Comparing NGENUITY vs Conventional Microscope in a New Hydrus User

Comparing NGENUITY vs Conventional Microscope in a New Hydrus User

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05529966
Enrollment
32
Registered
2022-09-07
Start date
2023-02-13
Completion date
2024-03-22
Last updated
2025-04-17

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Glaucoma

Brief summary

To compare the potential efficiency and educational value of using NGENUITY (3D camera system, FDA approved for this indication) vs the conventional microscope for Hydrus microstent placement by new users.

Interventions

DEVICENGENUITY

Device for 3-D viewing during placement of micro-stent placement.

Device for stereoscopic viewing during placement of micro-stent placement.

Sponsors

Alcon Research
CollaboratorINDUSTRY
Stanford University
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
OTHER
Masking
SINGLE (Subject)

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Patients with mild to moderate primary open angle glaucoma as defined by AGS with visually significant age-related cataract undergoing uncomplicated cataract surgery and Hydrus microstent * Either eye or both eyes may participate

Exclusion criteria

* Lack of patient cooperation severely affecting ability to place stent with any visualization technique * Angle abnormalities including PAS in nasal quadrant * Unable to properly visualize the angle due to complicated cataract surgery

Design outcomes

Primary

MeasureTime frame
Time of First Cannula in Until Satisfactory Stent Placement and Cannula Outup to 1 hour

Secondary

MeasureTime frameDescription
Number of Trabecular Meshwork Piercing Attempts Needed to Place Stentup to 1 hour
Number of Anterior Chamber Entry and Reentry Attempts During Stent Placementup to 1 hour
Number of Microscope Focus or Position Readjustments Needed During Stent Placementup to 1 hour
Change in Attending Questionnaire ScoreUp to 5 minutesProviders rate their comfort, visibility, image quality, focus, depth perception, staff engagement, teaching and level of confidence in using the system and system performance. Ratings are on a Likert scale of 1 to 5. Higher scores correspond to excellent ratings, higher confidence in using the system, and better system performance; for Scope Preference only, lower numbers (either group) correspond to a preference for NGenuity visualization.
Change in Fellow Questionnaire ScoreUp to 5 minutesProviders rate their comfort, visibility, image quality, focus, depth perception, staff engagement, teaching and level of confidence in using the system and system performance. Ratings are on a Likert scale of 1 to 5. Higher scores correspond to excellent ratings, higher confidence in using the system, and better system performance; for Scope Preference only, lower numbers (either group) correspond to a preference for NGenuity visualization.

Countries

United States

Participant flow

Pre-assignment details

32 participants signed informed consent, 28 were randomized to either study arm.

Participants by arm

ArmCount
Conventional Microscope
Participants randomized to the use of a conventional microscope for the first eye procedure. Participants who had a procedure for their second eye were randomized again to either the conventional or NGENUITY microscope.
12
NGENUITY
Participants randomized to the use of a NGENUITY microscope for the first eye procedure. Participants who had a procedure for their second eye were randomized again to either the conventional or NGENUITY microscope.
13
Total25

Baseline characteristics

CharacteristicNGENUITYTotalConventional Microscope
Age, Continuous76.1 years
STANDARD_DEVIATION 7.3
75.9 years
STANDARD_DEVIATION 7.5
75.6 years
STANDARD_DEVIATION 8.1
Ethnicity (NIH/OMB)
Hispanic or Latino
1 Participants1 Participants0 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
12 Participants24 Participants12 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants0 Participants0 Participants
Race/Ethnicity, Customized
American Indian
1 Participants1 Participants0 Participants
Race/Ethnicity, Customized
Asian
2 Participants8 Participants6 Participants
Race/Ethnicity, Customized
Black
1 Participants1 Participants0 Participants
Race/Ethnicity, Customized
Other
0 Participants2 Participants2 Participants
Race/Ethnicity, Customized
White
9 Participants13 Participants4 Participants
Region of Enrollment
United States
13 Participants25 Participants12 Participants
Sex: Female, Male
Female
6 Participants13 Participants7 Participants
Sex: Female, Male
Male
7 Participants12 Participants5 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
EG002
affected / at risk
EG003
affected / at risk
deaths
Total, all-cause mortality
0 / 140 / 140 / 10 / 3
other
Total, other adverse events
0 / 140 / 140 / 10 / 3
serious
Total, serious adverse events
0 / 140 / 140 / 10 / 3

Outcome results

Primary

Time of First Cannula in Until Satisfactory Stent Placement and Cannula Out

Time frame: up to 1 hour

Population: Participants/eyes evaluable for primary and secondary outcome measure analysis

ArmMeasureValue (MEAN)Dispersion
Conventional MicroscopeTime of First Cannula in Until Satisfactory Stent Placement and Cannula Out2.55 minutesStandard Deviation 1.36
NGENUITYTime of First Cannula in Until Satisfactory Stent Placement and Cannula Out3.06 minutesStandard Deviation 2.17
p-value: 0.45490734t-test, 2 sided
Secondary

Change in Attending Questionnaire Score

Providers rate their comfort, visibility, image quality, focus, depth perception, staff engagement, teaching and level of confidence in using the system and system performance. Ratings are on a Likert scale of 1 to 5. Higher scores correspond to excellent ratings, higher confidence in using the system, and better system performance; for Scope Preference only, lower numbers (either group) correspond to a preference for NGenuity visualization.

