Skip to content

Closed Loop Glucose Control in Patients With Type 1 and Type 2 Diabetes

Demonstration of an Artificial Intelligence Based Closed Loop Glucose Control System as a Therapeutic Modality in Type 1 and Type 2 Diabetic Patients

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05644730
Enrollment
7
Registered
2022-12-09
Start date
2022-11-08
Completion date
2024-06-21
Last updated
2025-07-17

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes Mellitus, Type 1, Diabetes Mellitus, Type 2

Keywords

Closed Loop Glucose Control, Artificial Pancreas, Artificial Intelligence, Glucose, Intensive Care Unit

Brief summary

This is a proof of concept safety study of an artificial intelligence based closed loop glucose control system designed for use in the intensive care unit setting. The type 1 and type 2 diabetic subjects in this study will have their glucose controlled to a range of 100-140 mg/dL by a novel artificial intelligence based closed loop glucose control system for a period of 24 hours. The subjects will consume three standardized meals during the 24 hour study period.

Detailed description

Tight glucose control in the intensive care unit (ICU) setting is difficult to achieve. The investigators hypothesize that a closed loop glucose control system based on artificial intelligence (AI) will improve upon the glucose control currently achieved by the current open loop manual methods, and that this improved glucose control may improve the outcomes of critically ill patients, including those with COVID-19. This Earl Feasibility Study will test the ability of a prototype artificial intelligence based closed loop glucose control system named FUSION, to provide safe and effective glucose control in subjects with type 1 and type 2 diabetes in a clinical research center (CRC) setting. Subjects with type 1 diabetes have been chosen as safe and effective glucose control is difficult to achieve in these subjects during meal challenges. Subjects with type 2 diabetes have been chosen as they are insulin resistant, which makes their insulin resistance profile similar to that of ICU patients. As this is the first in human study of a new medical device, the controlled environment of the CRC is preferable to the less controlled environment of an ICU setting. The prototype FUSION system to be used in this study will consist to two Dexcom G6 continuous glucose monitors (CGM), the AI-based glucose control software run on an all-in-one medical computer, and two syringe pumps. The prototype system is housed on a medical cart. Based on the average glucose value of the two Dexcom G6 CGM's, and the rules of the FUSION systems AI-based glucose control software, the FUSION system will make rate adjustments every 5-10 minutes to the intravenous infusion rates of short acting insulin (NovoLog) and dextrose (D10NS) under its control, in an attempt to keep the subjects glucose in the range of 100-140 mg/dL. The FUSION system only requires entry of the subjects study identification number and weight in kilograms to initiate the system. For safety reasons, the subjects will have their blood glucose independently measured every 10-60 minutes on the point of care Nova StatStrip system, throughout the 24 hour study period. The study has halting criteria to avoid recurrent instances of severe hypoglycemia (\< 54 mg/dL). The average of the two CGM's, that is used by the FUSION system for glucose control, will be used for statistical analysis.

Interventions

The FUSION system will be used to control the subjects glucose to a range of 100-140 mg/dL. Data will be collected for up to 24 hours, or upon early termination of the study session.

Sponsors

Emory University
CollaboratorOTHER
Ideal Medical Technologies
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Intervention model description

Early feasibility safety study

Eligibility

Sex/Gender
ALL
Age
18 Years to 70 Years
Healthy volunteers
No

Inclusion criteria

1. Are 18-70 years of age, inclusive. 2. Can understand and sign an informed consent, communicate with the investigator, and understand and comply with the protocol requirements. 3. Have had a diagnosis of type 1 or type 2 diabetes for a period of at least 1 year. 4. Use insulin injections at home for glucose control and are on a stable insulin regimen without more than a 20% change in their total daily insulin dose during the previous 3 months. Their total daily insulin dose during the previous 3 months will be confirmed through a review of the subject's electronic health record, insulin prescriptions, and insulin pump settings (if applicable). 5. Have a hemoglobin A1c (HbA1c) in the range of 7.0 - 10.0%. 6. Have a hemoglobin in the normal range for sex: 1. Females: 12-15.5 grams/dL. 2. Males: 13.5-17.5 grams/dl. 7. Have adequate venous access sites in upper extremities. 8. Body weight between 40 - 150 kg.

