Skip to content

Dexcom G6 for Total Pancreatectomy With Islet Autotransplantation (TPIAT) Patients

Accuracy of Dexcom G6 After Radiation Exposure-A Cross Sectional and Longitudinal Analysis

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05607576
Enrollment
18
Registered
2022-11-07
Start date
2022-03-01
Completion date
2024-05-01
Last updated
2025-05-06

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes

Keywords

Dexcom G6

Brief summary

A single-center prospective trial on the effects of standard of care radiation exposure on the Dexcom G6 continuous glucose monitoring (CGM) device.

Detailed description

To determine if standard of care radiation exposure longitudinally effects the accuracy and precision of the G6 transmitter in the post TPIAT in-patient population. Reduce burden for patients to remove and waste a functional and expensive Dexcom G6 CGM for radiologic procedures.

Interventions

OTHERUncovered CGM during radiologic procedures

Dexcom G6 CGM will not be covered during radiologic procedures.

Sponsors

DexCom, Inc.
CollaboratorINDUSTRY
Deborah Elder
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Masking description

This study is not blinded

Eligibility

Sex/Gender
ALL
Age
11 Years to 25 Years
Healthy volunteers
No

Inclusion criteria

* Planned TPIAT at CCHMC * Staying in-patient at CCHMC post TPIAT * Planned to receive a Dexcom G6 CGM

Exclusion criteria

• Hemo- or peritoneal dialysis

Design outcomes

Primary

MeasureTime frameDescription
Dexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)3 monthsBlood Glucose levels (mg/dL) obtained by the Dexcom G6 sensor and transmitter compared to blood glucose levels (mg/dL) obtained by a handheld glucometer until the transmitter is changed. The Clarke Error Grid Analysis method was used to compare absolute values. Accuracy reported as percentage of comparators in Zone A and B.
Post Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)3 monthsTransmitter functional duration is determined by time from TPIAT procedure and CGM placement (baseline) to time for Transmitter changed (approximately 3 months), Functionality is determined by analysis of Blood Glucose (mg/dL) in month groupings (0-1month, 1-2months, 2+ months from baseline. Blood Glucose levels (mg/dL) obtained by the Dexcom G6 sensor and transmitter compared to blood glucose levels (mg/dL) obtained by a handheld glucometer until the transmitter is changed. The Clarke Error Grid Analysis method was used to compare absolute values. Accuracy reported as percentage of comparators in Zone A and B.

Countries

United States

Participant flow

Participants by arm

ArmCount
Uncovered
Dexcom G6 CGM will be uncovered during standard of care radiologic procedures. Uncovered CGM during radiologic procedures: Dexcom G6 CGM will not be covered during radiologic procedures.
14
Total14

Withdrawals & dropouts

PeriodReasonFG000
Overall Studyimmediate sensor error2
Overall StudyPhysician Decision2

Baseline characteristics

CharacteristicUncovered
Age, Categorical
<=18 years
14 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
0 Participants
BMI percentile79.4 Percentile
Ethnicity (NIH/OMB)
Hispanic or Latino
3 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
10 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
1 Participants
Insurance
Medicaid
5 Participants
Insurance
Private
8 Participants
Insurance
Private and Medicaid
1 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
0 Participants
Race (NIH/OMB)
Black or African American
0 Participants
Race (NIH/OMB)
More than one race
0 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
2 Participants
Race (NIH/OMB)
White
12 Participants
Sex: Female, Male
Female
9 Participants
Sex: Female, Male
Male
5 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 14
other
Total, other adverse events
0 / 14
serious
Total, serious adverse events
0 / 14

Outcome results

Primary

Dexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)

Blood Glucose levels (mg/dL) obtained by the Dexcom G6 sensor and transmitter compared to blood glucose levels (mg/dL) obtained by a handheld glucometer until the transmitter is changed. The Clarke Error Grid Analysis method was used to compare absolute values. Accuracy reported as percentage of comparators in Zone A and B.

Time frame: 3 months

Population: 4168 CGM-glucometer matched data points analyzed, 67% of which were collected from the HG (ACCU-CHEK®). A total of 1375 matched data points were collected from the participants in the outpatient setting using the SIG (Contour®). There were 1435 matched pairs before radiation exposure and 2,733 matched data points after radiation exposure. Absolute blood glucose readings were plotted and analyzed using the Clarke Error Grid Analysis. Data points are matched blood glucose readings.

