Validation of the Masimo INVSENSOR00063 in Febrile and Afebrile Subejcts

Validation of the Masimo INVSENSOR00063 in Febrile and Afebrile Subejcts

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05787782
Enrollment
60
Registered
2023-03-28
Start date
2023-02-23
Completion date
2023-03-02
Last updated
2025-10-31

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Fever

Brief summary

This is a prospective, non-randomized data collection study to evaluate the performance of the Masimo INVSENSOR00063 in obtaining temperature measurements.

Detailed description

Individual sites are only representative of a subset of the overall population, per ISO 80601-2-56:2017(Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement), and therefore clinical bias and limits of agreement should not be evaluated for the subsets. Publication of results should include repeatability from individual sites and the combined clinical bias and limits of agreement (NCT05787782, NCT05674344 and NCT05779397).

Interventions

DEVICEMasimo INVSENSOR00063

Masimo INVSENSOR00063 will be used to measure temperature

Sponsors

Masimo Corporation
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
1 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* Participant age ≥ 1 year * Afebrile or febrile at time of enrollment: For febrile subjects, ≥ 99.5 F/ 37.5 C for sublingual temperature reference; ≥ 100.4 F/ 38 C for temporal artery temperature reference. * English-or Spanish-speaking patient or parent/Legal Guardian

Exclusion criteria

* Participants deemed not suitable for the study at the discretion of the investigator

Design outcomes

Primary

MeasureTime frameDescription
Accuracy of Temperature Measurements Using Masimo INVSENSOR00063About 15-30 minutesMasimo INVSENSOR00063 temperature measurements will be compared to reference temperature measurements. Accuracy of Masimo INVSENSOR00063 will be measured using clinical bias, limits of agreement and repeatability.

Countries

United States

Participant flow

Participants by arm

ArmCount
Masimo INVSENSOR00063
All subjects are enrolled into this arm and will have temperature measurements obtained. Masimo INVSENSOR00063: Masimo INVSENSOR00063 will be used to measure temperature
60
Total60

Baseline characteristics

CharacteristicMasimo INVSENSOR00063
Age, Categorical
<=18 years
31 Participants
Age, Categorical
>=65 years
1 Participants
Age, Categorical
Between 18 and 65 years
28 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
21 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
39 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
2 Participants
Race (NIH/OMB)
Black or African American
9 Participants
Race (NIH/OMB)
More than one race
5 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
3 Participants
Race (NIH/OMB)
Unknown or Not Reported
3 Participants
Race (NIH/OMB)
White
38 Participants
Sex: Female, Male
Female
33 Participants
Sex: Female, Male
Male
27 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 60
other
Total, other adverse events
0 / 60
serious
Total, serious adverse events
0 / 60

Outcome results

Primary

Accuracy of Temperature Measurements Using Masimo INVSENSOR00063

Masimo INVSENSOR00063 temperature measurements will be compared to reference temperature measurements. Accuracy of Masimo INVSENSOR00063 will be measured using clinical bias, limits of agreement and repeatability.

Time frame: About 15-30 minutes

Population: Data for 1 subject was excluded as they did not complete one round of measurements.~Data for 1 subject was excluded as the data suggested existence of external conditions that interefere with RCT (reference clinical thermometer) or Masimo INVSENSOR00063 measurements.

ArmMeasureGroupValue (NUMBER)
Masimo INVSENSOR00063Accuracy of Temperature Measurements Using Masimo INVSENSOR00063Clinical Bias-0.24 Degrees Celsius
Masimo INVSENSOR00063Accuracy of Temperature Measurements Using Masimo INVSENSOR00063Limits of Agreement0.85 Degrees Celsius
Masimo INVSENSOR00063Accuracy of Temperature Measurements Using Masimo INVSENSOR00063Repeatability0.05 Degrees Celsius