Exclusion criteria

1. Have participated in an interventional medical, surgical, or pharmaceutical study within 30 days of screening. 2. Have a known hypersensitivity to any of the components of study treatment. 3. Have skin disease/injury at Dexcom G6 CGM insertion site(s) that would prevent insertion of the CGM. 4. Currently abuses drugs or alcohol or has a history of abuse that in the investigator's opinion would cause the individual to be noncompliant. 5. Have a medical condition that in the opinion of the investigator could affect study participation and/or personal well-being. 6. Have a clinically significant history or presence of any of the following conditions: 1. Hepatic failure or has alanine aminotransferase (ALT) greater than 3 times the upper limit of normal. 2. Has an estimated glomerular filtration rate (GFR) \<60 ml/min/1.73 m2 or End Stage Kidney Disease on renal replacement therapy. 3. Type 2 diabetic subjects who have a C-peptide level less than 0.2 nmol/L (these subjects will be referred to their primary care doctor or endocrinologist for further work up). 4. Have congestive heart failure of class 1 or greater on the New York Heart Association (NYHA) classification system. 5. Have a history of seizures. 6. Have a history of cerebrovascular accident. 7. Have a history of ischemic heart disease. 7. For female subjects of potential childbearing age (age 18 to 55) they will be excluded if: 1. Pregnant. 2. Refuse to agree to a pregnancy test at the time of enrollment. 3. Have a positive urine pregnancy test at the time of enrollment. 8. Have a positive COVID-19 test within 14 days of visit 3. 9. Have any COVID-19 related symptoms in the 14-day period prior to visit 3. 10. Have a known unprotected COVID-19 exposure in the 14-day period prior to visit 3.

Design outcomes

Primary

MeasureTime frameDescription
Primary Safety OutcomeFrom beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time up to 24 hoursPercent of All Glucose Values Within Glucose Range \< 70 mg/dL
Primary Efficacy OutcomeFrom beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hoursPercent of All Glucose Values Within Glucose Range 70-180 mg/dL

Secondary

MeasureTime frameDescription
Percent of All Glucose Values Within Glucose Range > 180 mg/dLFrom beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hoursPercentage of all glucose values that are greater than 180 mg/dL
Percent of All Glucose Values Within Glucose Range > 250 mg/dLFrom beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hoursThe percentage of all glucose values that are greater than 250 mg/dL
Percent Coefficient of Variation of GlucoseFrom beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hoursThe percent coefficient of variation of glucose will be calculated by dividing the standard deviation of the average glucose value in mg/dL by the average glucose value in mg/dL, then multiplying this value by 100
Weight in KilogramsFrom beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hoursAverage weight in kilograms of the subjects
Average Glucose Value in mg/dLFrom beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hoursThe average of all measured glucose values in mg/dL
Total Daily Dose of Insulin in Units/KilogramAs measured prior to the start of the closed loop glucose control sessionThe subjects total daily dose of insulin used prior to the start of the closed loop glucose control session was calculated by adding up all of the insulin doses used by the subjects (both short and long acting insulin) in units and dividing this value by the subjects weight in Kilograms
Hemoglobin A1c Percentage LevelAs measured within 2 weeks of the start of the closed loop glucose control sessionThe subjects hemoglobin A1c measurement from a blood draw taken within 2 weeks of the start of the closed loop glucose control session
C-Peptide Level in ng/mLAs measured from a blood draw taken within 2 weeks of the start of the closed loop glucose control sessionC-Peptide level as measured from a blood draw
Body Mass IndexAs measured at the start of the closed loop glucose control sessionThe body mass index was calculated by dividing the subjects weight in kilograms by their height in meters squared
Percent of All Glucose Values Within Glucose Range < 54 mg/dLFrom beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hoursThe percentage of all glucose values that less than 54 mg/dL

Countries

United States

Participant flow

Participants by arm

ArmCount
FUSION Closed Loop Glucose Control System Safety Study
All subjects will be treated with the FUSION closed loop glucose control system for up to 24 hours FUSION closed loop glucose control system: The FUSION system will be used to control the subjects glucose to a range of 100-140 mg/dL. Data will be collected for up to 24 hours, or upon early termination of the study session.
7
Total7

Baseline characteristics

CharacteristicFUSION Closed Loop Glucose Control System Safety Study
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
1 Participants
Age, Categorical
Between 18 and 65 years
6 Participants
Age, Continuous52.4 Years
STANDARD_DEVIATION 7.7
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
0 Participants
Race (NIH/OMB)
Black or African American
6 Participants
Race (NIH/OMB)
More than one race
0 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
1 Participants
Race (NIH/OMB)
White
0 Participants
Region of Enrollment
United States
7 participants
Sex: Female, Male
Female
5 Participants
Sex: Female, Male
Male
2 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 7
other
Total, other adverse events
1 / 7
serious
Total, serious adverse events
0 / 7

Outcome results

Primary

Primary Efficacy Outcome

Percent of All Glucose Values Within Glucose Range 70-180 mg/dL

Time frame: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours

ArmMeasureValue (MEAN)Dispersion
FUSION Closed Loop Glucose Control System Safety StudyPrimary Efficacy Outcome79.0 Percentage of all glucose valuesStandard Deviation 21.3
Primary