Time frame: Up to 5 minutes

Population: Participants/eyes evaluable for primary and secondary outcome measure analysis

ArmMeasureGroupValue (MEAN)Dispersion
Conventional MicroscopeChange in Attending Questionnaire ScoreComfort3.14 score on a scaleStandard Deviation 0.53
Conventional MicroscopeChange in Attending Questionnaire ScoreVisibility3.57 score on a scaleStandard Deviation 0.85
Conventional MicroscopeChange in Attending Questionnaire ScoreImage Quality3.14 score on a scaleStandard Deviation 0.77
Conventional MicroscopeChange in Attending Questionnaire ScoreFocus3.14 score on a scaleStandard Deviation 0.66
Conventional MicroscopeChange in Attending Questionnaire ScoreDepth Perception2.93 score on a scaleStandard Deviation 0.73
Conventional MicroscopeChange in Attending Questionnaire ScoreStaff Engagement3.36 score on a scaleStandard Deviation 0.5
Conventional MicroscopeChange in Attending Questionnaire ScoreTeaching3.29 score on a scaleStandard Deviation 0.73
Conventional MicroscopeChange in Attending Questionnaire ScoreEn Face Confidence3.43 score on a scaleStandard Deviation 0.51
Conventional MicroscopeChange in Attending Questionnaire ScoreHydrus Placement Confidence3.71 score on a scaleStandard Deviation 0.47
Conventional MicroscopeChange in Attending Questionnaire ScoreScope Preference (1 = 3D.)2.71 score on a scaleStandard Deviation 0.61
NGENUITYChange in Attending Questionnaire ScoreEn Face Confidence4.07 score on a scaleStandard Deviation 0.96
NGENUITYChange in Attending Questionnaire ScoreComfort3.93 score on a scaleStandard Deviation 0.8
NGENUITYChange in Attending Questionnaire ScoreStaff Engagement4.27 score on a scaleStandard Deviation 0.8
NGENUITYChange in Attending Questionnaire ScoreVisibility4.53 score on a scaleStandard Deviation 0.64
NGENUITYChange in Attending Questionnaire ScoreScope Preference (1 = 3D.)1.80 score on a scaleStandard Deviation 0.86
NGENUITYChange in Attending Questionnaire ScoreImage Quality4.6 score on a scaleStandard Deviation 0.74
NGENUITYChange in Attending Questionnaire ScoreTeaching4.67 score on a scaleStandard Deviation 0.62
NGENUITYChange in Attending Questionnaire ScoreFocus4.0 score on a scaleStandard Deviation 1.25
NGENUITYChange in Attending Questionnaire ScoreHydrus Placement Confidence4.20 score on a scaleStandard Deviation 1.08
NGENUITYChange in Attending Questionnaire ScoreDepth Perception4.67 score on a scaleStandard Deviation 0.62
Comparison: Analysis of comfort scorep-value: 0.0042494t-test, 2 sided
Comparison: Analysis of Visibility scorep-value: 0.00223547t-test, 2 sided
Comparison: Analysis of Image Quality scorep-value: 0.00001858t-test, 2 sided
Comparison: Analysis of Focus scorep-value: 0.03003194t-test, 2 sided
Comparison: Analysis of Depth Perception scorep-value: 2.8e-7t-test, 2 sided
Comparison: Analysis of Staff Engagement scorep-value: 0.00112148t-test, 2 sided
Comparison: Analysis of Teaching scorep-value: 951t-test, 2 sided
Comparison: Analysis of En Face Confidence scorep-value: 0.03493564t-test, 2 sided
Comparison: Analysis of Hydrus Placement Confidence scorep-value: 0.12895248t-test, 2 sided
Comparison: Analysis of Scope Preference scorep-value: 0.00279611t-test, 2 sided
Secondary

Change in Fellow Questionnaire Score

Providers rate their comfort, visibility, image quality, focus, depth perception, staff engagement, teaching and level of confidence in using the system and system performance. Ratings are on a Likert scale of 1 to 5. Higher scores correspond to excellent ratings, higher confidence in using the system, and better system performance; for Scope Preference only, lower numbers (either group) correspond to a preference for NGenuity visualization.

