Skip to content

Clinical Performance of Masimo INVSENSOR00057 for Heart Rate Measurements

Clinical Performance of Masimo INVSENSOR00057 for Heart Rate Measurements

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05462886
Enrollment
86
Registered
2022-07-18
Start date
2022-07-08
Completion date
2022-07-21
Last updated
2024-09-24

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Healthy

Brief summary

This study is designed to compare the accuracy of a noninvasive measurement of heart rate compared to reference values obtained by a standard of care ECG monitor.

Interventions

Noninvasive wearable health monitoring device

Sponsors

Masimo Corporation
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 80 Years
Healthy volunteers
Yes

Inclusion criteria

* Subject is 18 to 80 years of age. * Subject is able to read and communicate in English.

Exclusion criteria

* Subject has open wounds, inflamed tattoos or piercings on the area of device placement, and/or has any visible healing wounds that the investigator and/or medical professional determines may place them at an increased risk for participation. * Subject has any medical condition which in the judgment of the investigator and/or study staff, renders them ineligible for participation in this study or subject is deemed ineligible by the discretion of the investigator/study staff.

Design outcomes

Primary

MeasureTime frameDescription
Heart Rate Accuracy of Masimo INVSENSOR000571-3 hoursHeart rate accuracy was determined by calculating the accuracy root mean-squared (Arms) difference between the measured values from Masimo INVSENSOR00057 (TestHR\_i) to the reference ECG monitor values (RefHR\_i). Arms= √(∑(i=1 to n) ((TestHR\_i-RefHR\_i )\^2 ))/n

Countries

United States

Participant flow

Participants by arm

ArmCount
Masimo INVSENSOR00057
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Masimo INVSENSOR00057 device. Masimo INVSENSOR00057: noninvasive wearable health monitoring device
79
Total79

Baseline characteristics

CharacteristicMasimo INVSENSOR00057
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
79 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
21 Participants
Race (NIH/OMB)
Black or African American
5 Participants
Race (NIH/OMB)
More than one race
8 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
24 Participants
Race (NIH/OMB)
White
21 Participants
Region of Enrollment
United States
79 participants
Sex: Female, Male
Female
36 Participants
Sex: Female, Male
Male
43 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 86
other
Total, other adverse events
0 / 86
serious
Total, serious adverse events
0 / 86

Outcome results

Primary

Heart Rate Accuracy of Masimo INVSENSOR00057

Heart rate accuracy was determined by calculating the accuracy root mean-squared (Arms) difference between the measured values from Masimo INVSENSOR00057 (TestHR\_i) to the reference ECG monitor values (RefHR\_i). Arms= √(∑(i=1 to n) ((TestHR\_i-RefHR\_i )\^2 ))/n

Time frame: 1-3 hours

Population: 17 subjects not included in analysis due to data collection error.

ArmMeasureValue (NUMBER)
INVSENSOR00057Heart Rate Accuracy of Masimo INVSENSOR000571.38 bpm

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026