Evaluation of Intravascular Monitoring of Partial PRessure of Oxygen for Neonatal intensiVE Care Patients -2

Evaluation of Intravascular Monitoring of Partial PRessure of Oxygen for Neonatal intensiVE Care Patients -2

Status
Terminated
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05433779
Acronym
IMPROVE-2
Enrollment
8
Registered
2022-06-27
Start date
2022-12-15
Completion date
2023-06-16
Last updated
2023-09-21

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Catheter, Lung Diseases, Newborn, Infant, Disease

Keywords

Neonatal, Umbilical Arteries, Oxygen Partial Pressure

Brief summary

The study is a non-blinded trial with one single intervention in sick newborn infants. The Neosense umbilical catheter will be used instead of the routinely used umbilical catheter on infants where the treating physician has ordered use of an umbilical catheter. The catheter will be connected to the Neosense monitor/measuring unit. Blood samples for blood gas analysis are collected from the patients according to the clinical routine (every 4th to 6th hour). Oxygen tension data from the Neosense measuring system, from the time points when a blood gas sample was collected, will be recorded. This data will be compared to the oxygen tension value from the blood gas sample collected at the same time point. The blood gas samples will be analysed according to clinical and laboratory routine. The infants will remain in the study as long as the Neosense measuring system is used

Interventions

The Neosense umbilical catheter is used instead of the routinely used umbilical catheter

Sponsors

Neosense Technologies
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Healthy volunteers
No

Inclusion criteria

Sick newborn infants where the clinical routine indicates use of an umbilical catheter. At least one of the criteria below should be fulfilled: * The infant needs invasive measurement of blood pressure. * The infant needs repetitive sampling of blood. * The infant needs prolonged infusion(s) \> 2 days. * The infant needs infusion of vessel irritating and potentially vessel harming solutions. * The infant is born extremely preterm (before 28+0 weeks gestational age). * A newborn infant with severe respiratory disorder, requiring oxygen treatment (more than 40% Fraction of inspired oxygen (FiO2)). * The infant is undergoing therapeutic hypothermia following asphyxia (oxygen deficiency at birth). * The infant has a severe infection/sepsis. Also: • Signed informed consent form by both parents (or guardians), (if the infant has only one parent (or guardian) one informed consent is required).

Exclusion criteria

* Gastroschisis * Omphalocele * Peritonitis * Necrotizing enterocolitis * Omphalitis * The infant has a severe infection/sepsis * Expected MRI investigation within the 7 first days after birth * The infants birth weight is below 750 g * The infant is born before week 25+0 * Local vascular compromise in lower limbs or buttocks area, or portal venous hypertension

Design outcomes

Primary

MeasureTime frameDescription
Measurement performance (continuously) of an intravascular oxygen sensor in the Neosense systemMeasurement comparisons done at every blood gas sample, i.e. every 4th - 6th hour during the study (up to 8 days)The oxygen tension measured by the Neosense system, compared to the oxygen tension measured by a blood gas analyser.

Secondary

MeasureTime frameDescription
Clinical complications with the Neosense catheter.Time Frame: During study, up to 8 daysThe number of occasions when the Neosense catheter had to be removed and the underlying reason. This includes all anticipated device effects as well as placing the catheter in the wrong vessel, impossible to measure blood pressure, collect blood samples and/or administer fluids.
Any deviations from normal procedures when inserting an umbilical catheterDuring day 1Number of attempts to place the catheter for each patient.
Number of monitor lock up or battery depletionsDuring study, in average 4-5 daysTimepoint for monitor lock up or battery depletion, in case of occurrence
Adverse device effects of the Neosense monitor.During study, in average 4-5 daysOccurence of anticipated adverse device effects for the monitor.
Any deviations from normal procedures when inserting an umbilical catheter.During day 1Any accessories needed to introduce the Neosense catheter.

Countries

Sweden

Outcome results

None listed