Primary Safety Outcome

Percent of All Glucose Values Within Glucose Range \< 70 mg/dL

Time frame: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time up to 24 hours

ArmMeasureValue (MEAN)Dispersion
FUSION Closed Loop Glucose Control System Safety StudyPrimary Safety Outcome1.4 Percentage of all glucose valuesStandard Deviation 2.7
Secondary

Average Glucose Value in mg/dL

The average of all measured glucose values in mg/dL

Time frame: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours

ArmMeasureValue (MEAN)Dispersion
FUSION Closed Loop Glucose Control System Safety StudyAverage Glucose Value in mg/dL148.4 mg/dLStandard Deviation 25.1
Secondary

Body Mass Index

The body mass index was calculated by dividing the subjects weight in kilograms by their height in meters squared

Time frame: As measured at the start of the closed loop glucose control session

ArmMeasureValue (MEAN)Dispersion
FUSION Closed Loop Glucose Control System Safety StudyBody Mass Index31.0 Kilograms/Height in meters squaredStandard Deviation 6.1
Secondary

C-Peptide Level in ng/mL

C-Peptide level as measured from a blood draw

Time frame: As measured from a blood draw taken within 2 weeks of the start of the closed loop glucose control session

ArmMeasureValue (MEAN)Dispersion
FUSION Closed Loop Glucose Control System Safety StudyC-Peptide Level in ng/mL1.5 ng/mLStandard Deviation 1.5
Secondary

Hemoglobin A1c Percentage Level

The subjects hemoglobin A1c measurement from a blood draw taken within 2 weeks of the start of the closed loop glucose control session

Time frame: As measured within 2 weeks of the start of the closed loop glucose control session

ArmMeasureValue (MEAN)Dispersion
FUSION Closed Loop Glucose Control System Safety StudyHemoglobin A1c Percentage Level8.5 Percentage of glycated hemoglobinStandard Deviation 0.8
Secondary

Percent Coefficient of Variation of Glucose

The percent coefficient of variation of glucose will be calculated by dividing the standard deviation of the average glucose value in mg/dL by the average glucose value in mg/dL, then multiplying this value by 100

Time frame: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours

ArmMeasureValue (MEAN)Dispersion
FUSION Closed Loop Glucose Control System Safety StudyPercent Coefficient of Variation of Glucose21.3 Percentage of Coefficient of VariationStandard Deviation 5.3
Secondary

Percent of All Glucose Values Within Glucose Range > 180 mg/dL

Percentage of all glucose values that are greater than 180 mg/dL

Time frame: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours

ArmMeasureValue (MEAN)Dispersion
FUSION Closed Loop Glucose Control System Safety StudyPercent of All Glucose Values Within Glucose Range > 180 mg/dL19.6 Percentage of all glucose valuesStandard Deviation 22
Secondary

Percent of All Glucose Values Within Glucose Range > 250 mg/dL

The percentage of all glucose values that are greater than 250 mg/dL

Time frame: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours

ArmMeasureValue (MEAN)Dispersion
FUSION Closed Loop Glucose Control System Safety StudyPercent of All Glucose Values Within Glucose Range > 250 mg/dL2.9 Percentage of all glucose valuesStandard Deviation 4.4
Secondary

Percent of All Glucose Values Within Glucose Range < 54 mg/dL

The percentage of all glucose values that less than 54 mg/dL

Time frame: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours

ArmMeasureValue (MEAN)Dispersion
FUSION Closed Loop Glucose Control System Safety StudyPercent of All Glucose Values Within Glucose Range < 54 mg/dL0 Percentage of all glucose valuesStandard Deviation 0
Secondary

Total Daily Dose of Insulin in Units/Kilogram

The subjects total daily dose of insulin used prior to the start of the closed loop glucose control session was calculated by adding up all of the insulin doses used by the subjects (both short and long acting insulin) in units and dividing this value by the subjects weight in Kilograms

Time frame: As measured prior to the start of the closed loop glucose control session

ArmMeasureValue (MEAN)Dispersion
FUSION Closed Loop Glucose Control System Safety StudyTotal Daily Dose of Insulin in Units/Kilogram0.9 Units Insulin/Kilogram per dayStandard Deviation 0.4
Secondary

Weight in Kilograms

Average weight in kilograms of the subjects

Time frame: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours

ArmMeasureValue (MEAN)Dispersion
FUSION Closed Loop Glucose Control System Safety StudyWeight in Kilograms88.2 KilogramsStandard Deviation 14.7

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026