Clareon Toric vs Eyhance Toric

Clareon Toric vs Eyhance Toric

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05481125
Enrollment
203
Registered
2022-08-01
Start date
2022-10-10
Completion date
2023-10-16
Last updated
2024-10-30

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Aphakia, Cataract

Keywords

Cataracts, Phacoemulsification, IOL

Brief summary

The primary purpose of this study is to compare the Clareon/Clareon Toric Intraocular Lenses (IOLs) to the Eyhance/Eyhance Toric IOLs in binocular Best Corrected Distance Visual Acuity (BCDVA) at 3 months postoperative.

Detailed description

This study will enroll adults 22 years of age and older diagnosed with cataracts in both eyes with planned bilateral cataract removal by routine small incision phacoemulsification surgery. Subjects will attend up to 7 scheduled visits: A screening visit, two operative visits, and four post-operative visits. The expected individual duration of participation in the study is 3 months.

Interventions

DEVICEClareon IOL

Alcon Clareon Aspheric Hydrophobic Acrylic IOL for the visual correction of aphakia in adult patients in whom a cataractous lens has been removed. The lens is indicated for primary implantation in the capsular bag in the posterior chamber of the eye, replacing the natural crystalline lens. IOLs are implantable medical devices and are intended for long-term use over the lifetime of the pseudophakic subject.

Clareon Toric Aspheric Hydrophobic Acrylic Toric IOL for the visual correction of aphakia with pre-existing corneal astigmatism in adult patients in whom a cataractous lens has been removed. The lens is indicated for primary implantation in the capsular bag in the posterior chamber of the eye, replacing the natural crystalline lens. IOLs are implantable medical devices and are intended for long-term use over the lifetime of the pseudophakic subject.

TECNIS Eyhance IOL with TECNIS Simplicity Delivery System for the visual correction of aphakia in adult patients in whom a cataractous lens has been removed by extracapsular cataract extraction. The lens is intended to be placed in the capsular bag. IOLs are implantable medical devices and are intended for long-term use over the lifetime of the pseudophakic subject.

DEVICEEyhance Toric IOL

TECNIS Eyhance Toric II IOL with TECNIS Simplicity Delivery System for the visual correction of aphakia and pre-existing corneal astigmatism of one diopter or greater in adult patients with or without presbyopia in whom a cataractous lens has been removed by phacoemulsification and who desire reduction in residual refractive cylinder. The lens is intended to be placed in the capsular bag. IOLs are implantable medical devices and are intended for long-term use over the lifetime of the pseudophakic subject.

Cataract removal by routine small incision phacoemulsification surgery

Sponsors

Alcon Research
Lead SponsorINDUSTRY

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
TREATMENT
Masking
DOUBLE (Subject, Investigator)

Eligibility

Sex/Gender
ALL
Age
22 Years to No maximum
Healthy volunteers
No

Inclusion criteria

Key Inclusion Criteria: * Planned bilateral cataract removal by routine small incision phacoemulsification surgery; * Planned implantation of either test or comparator IOL (per randomization) in at least one eye with approximately 1.00 - 3.00 diopter (D) of preoperative corneal astigmatism; * Predicted postoperative astigmatism of ≤ 0.5 D in both operative eyes based on a Toric calculator; * Other protocol-specified inclusion criteria may apply. Key

Exclusion criteria

* Any disease or pathology, other than cataract, that (in the investigator's opinion) may reduce potential Best Corrected Distance Visual Acuity (BCDVA) to a level worse than 0.2 logarithm minimum angle of resolution (logMAR); * Clinically significant (in the investigator's opinion) corneal pathology, dry-eye or ocular surface disease that would adversely affect a) the biometry measures and/or toric calculations and b) the visual outcome; * History of previous intraocular or corneal surgery (including laser-assisted in situ keratomileusis (LASIK)); * Any other planned ocular surgical procedures including but not limited to limbal relaxing incision (LRI), astigmatic keratotomy, LASIK, and retinal laser treatment within the study time frame; * Other protocol-specified

Design outcomes

Primary

MeasureTime frameDescription
Mean Binocular Best-corrected Distance Visual Acuity (BCDVA) at 4 Meters Under Bright Lighting ConditionsMonth 3 postoperative visitVisual acuity was tested at a distance of 4 meters using a refraction chart and the correction obtained from a manifest refraction. BCDVA was measured in logarithm minimum angle of resolution (logMAR) units, with 0.00 logMar equating to 20/20 Snellen, which is considered normal distance eyesight. A lower logMAR value (more negative value) indicates better eyesight. This was a subject based assessment, and subject assigned a single score for both eyes.

