Clinical Evaluation of Daily Disposable Etafilcon A Cosmetic Contact Lenses (2022)

Clinical Evaluation of Daily Disposable Etafilcon A Cosmetic Contact Lenses

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05480514
Enrollment
89
Registered
2022-07-29
Start date
2022-08-27
Completion date
2022-11-16
Last updated
2024-01-03

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Visual Acuity

Brief summary

This is a multi-site, 4-visit, brand-masked, bilateral, 2x2 cross-over dispensing study to evaluate visual acuity.

Interventions

DEVICETEST LENS

etafilcon A with cosmetic pattern

Acuvue 1-Day Define Fresh Honey

Sponsors

Johnson & Johnson Vision Care, Inc.
Lead SponsorINDUSTRY

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
TREATMENT
Masking
NONE

Masking description

Brand masking: subjects, investigators, and clinical site personnel will be unaware of the identity (brand) of both study lenses.

Eligibility

Sex/Gender
FEMALE
Age
18 Years to 29 Years
Healthy volunteers
Yes

Inclusion criteria

Potential subjects must satisfy all of the following criteria to be enrolled in the study: 1. The subject must read, understand, and sign the STATEMENT OF INFORMED CONSENT and receive a fully executed copy of the form 2. Females between 18 and 29 (inclusive) years of age at the time of screening 3. Appear able and willing to adhere to the instructions set forth in this clinical protocol (i.e. willing to wear only the study lenses and not use habitual lenses during the dispensing periods) 4. Be a current soft contact lens wearer in both eyes with a minimum of 6 hours/day and 5 days/week over the last 30 days by self-report 5. Be a current wearer of cosmetic/circle lenses in the last 6 months, by self-report 6. The subject must be willing to be photographed and/or video-taped 7. The subject's vertex corrected spherical equivalent distance refraction must be in the range of -1.00 D to -6.00 D (inclusive) in each eye 8. The subject's refractive cylinder must be less than or equal to 0.75 D (inclusive) in each eye 9. Have spherical best corrected visual acuity of 20/25 or better in each eye.

Exclusion criteria

Potential subjects who meet any of the following criteria will be excluded from participating in the study: 1. Currently pregnant or lactating 2. Any systemic disease (eg, Sjögren's Syndrome), allergies, infectious disease (eg, hepatitis, tuberculosis), contagious immunosuppressive diseases (eg, HIV), autoimmune disease (eg, rheumatoid arthritis), or other diseases, by self-report, which are known to interfere with contact lens wear and/or participation in the study (at the investigators discretion) 3. Use of systemic medications (eg, chronic steroid use) that are known to interfere with contact lens wear (at the investigators discretion) 4. Any previous, or planned (during the study) ocular surgery (eg, radial keratotomy, PRK, LASIK, etc.) 5. Any previous history or signs of a contact lens-related corneal inflammatory event (eg, past peripheral ulcer or round peripheral scar), or any other ocular abnormality that would contraindicate contact lens wear (at the investigators discretion). 6. Participation in any contact lens or lens care product clinical trial within seven (7) days prior to study enrollment 7. Employee or family members of clinical site (eg, Investigator, Coordinator, Technician) 8. Any ocular allergies, infections or other ocular abnormalities that are known to interfere with contact lens wear and/or participation in the study. This may include, but not be limited to entropion, ectropion, extrusions, chalazia, recurrent styes, glaucoma, history of recurrent corneal erosions, aphakia, or corneal distortion (at the investigators discretion) 9. Clinically significant (Grade 3 or 4 on FDA scale) tarsal abnormalities, bulbar injection, corneal edema, corneal vascularization, corneal staining, or any other abnormalities of the cornea which would contraindicate contact lens wear.

Design outcomes

Primary

MeasureTime frameDescription
Proportion of Eyes With LogMAR Visual Acuity Less Than 0.176 (20/30 Snellen Acuity)5-minutes post-lens-fittingVisual acuity on logMAR (Logarithm of Minimal Angle of Resolution) scale was evaluated at distance (4 meters) under high luminance low contrast and low luminance high contrast lighting conditions using ETDRS (Early Treatment Diabetic Retinopathy Study) charts. Letter-by-letter results calculated the visual performance score for each chart read. logMAR scores closer to zero, or below zero, indicate a better visual acuity. However, the data was dichotomized as Y=1 if logMAR visual acuity was less than 0.176 logMAR; and Y=0 otherwise. A value of 0.176 logMAR ≅20/30 Snellen Visual Acuity. The proportion of eyes with visual acuity lower than 0.176 logMAR was reported.

Countries

Hong Kong

Participant flow

Recruitment details

A total of 89 subjects were enrolled in this study. Of those enrolled, 77 subjects were dispensed at least one study lens. Of those dispensed, 54 subjects completed the study while 23 subjects were discontinued.

Participants by arm

ArmCount
Dispensed Subjects
All subjects dispensed one study lens.
77
Total77

Withdrawals & dropouts

PeriodReasonFG000FG001
Period 1Covid-19 Related10
Period 1Pd-Incorrect Lens Power Assigned23
Period 1Subject No Longer Meets Eligibility Criteria01
Period 1Unsatisfactory Lens Fitting Due to Test Article35
Period 2Unsatisfactory Lens Fitting Due to Test Article44

Baseline characteristics

CharacteristicDispensed Subjects
Age, Continuous24.4 Years
STANDARD_DEVIATION 2.51
Race/Ethnicity, Customized
Asian
77 Participants
Region of Enrollment
Hong Kong, S.A.R., China
77 Participants
Sex: Female, Male
Female
77 Participants
Sex: Female, Male
Male
0 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 680 / 71
other
Total, other adverse events
0 / 680 / 71
serious
Total, serious adverse events
0 / 680 / 71

Outcome results

Primary

Proportion of Eyes With LogMAR Visual Acuity Less Than 0.176 (20/30 Snellen Acuity)

Visual acuity on logMAR (Logarithm of Minimal Angle of Resolution) scale was evaluated at distance (4 meters) under high luminance low contrast and low luminance high contrast lighting conditions using ETDRS (Early Treatment Diabetic Retinopathy Study) charts. Letter-by-letter results calculated the visual performance score for each chart read. logMAR scores closer to zero, or below zero, indicate a better visual acuity. However, the data was dichotomized as Y=1 if logMAR visual acuity was less than 0.176 logMAR; and Y=0 otherwise. A value of 0.176 logMAR ≅20/30 Snellen Visual Acuity. The proportion of eyes with visual acuity lower than 0.176 logMAR was reported.

Time frame: 5-minutes post-lens-fitting

Population: All subject dispensed at least one study lens.

ArmMeasureValue (NUMBER)
Test [Etafilcon A]Proportion of Eyes With LogMAR Visual Acuity Less Than 0.176 (20/30 Snellen Acuity)1 Proportion of eyes
Control [Etafilcon A]Proportion of Eyes With LogMAR Visual Acuity Less Than 0.176 (20/30 Snellen Acuity)1 Proportion of eyes
95% CI: [0.9445, 0.9926]Bayesian beta-binomial model