Time frame: Up to 5 minutes

Population: Participants/eyes evaluable for primary and secondary outcome measure analysis

ArmMeasureGroupValue (MEAN)Dispersion
Conventional MicroscopeChange in Fellow Questionnaire ScoreImage Quality4.21 score on a scaleStandard Deviation 0.8
Conventional MicroscopeChange in Fellow Questionnaire ScoreStaff Engagement3.57 score on a scaleStandard Deviation 0.94
Conventional MicroscopeChange in Fellow Questionnaire ScoreTeaching4.00 score on a scaleStandard Deviation 0.78
Conventional MicroscopeChange in Fellow Questionnaire ScoreVisibility4.21 score on a scaleStandard Deviation 0.8
Conventional MicroscopeChange in Fellow Questionnaire ScoreEn Face Confidence4.00 score on a scaleStandard Deviation 0.88
Conventional MicroscopeChange in Fellow Questionnaire ScoreFocus3.93 score on a scaleStandard Deviation 0.83
Conventional MicroscopeChange in Fellow Questionnaire ScoreHydrus Placement Confidence4.29 score on a scaleStandard Deviation 0.61
Conventional MicroscopeChange in Fellow Questionnaire ScoreComfort4.21 score on a scaleStandard Deviation 0.89
Conventional MicroscopeChange in Fellow Questionnaire ScoreScope Preference (1 = 3D..)3.00 score on a scaleStandard Deviation 1.11
Conventional MicroscopeChange in Fellow Questionnaire ScoreDepth Perception4.29 score on a scaleStandard Deviation 0.61
NGENUITYChange in Fellow Questionnaire ScoreScope Preference (1 = 3D..)1.93 score on a scaleStandard Deviation 1.1
NGENUITYChange in Fellow Questionnaire ScoreComfort4.20 score on a scaleStandard Deviation 0.56
NGENUITYChange in Fellow Questionnaire ScoreVisibility4.60 score on a scaleStandard Deviation 0.83
NGENUITYChange in Fellow Questionnaire ScoreImage Quality4.60 score on a scaleStandard Deviation 0.83
NGENUITYChange in Fellow Questionnaire ScoreFocus4.40 score on a scaleStandard Deviation 0.91
NGENUITYChange in Fellow Questionnaire ScoreDepth Perception4.73 score on a scaleStandard Deviation 0.59
NGENUITYChange in Fellow Questionnaire ScoreTeaching4.80 score on a scaleStandard Deviation 0.56
NGENUITYChange in Fellow Questionnaire ScoreEn Face Confidence4.33 score on a scaleStandard Deviation 0.82
NGENUITYChange in Fellow Questionnaire ScoreHydrus Placement Confidence4.53 score on a scaleStandard Deviation 0.83
NGENUITYChange in Fellow Questionnaire ScoreStaff Engagement4.73 score on a scaleStandard Deviation 0.59
Comparison: Analysis of Comfort scorep-value: 0.959638658t-test, 2 sided
Comparison: Analysis of Visibility scorep-value: 0.213464715t-test, 2 sided
Comparison: Analysis of Image Quality scorep-value: 0.213464715t-test, 2 sided
Comparison: Analysis of Focus scorep-value: 0.155899777t-test, 2 sided
Comparison: Analysis of Depth Perception scorep-value: 0.055960023t-test, 2 sided
Comparison: Analysis of Staff Engagement scorep-value: 0.000685499t-test, 2 sided
Comparison: Analysis of Teaching scorep-value: 0.004526118t-test, 2 sided
Comparison: Analysis of En Face Confidence scorep-value: 0.299921137t-test, 2 sided
Comparison: Analysis of Hydrus Placement Confidence scorep-value: 0.368082084t-test, 2 sided
Comparison: Analysis of Scope Preference scorep-value: 0.015036417t-test, 2 sided
Secondary

Number of Anterior Chamber Entry and Reentry Attempts During Stent Placement

Time frame: up to 1 hour

Population: Participants/eyes evaluable for primary and secondary outcome measure analysis

ArmMeasureValue (MEAN)Dispersion
Conventional MicroscopeNumber of Anterior Chamber Entry and Reentry Attempts During Stent Placement1.14 Number of Anterior Chamber EntriesStandard Deviation 0.36
NGENUITYNumber of Anterior Chamber Entry and Reentry Attempts During Stent Placement1.13 Number of Anterior Chamber EntriesStandard Deviation 0.35
p-value: 0.94342221t-test, 2 sided
Secondary

Number of Microscope Focus or Position Readjustments Needed During Stent Placement

Time frame: up to 1 hour

Population: Participants/eyes evaluable for primary and secondary outcome measure analysis

ArmMeasureValue (MEAN)Dispersion
Conventional MicroscopeNumber of Microscope Focus or Position Readjustments Needed During Stent Placement1.43 Number of Focus/Microscope adjustmentsStandard Deviation 1.28
NGENUITYNumber of Microscope Focus or Position Readjustments Needed During Stent Placement1.93 Number of Focus/Microscope adjustmentsStandard Deviation 1.79
p-value: 0.38902793t-test, 2 sided
Secondary

Number of Trabecular Meshwork Piercing Attempts Needed to Place Stent

Time frame: up to 1 hour

Population: Participants/eyes evaluable for primary and secondary outcome measure analysis

ArmMeasureValue (MEAN)Dispersion
Conventional MicroscopeNumber of Trabecular Meshwork Piercing Attempts Needed to Place Stent1.50 piercingStandard Deviation 0.94
NGENUITYNumber of Trabecular Meshwork Piercing Attempts Needed to Place Stent1.53 piercingStandard Deviation 0.92
p-value: 0.9237611t-test, 2 sided

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026