary

Diastolic Blood Pressure

Difference in measurement of Diastolic Blood Pressure using the wrist-based PPG sensor versus invasive measurement of blood pressure measured in millimeters of mercury (mmHg).

Time frame: Baseline

Population: Zero subjects were enrolled in the hospitalized diabetic arm as the study was terminated early due to a lack of staffing support.

ArmMeasureValue (MEAN)Dispersion
Electrophysiologic (EP) GroupDiastolic Blood Pressure8.42 mmHgStandard Deviation 10.56
Primary

Systolic Blood Pressure

Difference in measurement of Systolic Blood Pressure using the wrist-based PPG sensor versus invasive measurement of blood pressure measured in millimeters of mercury (mmHg).

Time frame: Baseline

Population: Zero subjects were enrolled in the hospitalized diabetic arm as the study was terminated early due to a lack of staffing support.

ArmMeasureValue (MEAN)Dispersion
Electrophysiologic (EP) GroupSystolic Blood Pressure13.08 mmHgStandard Deviation 17.22