Biliary Disease, Biliary Obstruction, Biliary Tract Neoplasms, Choledocholithiasis, Pancreatic Disease, Pancreatic Neoplasms
Conditions
Keywords
ERCP, Duodenoscope, Duodenoscope with single-use distal cover
Brief summary
The purpose of the study is to compare the success rates of procedure success and complication rates in patients undergoing Endoscopic retrograde cholangiopancreatography (ERCP) using two types of currently available endoscopes. These are (1) duodenoscope with a Single-use distal cover or (2) a conventional reusable duodenoscope.
Detailed description
The risks associated with the ERCP procedure will be explained to patients who meets the eligibility criteria. Consenting patients will be randomized to undergo Endoscopic retrograde cholangiopancreatography (ERCP) using either (1) a duodenoscope with a Single-use distal cover or (2) a conventional reusable duodenoscope. Assessments will be made during, and immediately after the procedure, then patients will be called after 1 week, 1 month, and 3 months after the procedure to get information about adverse events.
Interventions
Newer duodenoscope with a single-use distal cover
Conventional duodenoscope with standard methods of reprocessing
Sponsors
Study design
Eligibility
Inclusion criteria
* Patients undergoing ERCP for biliary and pancreatic conditions * Age \> 18 years old
Exclusion criteria
* Vulnerable patient groups including pregnant women * Patients who are unable to consent * Patients with known cholangitis * Patients with known Carbapenem-Resistant Enterobacteriaceae (CRE) or any other multi-drug resistant organisms (MDRO) infection * Surgically altered anatomy except for Billroth I
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Number of Participants With Procedure-related Adverse Events | 3 months | Patients undergoing Endoscopic retrograde cholangiopancreatography (ERCP) using either of the 2 duodenoscopes were assessed for superficial mucosal injury, perforation of the stomach or intestine, mucosal bleeding, pancreatitis, infection, death or any other symptom reported by the patient. |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Duodenoscope Maneuverability: Passage Into Stomach | Maneuverability will be assessed only during the procedure (approximately 10 minutes) | Providers assessed the 2 duodenoscopes for maneuverability during the procedure in patients undergoing ERCP, using a previously defined scoring system which uses a 5-point scale ranging from 1 to 5, with 1 being easy to perform, 2 minimal difficulty, 3 moderate difficulty, 4 severe difficulty and 5 unable to complete the requisite maneuver. |
| Duodenoscope Maneuverability: Navigation Across Pylorus | Maneuverability will be assessed only during the procedure (approximately 10 minutes) | Providers assessed the 2 duodenoscopes for maneuverability during the procedure in patients undergoing ERCP, using a previously defined scoring system which uses a 5-point scale ranging from 1 to 5, with 1 being easy to perform, 2 minimal difficulty, 3 moderate difficulty, 4 severe difficulty and 5 unable to complete the requisite maneuver. |
| Duodenoscope Maneuverability: Achieving Short Position of the Scope | Maneuverability will be assessed only during the procedure (approximately 10 minutes) | Providers assessed the 2 duodenoscopes for maneuverability during the procedure in patients undergoing ERCP, using a previously defined scoring system which uses a 5-point scale ranging from 1 to 5, with 1 being easy to perform, 2 minimal difficulty, 3 moderate difficulty, 4 severe difficulty and 5 unable to complete the requisite maneuver. |
| Duodenoscope Maneuverability: Achieving Optimal Papillary Position | Maneuverability will be assessed only during the procedure (approximately 10 minutes) | Providers assessed the 2 duodenoscopes for maneuverability during the procedure in patients undergoing ERCP, using a previously defined scoring system which uses a 5-point scale ranging from 1 to 5, with 1 being easy to perform, 2 minimal difficulty, 3 moderate difficulty, 4 severe difficulty and 5 unable to complete the requisite maneuver. |
| Duodenoscope Mechanical Characteristics Profile: Scope Stiffness | Mechanical characteristics assessed only during the procedure (approximately 15 minutes) | Providers assessed the 2 duodenoscopes for their mechanical characteristics during the procedure in patients undergoing ERCP. Each characteristic was assessed using a published 5-point scoring system, with 1 being easy to perform, 2 minimal difficulty, 3 moderate difficulty, 4 severe difficulty and 5 unable to complete the requisite maneuver. |
