Accuracy of Pulse Oximeters During Profound Hypoxia

Accuracy of Pulse Oximeters During Profound Hypoxia

Status
UNKNOWN
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05790616
Acronym
APO
Enrollment
66
Registered
2023-03-30
Start date
2023-01-28
Completion date
2023-11-01
Last updated
2023-03-30

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hypoxia

Keywords

SpO2, photoplethysmography, camera based photoplethysmography

Brief summary

This study is designed to calibrate and determine the accuracy for SpO2, pulse rate and respiratory rate of the newly in-house build Test Device wrist 1 (TDw1, or EVA) at Philips. SpO2, pulse rate and respiratory rate during hypoxia will be calculated by using data of well-known reference devices, including: * A reference SpO2 sensor of Nellcor placed at a fingertip, that reflects also continuously the pulse rate Will be used to compare with the test device. * A reference respiratory rate device that calculates the respiratory rate based on detection of end-tidal CO2 peaks by capnography. * Oxygen saturation in arterial blood samples (SaO2), determined by a co-oximeter will be used to calculate the accuracy of the test device. During the study the following devices will be additionally used by the volunteers: * AppleWatch 7 * TDw2, watch build by philips, using the PPG and software technology developed by Philips * A smartphone that detects reflected PPG signals from the reflected screen at the handpalm, by the build in frontfacing camera (TDc) of the smartphone Volunteers will undergo progressive hypoxia (9 min/% O2) from 21 to 10% O2 in an altitude room, resulting in a volunteer's SpO2 of 73%. During this deliberated hypoxia, the volunteers wear the test and reference devices. This study consists of 4 sub-studies (NI = non-invasive; IN = invasive with an arterial line): * NI (Fast-Sitting): volunteers are seated in the hypoxia room in which the ambient oxygen concentration decreases at a speed of 9 min/% O2. If the volunteer reached a SpO2 ≤73% for more than 1 minute, he/she leaves the hypoxia room. And will breath air with 21% oxygen. Volunteers wear TDw1 and TDw2 and the reference devices. * NI(Fast-Lying): identical to NI(Fast-Sitting) but volunteers lay on a mattress. Volunteers wear TDw1 and AppleWatch 7 and the reference devices. * NI (Slow-Sitting): identical to NI (Fast-Sitting), but after one of the volunteers reaches a SpO2 ≤73% for more than one minute, oxygen in the room increases at a speed of 9 min/% O2 until normal ambient air oxygen concentration of 21%. Volunteers wear TDw1 and AppleWatch 7 and the reference devices. * IN(Fast-Sitting): identical to NI(Fast-Sitting) but the volunteer's oxygen saturation in blood samples withdrawn via an arterial line is measured in the laboratory. The NI studies include 18 healthy participants in each sub-study. After the first volunteers have completed the study, small adaptations in the software of the study devices is still possible, e.g. to increase the quality of the PPG-signals. After the three sub-studies are completed, the algorithm for conversion of raw PPG signals to SpO2, pulse rate and respiratory rate will be defined and fixed for the test devices. During the IN-study, which can only be started after completion of all NI studies, an arterial catheter will be inserted in the radial artery of the 12 participating volunteers, in order to take several blood samples to measure oxygen saturation in the blood (25 samples at well-defined moments during the study per volunteer). Using these results of arterial oxygen saturation, the accuracy of the test devices can be calculated.

Interventions

For the IN(Fast-Sitting) arm arterial blood will be collected to derive Gold-standard SaO2 reference values.

OTHEROxygen reduction

Volunteers undergo an ambient oxygen decrease from 21% to 10%

Sponsors

Philips Clinical & Medical Affairs Global
Lead SponsorINDUSTRY

Study design

Allocation
NON_RANDOMIZED
Intervention model
SINGLE_GROUP
Primary purpose
SCREENING
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 65 Years
Healthy volunteers
Yes

Inclusion criteria

• Healthy subjects (ASA 1 and ASA 2) between 18 and 65 years of age upon signing the informed consent. (ASA: American Society of Anesthesiologist's health score). In each group, at least 30% of one gender need to be present, and at least 15% of participants have a Fitzpatrick's skin color scale of 5 or 6.

Exclusion criteria

* Heavy smokers or individuals exposed to high levels of carbon monoxide that result in elevated carboxyhemoglobin levels. * Individuals subject to conditions that result in elevated levels of methemoglobin. * Individuals with low SpO2 or SaO2 (\<95%) at FiO2 = 0,21 * Moderate to severe claustrophobia for small rooms * Severe obesity (BMI \> 35 kg/m2) * Hemoglobinopathies * Arteriopathies, including Raynaud disease * ASA 3 to 5 subjects * Tattoo's, skin lesions,use of skin lotions or use of nail polish/fake nails at the region of interest for TDw and TDc * Pregnancy

Design outcomes

Primary

MeasureTime frameDescription
Algorithm validation90 minutes per test per participant1. the calculation and validation of the algorithm for 1. conversion of the modulation ratio R of theraw PPG signals to SpO2 for all wrist Test Devices. 2. conversion of PPG signals to PR and RR. 2. determining the accuracy of TDw1 and TDc with cooximetry as golden benchmark in the IN study. 3. Looking for the feasibility of TDw2 for further development The accuracy is represented by A(rms), and specifically for the NI study, the SpO2 range is continuous (SpO2 100- 73%, about 16 000 samples per volunteer). During the IN study the accuracy is represented by A(rms) and calculated for SpO2 versus Sa O2 (about 20 to 25 samples per volunteer).

Secondary

MeasureTime frameDescription
Plateau 190 minutes per test per participantBias and 95% limits of agreement (LOA) of TDw1 SpO2 - Nellcor SpO2 in 90-100% Nellcor SpO2 range
Plateau 390 minutes per test per participantBias and 95% LOA of TDw1 SpO2 - Nellcor SpO2 in 70-80% Nellcor SpO2 range
Plateau 490 minutes per test per participantBias and 95% LOA of TDw1 SpO2 - Nellcor SpO2 in 70-100% Nellcor SpO2 range
Plateau 290 minutes per test per participantBias and 95% LOA of TDw1 SpO2 - Nellcor SpO2 in 80-90% Nellcor SpO2 range
Statistics individual subject90 minutes per test per participantBland-Altman of Bias for each individual subject
Statistics individual subject 290 minutes per test per participantLine listing for each individual subject
Statistics pooled subjects90 minutes per test per participantBland-Altman of pooled subjects' TDw1 SpO2 - Nellcor SpO2