Skip to content

Pilot Study of the Endolumik Gastric Calibration Tube for Bariatric Surgery

Pilot Study of the Endolumik Gastric Calibration Tube for Bariatric Surgery

Status
UNKNOWN
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05486325
Enrollment
30
Registered
2022-08-03
Start date
2022-08-31
Completion date
2022-12-31
Last updated
2022-08-03

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Morbid Obesity

Brief summary

The pilot study will be a First In Human Clinical Trial to evaluate the Endolumik Gastric Calibration Tube and associated methods in a human clinical cohort of 20-30 bariatric patients at two high-volume bariatric programs. During this clinical study, the single-use, disposable Endolumik Gastric Calibration Tube will be used during twenty sleeve gastrectomy operations and gastric bypass operations.

Detailed description

Endolumik Inc. has developed a fluorescence-guided calibration device, the Endolumik Gastric Calibration Tube, to improve visualization during bariatric operations. The present pilot study will be a First In Human Clinical Trial to evaluate the Endolumik Gastric Calibration Tube and associated methods in a human clinical cohort of 20-30 bariatric patients at two high-volume bariatric programs. During this clinical study, the single-use, disposable Endolumik Gastric Calibration Tube will be used during twenty sleeve gastrectomy operations and gastric bypass operations. The goal of the present study is to confirm that the Endolumik Gastric Calibration Tube performs similar to predicate devices currently on the market, as well as to get end-user feedback prior to submitting a 510(k) application to the U.S. Food and Drug Administration (FDA). In this multicenter trial, both study sites are certified as Bariatric Centers of Excellence through the American Society of Metabolic and Bariatric Surgery. The first clinical site is Ruby Memorial Hospital, part of West Virginia University; the second clinical site is Durham Regional Hospital, part of Duke University. To date, using funding from a West Virginia Clinical & Translational Science Institute Launch Pilot Grant, our team has validated the device and associated methods of fluorescence-guided sleeve gastrectomy and fluorescence-guided gastric bypass in ex-vivo and in-vivo porcine models, as well as human cadaveric tissue models. Based on FDA criteria, this investigational device meets the definition of a Non-Significant Risk (NSR) device; the risk classification document is included with submission materials. The investigational device provided will have met design controls per FDA code of federal regulation (CFR): 21 CFR 820.30. Endolumik Inc. will serve as the industry sponsor for this study and will donate the devices being used for the protocol to the two participating study sites. Development of the technology in this research is supported by the West Virginia University Department of Surgery.

Interventions

DEVICEEndolumik Gastric Calibration Tube (EGCT)

Up to twenty participants will undergo laparoscopic or robotic sleeve gastrectomy using the EGCT. This operation will take approximately 60 minutes. The portions of the sleeve gastrectomy operation where the EGCT will be used include: * To evacuate gastric contents * During construction of the gastric sleeve * To perform the leak test of the gastric sleeve Up to ten participants will undergo laparoscopic gastric bypass using the EGCT. This operation will take approximately 120 minutes. The portions of the gastric bypass operation where the EGCT will be used include: * To evacuate gastric contents * During construction of the gastric pouch * To calibrate closure of the gastrojejunal anastomosis * To perform a leak test of the gastrojejunal anastomosis In both procedures, The amount of time the EGCT will be inside the patient is approximately 20 minutes. The device is removed prior to the conclusion of the operation, while the patient is under anesthesia.

Sponsors

West Virginia University
CollaboratorOTHER
Endolumik, Inc
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Intervention model description

The sample size (n) will be 20-30 Bariatric patients. Twenty participants will undergo laparoscopic or robotic sleeve gastrectomy using the Endolumik Gastric Calibration Tube instead of the standard bougie calibration tube. Ten participants will undergo laparoscopic gastric bypass using the Endolumik Gastric Calibration Tube instead of the standard calibration tube. The current pilot protocol is a qualitative, not quantitative, study design, thus no power calculation is indicated. The goal of the present pilot study is to serve as the First In Human trial and confirm that the Endolumik Gastric Calibration Tube performs similar to predicate devices currently on the market, as well as to get end-user feedback prior to 510(k) application to the FDA. The sample size was chosen because 20 to 30 uses will allow the development team to appropriately evaluate device performance.

Eligibility

Sex/Gender
ALL
Age
18 Years to 65 Years
Healthy volunteers
No

Inclusion criteria

* English speaking * Adults (between age of 18 and 65 years old) * Have body mass index (BMI) between 35 and 65.99 kg/m2 * Male or female bariatric patients undergoing bariatric surgery at one of the two sites included in the protocol.

Exclusion criteria

* Nonbariatric patients * Patients who have conditions that preclude a sleeve gastrectomy or gastric bypass operation * Patients with esophageal stricture * Patients with esophageal varices * Patients with a Zenker's diverticulum * Patients aged 66 and older * Patients with a BMI of 66 kg/m2 or greater * Pregnant Women * Lactating Women * Other vulnerable patient populations: pregnant and lactating women, prisoners, cognitively impaired and critically ill subjects.

Design outcomes

Primary

MeasureTime frameDescription
Intraoperative adverse eventsDuring surgical treatmentAdverse events occurring during surgical treatment
Operative timeDuring surgical treatmentDuration of surgical treatment
30-day morbidity30 days post surgical treatmentany 30-day Metabolic and Bariatric Surgery Accreditation and Quality Improvement Program (MBSAQIP) Occurrences

Countries

United States

Contacts

Primary ContactMara McFadden
mcfadden@endolumik.com5104684426
Backup ContactNova Szoka, Dr.
szoka@endolumik.com13049067613

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026