Evaluation of CSF-3 in Performing ECG

A Prospective Study for Evaluating the Safety and Performance of CSF-3 ECG Lead-I, HR and PR When Compared to ECG Holter

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05591157
Acronym
ECG
Enrollment
47
Registered
2022-10-24
Start date
2022-11-16
Completion date
2023-02-01
Last updated
2023-02-10

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

ECG, Healthy

Brief summary

CSF-3 is a wrist-worn medical device similar to a watch. It is designed to monitor a user's heart rate based on ECG (electrocardiogram) and PPG (Photoplethysmograph) readings, using off-the-shelf ECG and unique PPG chipsets with proprietary algorithms. ECG and PPG are the two primary technologies for measuring heart rate. ECG measures the bio-potential generated by electrical signals that control the expansion and contraction of heart chambers, while PPG uses light-based technology to sense the rate of blood flow as controlled by the heart's pumping action.

Detailed description

The proposed clinical study is a prospective, open, single-center study to evaluate the performance and accuracy of the ECG and heart rate calculation of the CSF-3 investigational Device. Subjects will be screened by the Principal Investigator. Subjects who meet the study's enrolment criteria will be recruited and fitted with an FDA cleared ECG Holter, and with the investigational Device (CSF-3). At least 40 eligible subjects will be enrolled in a single center. Data collected from the subjects during the study will be included in the overall performance analysis of the ECG output. Subjects will perform an ECG and CSF-3 Device simultaneously Subjects will remain connected to both the ECG Holter and the CSF-3 Device at the study site for at least 7 minutes duration. The ECG and Holter will be stopped simultaneously. Both recordings data (ECG and CSF-3) will be uploaded and then synchronized offline using a dedicated and validated tool.

Interventions

DEVICECSF-3 watch, designed to monitor a user's heart rate based on ECG (electrocardiogram) and PPG (Photoplethysmograph) readings

The subjects will be connected to an ECG Holter, and an investigational Device will be placed simultaneously on the subject's wrist. All subjects in this phase will be asked to participate in a single continuous data recording measurement for seven (7) minutes

Sponsors

CardiacSense Ltd.
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Intervention model description

clinical study is a prospective, open, single-center study to evaluate the performance and accuracy of the ECG and heart rate calculation of the CSF-3 investigational Device.

Eligibility

Sex/Gender
ALL
Age
21 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* Age ≥ 21 years old * Able and willing to understand and sign the informed consent and follow instructions * Screening ECG results of Sinus Rhythm * Wrist size 14 cm (5.5), Max 20 cm (7.8)

Exclusion criteria

* Subjects who are currently enrolled in other clinical investigation * Subjects diagnosed with a cardiac disorder or any cardiac symptoms * Subjects with SNR \< 100 * Subjects with ECG quality range \> 1 and \< 30 * Individuals with electrical pacing by a pacemaker * Subjects who are pregnant or breastfeeding * Subjects with callous on the index finger or thumb (at fingerprint area) * Subjects with active wound, prior soft tissue injury or skin conditions at the measurement location Sensitivity to the device material or ECG patches * Subjects with tremors or otherwise unable to remain still for 15 minutes * Subjects without two hands and sufficient fingers to complete the study * Subjects unable to participate based upon investigator discretion

Design outcomes

Primary

MeasureTime frameDescription
CSF-3 efficacy in ECG detection6 monthsPerformance endpoint is defined as the ability of CSF-3 to detect beat-by beat Heart Rate in ECG signal with a sensitivity of no less than 96%, considering a false detection rate of not higher than 2%, and PPV of 80%.

Secondary

MeasureTime frameDescription
CSF-3 HR calculation6 monthsPerformance endpoint is defined as the calculation of average Heart Rate with ARMS error\<3BPM

Countries

United States

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026