SpO2 Data Collection in Pediatric Patients Using INVSENSOR00061

SpO2 Data Collection in Pediatric Patients Using INVSENSOR00061

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05896267
Enrollment
64
Registered
2023-06-09
Start date
2023-05-15
Completion date
2023-06-22
Last updated
2025-11-12

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Spo2

Brief summary

This is a prospective, non-randomized data collection study to evaluate the form, fit, and function of INVSENSOR00061.

Interventions

DEVICEMasimo INVSENSOR00061

Masimo INVSENSOR00061 will be used to measure noninvasive SpO2 (Oxygen Saturation )

Sponsors

Masimo Corporation
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
37 Weeks to 18 Months
Healthy volunteers
Yes

Inclusion criteria

* • Subject is a full-term newborn (37 weeks) to 18 months of age.

Exclusion criteria

* • Subject has underdeveloped skin. * Subject has a skin condition and/or deformity at the planned application site, which would preclude sensor placement and measurements. * Subject has an absence or deformities of limbs or severe edema, which would interfere with sensor application or prevent the proper fit of the sensors. * Subject is not suitable for the investigation at the discretion of the Investigator.

Design outcomes

Primary

MeasureTime frameDescription
SpO2 Accuracy of INVSENSOR00061Approximately 15 minutesSpO2 accuracy will be determined by calculating the Arithmetic Root Mean Square (ARMs) value which will be reported as percent of oxygen saturated hemoglobin. The noninvasive oxygen saturation measurement (SpO2) from the INVSENSOR00061 will be compared to the noninvasive oxygen saturation measurement (SpO2) value from the RD SET Neo sensor.
Pulse Rate Accuracy of INVSENSOR00061Approximately 15 minutesPulse Rate accuracy will be determined by calculating the Arithmetic Root Mean Square (ARMs) value through the comparison of the noninvasive Pulse Rate to the noninvasive Pulse Rate (PR) value obtained from the RD SET Neo sensor which will be reported as beats per minute.

Countries

United States

Participant flow

Participants by arm

ArmCount
Masimo INVSENSOR00061
Experimental: INVSENSOR00061 All subjects are enrolled into this arm will have noninvasive blood oxygen measurements obtained. Masimo INVSENSOR00061: Masimo INVSENSOR00061 will be used to measure noninvasive SpO2 (Oxygen Saturation )
60
Total60

Baseline characteristics

CharacteristicMasimo INVSENSOR00061
Age, Categorical
<=18 years
60 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
0 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
32 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
27 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
1 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
3 Participants
Race (NIH/OMB)
Black or African American
14 Participants
Race (NIH/OMB)
More than one race
0 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
2 Participants
Race (NIH/OMB)
Unknown or Not Reported
3 Participants
Race (NIH/OMB)
White
38 Participants
Region of Enrollment
United States
60 participants
Sex: Female, Male
Female
25 Participants
Sex: Female, Male
Male
35 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 60
other
Total, other adverse events
0 / 60
serious
Total, serious adverse events
0 / 60

Outcome results

Primary

Pulse Rate Accuracy of INVSENSOR00061

Pulse Rate accuracy will be determined by calculating the Arithmetic Root Mean Square (ARMs) value through the comparison of the noninvasive Pulse Rate to the noninvasive Pulse Rate (PR) value obtained from the RD SET Neo sensor which will be reported as beats per minute.

Time frame: Approximately 15 minutes

Population: 30 subjects were excluded from the analysis and replaced with new participants because of incorrect sensor placement.

ArmMeasureValue (NUMBER)
Masimo INVSENSOR00061Pulse Rate Accuracy of INVSENSOR000614.45 beats per minute
Primary

SpO2 Accuracy of INVSENSOR00061

SpO2 accuracy will be determined by calculating the Arithmetic Root Mean Square (ARMs) value which will be reported as percent of oxygen saturated hemoglobin. The noninvasive oxygen saturation measurement (SpO2) from the INVSENSOR00061 will be compared to the noninvasive oxygen saturation measurement (SpO2) value from the RD SET Neo sensor.

Time frame: Approximately 15 minutes

Population: 30 subjects were excluded from the analysis and replaced with new participants because of incorrect sensor placement.

ArmMeasureValue (NUMBER)
Masimo INVSENSOR00061SpO2 Accuracy of INVSENSOR000610.87 percent of oxygen saturated hemoglobin

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026