Owlet Smart Sock 3 Accuracy on Infants

Owlet Smart Sock 3 Accuracy on Infants

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT05635513
Enrollment
28
Registered
2022-12-02
Start date
2021-07-14
Completion date
2021-12-03
Last updated
2023-02-22

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hypoxia Neonatal

Brief summary

The purpose of this research study is to evaluate the Owlet Smart Sock 3 wireless pulse oximeter performance in assessing the heart rate and oxygen saturation in neonatal population, while measuring the same parameter with a standard wired pulse oximeter simultaneously. The monitoring duration with the study device will be a maximum of 60 minutes, after which the recorded data from the Owlet Smart Sock 3 and the standard wired pulse oximeters will be de-identified and analyzed.

Detailed description

This prospective observational study compared the HR and SpO2 measurements of the OSS3 (Owlet, Utah, USA) with the Masimo SET® (Masimo Corporation, California, USA) in infants admitted to the University of Utah NICU. The inclusion criteria were infants who weighed 1.7-2.5 kg at the time of monitoring and were medically stable to tolerate routine care. The exclusion criteria were: infants with significant anomalies, congenital deformities, or medical instability. Infants with an intravenous catheter (IV) and skin lesions on either foot were excluded from the study. The wireless device being tested was the OSS3. The gold standard for validating the pulse oximeter's accuracy is arterial blood oxygen saturation (SaO 2 ) at the pulse oximeter sensor site. \[6\] However, obtaining SaO2 is an invasive arterial blood test for infants. To avoid this conflict, the OSS3 was tested against the reference FDA-approved wired pulse oximeter Masimo. The study was approved by the University of Utah Institutional Review Board, and informed consent was obtained from parents. During the measurement, OSS3 and Masimo were placed on the infants with one device on each foot. Since the data obtained from the study devices can not be recorded in the hospital electronic medical record, the hospital wired pulse oximeter (Philips, Amsterdam, Netherlands) continued to be used on one of the infant's hands during the study period. OSS3 and Masimo were calibrated appropriately against the hospital wired pulse oximeter with matching HR and SpO 2 readings. Pulse oximetry data were collected simultaneously for 60 minutes on the quiet infants. The two devices (OSS3 and Masimo) continuously recorded HR (beats per minute, bpm) and SpO2 (%) every two seconds. Data from the 2 devices were uploaded and compared according to their time stamped measurements. The paired data agreement was compared using Pearson's correlation coefficient (r), Bland-Altman, Average Root Mean Square (ARMS), and Prevalence and Bias Adjusted Kappa (PABAK) analyses.

Interventions

DEVICEOwlet OSS V3

A wearable pulse oximeter designed for monitoring infants and children in a home environment

Sponsors

Owlet Baby Care, Inc.
Lead SponsorINDUSTRY

Study design

Observational model
OTHER
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Healthy volunteers
No

Inclusion criteria

* Infant who weigh between 1.7kg and 2.3kg * Infant who is admitted to U of U NICU

Exclusion criteria

* Infant who has injuries, deformities, abnormalities, or intravenous catheter in the extremities that may prevent proper application of device * Infant who has unstable vital signs or clinical condition * No parental consent * Inability to achieve adequate fit with the OSS3

Design outcomes

Primary

MeasureTime frameDescription
Owlet Pulse Rate Accuracy PerformanceDuring the monitoring period (approx. 1 hour)Conduct a single-arm observational study to monitor heart rate (HR), while a standard wired pulse oximeter (WPOx) will simultaneously evaluate the same parameters in the University of Utah (U of U) Neonatal Intensive Care Unit (NICU).
Owlet Oxygen Saturation Accuracy PerformanceDuring the monitoring period (approx. 1 hour)Conduct a single-arm observational study to monitor oxygen saturation (SpO2) while a standard wired pulse oximeter (WPOx) will simultaneously evaluate the same parameters in the University of Utah (U of U) Neonatal Intensive Care Unit (NICU).