Countries

United States

Participant flow

Recruitment details

Subjects were recruited from 8 investigative sites located in the United States.

Pre-assignment details

Of the 203 enrolled, 14 subjects were exited prior to randomization as screen failures, and 14 subjects were discontinued after randomization and prior to attempted implantation. This reporting group includes all subjects with attempted implantation per actual treatment (175).

Participants by arm

ArmCount
Clareon/Clareon Toric
Phacoemulsification surgery, followed by implantation with Alcon Clareon Aspheric Hydrophobic Acrylic IOL or Alcon Clareon Aspheric Hydrophobic Acrylic Toric IOL, as indicated. The second eye surgery was recommended to occur within 14 days after the 1st eye surgery.
89
Eyhance/Eyhance Toric
Phacoemulsification surgery, followed by implantation with TECNIS Eyhance IOL or TECNIS Eyhance Toric II IOL, as indicated. The second eye surgery was recommended to occur within 14 days after the 1st eye surgery.
86
Total175

Withdrawals & dropouts

PeriodReasonFG000FG001
Overall StudyAdverse Event10
Overall StudyLost to Follow-up01
Overall StudyWithdrawal by Subject01

Baseline characteristics

CharacteristicTotalClareon/Clareon ToricEyhance/Eyhance Toric
Age, Continuous69.7 years
STANDARD_DEVIATION 7.69
69.0 years
STANDARD_DEVIATION 7.63
70.4 years
STANDARD_DEVIATION 7.73
Ethnicity (NIH/OMB)
Hispanic or Latino
23 Participants11 Participants12 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
151 Participants77 Participants74 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
1 Participants1 Participants0 Participants
Race/Ethnicity, Customized
American Indian or Alaska Native
0 Participants0 Participants0 Participants
Race/Ethnicity, Customized
Asian
1 Participants1 Participants0 Participants
Race/Ethnicity, Customized
Black or African American
24 Participants15 Participants9 Participants
Race/Ethnicity, Customized
Native Hawaiian or Other Pacific Islander
0 Participants0 Participants0 Participants
Race/Ethnicity, Customized
Other
6 Participants3 Participants3 Participants
Race/Ethnicity, Customized
White
144 Participants70 Participants74 Participants
Region of Enrollment
United States
175 participants89 participants86 participants
Sex: Female, Male
Female
101 Participants47 Participants54 Participants
Sex: Female, Male
Male
74 Participants42 Participants32 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
EG002
affected / at risk
EG003
affected / at risk
EG004
affected / at risk
EG005
affected / at risk
EG006
affected / at risk
deaths
Total, all-cause mortality
0 / 1750 / 890 / 860 / 890 / 860 / 840 / 86
other
Total, other adverse events
0 / 17511 / 898 / 860 / 896 / 867 / 840 / 86
serious
Total, serious adverse events
0 / 1752 / 890 / 862 / 892 / 860 / 843 / 86

Outcome results

Primary

Mean Binocular Best-corrected Distance Visual Acuity (BCDVA) at 4 Meters Under Bright Lighting Conditions

Visual acuity was tested at a distance of 4 meters using a refraction chart and the correction obtained from a manifest refraction. BCDVA was measured in logarithm minimum angle of resolution (logMAR) units, with 0.00 logMar equating to 20/20 Snellen, which is considered normal distance eyesight. A lower logMAR value (more negative value) indicates better eyesight. This was a subject based assessment, and subject assigned a single score for both eyes.

Time frame: Month 3 postoperative visit

Population: AAS with data at visit

ArmMeasureValue (MEAN)Dispersion
Clareon/Clareon ToricMean Binocular Best-corrected Distance Visual Acuity (BCDVA) at 4 Meters Under Bright Lighting Conditions-0.100 logMARStandard Deviation 0.088
Eyhance/Eyhance ToricMean Binocular Best-corrected Distance Visual Acuity (BCDVA) at 4 Meters Under Bright Lighting Conditions-0.055 logMARStandard Deviation 0.0716

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026