ArmMeasureGroupValue (COUNT_OF_UNITS)
CGM Uncovered -Pre Radiation ExposuresDexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section A806 Matched blood glucose readings
CGM Uncovered -Pre Radiation ExposuresDexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section B606 Matched blood glucose readings
CGM Uncovered -Pre Radiation ExposuresDexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section C6 Matched blood glucose readings
CGM Uncovered -Pre Radiation ExposuresDexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section D17 Matched blood glucose readings
CGM Uncovered - Post Radiation ExposureDexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section A1226 Matched blood glucose readings
CGM Uncovered - Post Radiation ExposureDexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section D31 Matched blood glucose readings
CGM Uncovered - Post Radiation ExposureDexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section B570 Matched blood glucose readings
CGM Uncovered - Post Radiation ExposureDexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section C7 Matched blood glucose readings
CGM Uncovered - Post Radiation Exposure >500 µGyDexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section D5 Matched blood glucose readings
CGM Uncovered - Post Radiation Exposure >500 µGyDexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section B187 Matched blood glucose readings
CGM Uncovered - Post Radiation Exposure >500 µGyDexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section C5 Matched blood glucose readings
CGM Uncovered - Post Radiation Exposure >500 µGyDexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section A704 Matched blood glucose readings
Handheld Clinical GlucometerDexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section A1715 Matched blood glucose readings
Handheld Clinical GlucometerDexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section B1033 Matched blood glucose readings
Handheld Clinical GlucometerDexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section D47 Matched blood glucose readings
Handheld Clinical GlucometerDexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section C56 Matched blood glucose readings
Handheld Home Glucometer.Dexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section D15 Matched blood glucose readings
Handheld Home Glucometer.Dexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section C69 Matched blood glucose readings
Handheld Home Glucometer.Dexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section B329 Matched blood glucose readings
Handheld Home Glucometer.Dexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)error grid section A1020 Matched blood glucose readings
p-value: <0.05Mixed Models Analysis
Primary

Post Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)

Transmitter functional duration is determined by time from TPIAT procedure and CGM placement (baseline) to time for Transmitter changed (approximately 3 months), Functionality is determined by analysis of Blood Glucose (mg/dL) in month groupings (0-1month, 1-2months, 2+ months from baseline. Blood Glucose levels (mg/dL) obtained by the Dexcom G6 sensor and transmitter compared to blood glucose levels (mg/dL) obtained by a handheld glucometer until the transmitter is changed. The Clarke Error Grid Analysis method was used to compare absolute values. Accuracy reported as percentage of comparators in Zone A and B.

Time frame: 3 months

Population: Data collected on functionality of transmitter over time (blood glucose (BG) mg/dL/months from TPIAT). Data points analyzed using the Clarke Error Grid analysis of matched blood glucose readings.

ArmMeasureGroupValue (COUNT_OF_UNITS)
CGM Uncovered -Pre Radiation ExposuresPost Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)error grid Zone A2065 Matched blood glucose readings
CGM Uncovered -Pre Radiation ExposuresPost Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)error grid Zone B1173 Matched blood glucose readings
CGM Uncovered -Pre Radiation ExposuresPost Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)error grid Zone C16 Matched blood glucose readings
CGM Uncovered -Pre Radiation ExposuresPost Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)error grid Zone D43 Matched blood glucose readings
CGM Uncovered - Post Radiation ExposurePost Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)error grid Zone D10 Matched blood glucose readings
CGM Uncovered - Post Radiation ExposurePost Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)error grid Zone A446 Matched blood glucose readings
CGM Uncovered - Post Radiation ExposurePost Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)error grid Zone C2 Matched blood glucose readings
CGM Uncovered - Post Radiation ExposurePost Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)error grid Zone B128 Matched blood glucose readings
CGM Uncovered - Post Radiation Exposure >500 µGyPost Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)error grid Zone D1 Matched blood glucose readings
CGM Uncovered - Post Radiation Exposure >500 µGyPost Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)error grid Zone B61 Matched blood glucose readings
CGM Uncovered - Post Radiation Exposure >500 µGyPost Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)error grid Zone C0 Matched blood glucose readings
CGM Uncovered - Post Radiation Exposure >500 µGyPost Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)error grid Zone A226 Matched blood glucose readings
p-value: <0.05Mixed Models Analysis

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026