Skip to content

Assessing Accuracy of Kavo Diagnodent

Comparison Between KAVO DIAGNOdent and Digital Radiography in Occlusal Caries Detection (A Diagnostic Accuracy Study)

Status
UNKNOWN
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05777057
Enrollment
42
Registered
2023-03-20
Start date
2022-11-16
Completion date
2023-03-30
Last updated
2023-03-20

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Condition

Brief summary

This study aims to evaluate the accuracy of kavodiagondent and compare with digital radiograph (vista scan).the gold standard is light microscope

Detailed description

Extracted teeth with intial caries (no cavitation)will be used to detect caries. The teeth will be numbered and each of the occlusal surfaces will be measured by DIAGNOdent with a conical probe in accordance with the manufacturer's instructions. The standard value for each individual tooth will be calibrated before each measurement by measuring in a region of sound tissue. The occlusal surfaces will be carefully scanned and measured 3 times when wet, and 3 times after they had been wiped with paper tissues and exposed to air for about 2 min in room temperature. The maximum value from each measurement will be registered and the mean values of the three data under wet and dry conditions, respectively, will be calculated. After an interval of 2 weeks, the procedure will be repeated under identical conditions. All occlusal surfaces will be photographed. The sites with the highest values will be indicated in the photographs in order to identify the location for the second measurement and for subsequent sectioning for light microscope examination. The teeth will be arranged in pairs and mounted in plaster blocks. Reproducible radiographs will be obtained. This will be done by digital radiography (vista scan sensor). The radiographs will be examined by 4 radiologists. The observers will be instructed to select one of five ratings to represent his or her level of confidence that a carious lesion was present or not in the occlusal surfaces of the teeth. The following scale was used: 1 = definitely not caries, 2 = probably not Caries, 3 = questionable, 4 = probably caries, and 5 = definitely caries. After the laser fluorescence measurements had been completed and the digital radiographs had been exposed, the teeth will be removed from the blocks and embedded in plastic. At the sites indicated in the photographs, buccolingual sections approximately 300 µm thick will be taken perpendicular to the occlusal surfaces using a water-cooled saw. The slices will then examined under light microscope. The lesion depth will be scored on a scale from 0 to 4, where 0 = sound, 1 = enamel caries limited to the outer half of the enamel, 2 = enamel caries in the inner half of the enamel, 3 = dentinal caries in the outer half of the dentine, and 4 =dentinal caries in the inner half of the dentine

Interventions

DIAGNOSTIC_TESTDiagnodent

using DIAGNDENT depends on the principle that when 655nm wavelength of Diode laser is irradiated on dental surface, intraoral bacteria metabolites will absorb it and then metabolites emit a red fluorescence. The dental surface will reflect this fluorescence and gives number between 0 and 99 on the screen of the DIAGNdent.

Sponsors

Cairo University
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Masking description

2 radiologist don't know the results of diagnodent and doctor in lab also don't know the the results of diagnodent and radiograph

Intervention model description

Extracted teeth are radiographed and examined by diagnodent and then sections are examined under light microscope

Eligibility

Sex/Gender
ALL
Age
12 Years to 60 Years
Healthy volunteers
Yes

Inclusion criteria

* Extracted posterior teeth Visually intact occlusal surfaces or a variety of occlusal carious lesions, but no obvious cavities * Removal of organic material and stains from the surfaces, the teeth will be immersed in 10% sodium hypochlorite solution for 20 min, followed by rinsing in distilled water for 20 min. The teeth will then stored in saturated thymol solution or formalin. * The teeth will be arranged in pairs and mounted in plaster blocks.

Exclusion criteria

Teeth with deep cavity \-

Design outcomes

Primary

MeasureTime frameDescription
To detect of occlusal caries4 monthsPrimary Outcome Measure: 1.To detect of occlusal caries To detect occlusal caries.42 teeth with initial non cavitated caries will be examined by kavodiagondent and then radiographed. The occlusal surfaces will be carefully scanned and measured 3 times when wet, and 3 times after they had been wiped with paper tissues and exposed to air for about 2 min in room temperature.the teeth will be radiographed.The radiographs will be examined by 4 radiologists. The observers will be instructed to select one of five ratings to represent his or her level of confidence that a carious lesion was present or not in the occlusal surfaces of the teeth. The following scale was used: 1 = definitely not caries, 2 = probably not Caries, 3 = questionable, 4 = probably caries, and 5 = definitely caries. the light microscope will be used as gold standard.

Countries

Egypt

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026