Sternotomy
Conditions
Keywords
Rigid Fixation, Sternal Closure, STERN FIX
Brief summary
The goal of this single arm observational registry is to evaluate the safety and performance of the STERN FIX device in normal conditions of use, in patients treated with medial sternotomy according to the instructions for use. The main goal of this registry is to evaluate the sternal stability in patients who had their sternum fixed with the STERN FIX system. Participants will have their sternotomy closed with the STERN FIX device at the end of their cardiothoracic surgery and will be followed up as per standard of care. At discharge and at the 1 month FU visit the sternal stability will be assessed using the sternal instability scale (SIS).
Interventions
Sternotomy closure with STERN FIX and supplemented with wires
Sponsors
Study design
Eligibility
Inclusion criteria
* Older than 18 years old patient * Patient undergoing surgery requiring medial sternotomy * Patient able and willing to provide informed consent to participate in the registry and/or who have provided written consent to have their clinical data added to the registry
Exclusion criteria
* Patients that are still growing. * Patients with Sternal anomalies preventing the use of the STERN FIX Sternal Stabilization System, such as bone tumors located in the implant region. * Patient with conditions previous to surgery that may impact healing, including bone blood supply limitations, insufficient quantity or quality of bone, or other severe structural bone damage. * Patients with suspected or known allergies or intolerances to the implant material. * Patients with severe osteoporosis or other degenerative bone diseases. * Any patient unwilling to, or incapable of, following postoperative care instructions. * Patients with any other medical or surgical process that may compromise or limit the adequate functionality of the implant. * Pregnant patients or planning to become pregnant during the 2 months following surgery. * Patients with life expectancy lower than 2 months. * Patients who are or have been participating in an interventionist trial during the last 4 weeks Intraoperative
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Sternal stability | One month postoperatively | The primary outcome for this registry will be the sternal stability assessed with the sternal instability scale (SIS). The SIS is a scale ranging from 0 to 3 with the following descriptors: * 0\. Clinically stable sternum (no detectable motion) - normal * 1\. Minimally separated sternum (slight increase in motion upon special testing\* - upper limbs, trunk) * 2\. Partially separated sternum - regional (moderate increase in motion upon special testing\*) * 3\. Completely separated sternum - entire length (marked increase in motion upon special testing\*) |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Prevalence and type of AE and SAE | Through study completion, an average of 6 months | All adverse events occurring during the study period related to the sternotomy must be recorded. The AE that should be recorded, include: * Any device complication (wire o STERN FIX fracture, rotation, migration, …) * Surgical complications: any injury caused by the devices or instruments. * Tissue reactions to implant materials. * Damage to the tissues surrounding the devices. * Osseous complications (fractures, sternal dehiscence, malunion, nonunion, instability, necrosis, calcification, decrease in bone density….). * Postoperative hematoma/ bleeding. * Infectious complications (superficial wound infections, mediastinitis, abscess …). * Increased fibrous tissue, keloid or seroma formation in the sternotomy area. * Any other AE that the investigator considers relevant for the registry. |
| Prevalence of device deficiencies. | Through study completion, an average of 6 months | Adverse event related to the use of an investigational medical device. This includes adverse events resulting from insufficient or inadequate instructions for use, deployment, implantation, installation, or operation, or any malfunction and events resulting from use error or from intentional misuse of the investigational medical device. |
| Sternal closure time | Surgery | Time elapsed from the measurement of the intercostal spaces to the closure of the last intercostal space. |
| Easiness of implantation of the device (assistance) | Immediately after surgery | The surgeon will indicate whether they requested external assistance during the implantation of the device |
| Surgeon satisfaction for the use of the device | Immediately after surgery | The surgeon general satisfaction with the sternotomy closure using STERN FIX will be assessed in comparison with the usual wire closure using the Likert scale (1-5) with the following descriptors: * 5\. Very satisfied * 4\. Satisfied * 3\. Neither satisfied or dissatisfied * 2\. Dissatisfied * 1\. Very dissatisfied |
| Easiness of implantation of the device (Likert) | Immediately after surgery | The easiness score assessed by the surgeon about the usage of the medical device will be registered according to the Likert scale (1-5) with the following descriptors: * 5\. Very easy * 4\. Easy * 3\. Neither easy or difficult * 2\. Difficult * 1\. Very difficult |
Other
| Measure | Time frame | Description |
|---|---|---|
| Performance of the device in high-risk patients (Stability) | One month postoperatively | The performance of the device (stability) in high-risk patients will be assessed in a sub-analysis. Patients considered of high-risk should comply at least one of the following criteria: * Pre-operative criteria: diabetes mellitus, COPD, obesity (BMI ≥30kg/m2), renal failure, chronic usage of steroids, immunosuppression, previous sternotomy, osteoporosis. * Intra-operative criteria: BIMA, cardiopulmonary bypass lasting more than 2 hours. The sternal stability assessed with the sternal instability scale (SIS). The SIS is a scale ranging from 0 to 3 with the following descriptors: * 0 Clinically stable sternum (no detectable motion) - normal * 1 Minimally separated sternum (slight increase in motion upon special testing\* - upper limbs, trunk) * 2 Partially separated sternum - regional (moderate increase in motion upon special testing\*) * 3 Completely separated sternum - entire length (marked increase in motion upon special testing\*) |
| Safety of the device in high-risk patients (AE and SAE prevalence) | Through study completion, an average of 6 months | The safety of the device (prevalence and type of AE and SAE) in high-risk patients will be assessed in a sub-analysis. All adverse events occurring during the study period related to the sternotomy must be recorded. The AE that should be recorded, include: * Any device complication (wire o STERN FIX fracture, rotation, migration, …) * Surgical complications: any injury caused by the devices or instruments. * Tissue reactions to implant materials. * Damage to the tissues surrounding the devices. * Osseous complications (fractures, sternal dehiscence, malunion, nonunion, instability, necrosis, calcification, decrease in bone density….). * Postoperative hematoma/ bleeding. * Infectious complications (superficial wound infections, mediastinitis, abscess …). * Increased fibrous tissue, keloid or seroma formation in the sternotomy area. * Any other AE that the investigator considers relevant for the registry. |