| Duodenoscope Maneuverability: Intubation Into Esophagus | Maneuverability will be assessed only during the procedure (approximately 10 minutes) | Providers assessed the 2 duodenoscopes for maneuverability during the procedure in patients undergoing ERCP, using a previously defined scoring system which uses a 5-point scale ranging from 1 to 5, with 1 being easy to perform, 2 minimal difficulty, 3 moderate difficulty, 4 severe difficulty and 5 unable to complete the requisite maneuver. |
| Duodenoscope Mechanical Characteristics Profile: Elevator Efficiency | Mechanical characteristics assessed only during the procedure (approximately 15 minutes) | Providers assessed the 2 duodenoscopes for their mechanical characteristics during the procedure in patients undergoing ERCP. Each characteristic was assessed using a published 5-point scoring system, with 1 being easy to perform, 2 minimal difficulty, 3 moderate difficulty, 4 severe difficulty and 5 unable to complete the requisite maneuver. |
| Duodenoscope Mechanical Characteristics Profile: Hand Strain | Mechanical characteristics assessed only during the procedure (approximately 15 minutes) | Providers assessed the 2 duodenoscopes for their mechanical characteristics during the procedure in patients undergoing ERCP. Each characteristic was assessed using a published 5-point scoring system, with 1 being no strain, 2 minimal strain, 3 moderate strain, 4 severe strain and 5 unable to complete the procedure. |
| Duodenoscope Imaging Characteristics Profile: Image Quality | Imaging characteristics were assessed only during the procedure (approximately 15 minutes) | Providers assessed the 2 duodenoscopes for their imaging characteristics during the procedure in patients undergoing ERCP. Each characteristic was assessed using a published 5-point scoring system, with 1 being superior, 2 good, 3 satisfactory, 4 suboptimal and 5 unable to visualize or achieve stability of image resulting in termination of procedure. |
| Duodenoscope Imaging Characteristics Profile: Image Stability | Imaging characteristics were assessed only during the procedure (approximately 15 minutes) | Providers assessed the 2 duodenoscopes for their imaging characteristics during the procedure in patients undergoing ERCP. Each characteristic was assessed using a published 5-point scoring system, with 1 being superior, 2 good, 3 satisfactory, 4 suboptimal and 5 unable to visualize or achieve stability of image resulting in termination of procedure. |
| Number of Participant With Procure Success (Ability to Perform Interventions) | 30-60 minutes | Patients undergoing Endoscopic retrograde cholangiopancreatography (ERCP) using either of the 2 duodenoscopes will be assessed for their ability to perform endoscopic maneuvers like sphincterotomy, balloon sweep, use of basket/mechanical lithotriptor, stone clearance, stricture dilation using balloon/catheter, stent insertion, stent removal and use of single operator cholangioscopy/pancreatoscopy. |
| Duodenoscope Mechanical Characteristics Profile: Air-water Button Functionality | Mechanical characteristics assessed only during the procedure (approximately 15 minutes) | Providers assessed the 2 duodenoscopes for their mechanical characteristics during the procedure in patients undergoing ERCP. Each characteristic was assessed using a published 5-point scoring system, with 1 being no water leakage or no difficulty in applying suction or inflating air, 2 minimal leakage or minimal difficulty with suction or inflation, 3 moderate leakage or difficulty, 4 severe leakage or difficulty and 5 unable to perform the requisite function. |
Countries
United States
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| Duodenoscope With Disposable Distal End Caps Patients will undergo ERCP using a duodenoscope with a single-use distal cover. | 159 |
| Conventional Reusable Duodenoscope Patients will undergo ERCP using a conventional duodenoscope. | 160 |
| Total | 319 |
Baseline characteristics
| Characteristic | Conventional Reusable Duodenoscope | Total | Duodenoscope With Disposable Distal End Caps |
|---|---|---|---|
| Age, Continuous | 65 years | 65 years | 64.5 years |
| Previous ERCP Index | 44 Participants | 89 Participants | 45 Participants |
| Previous ERCP Repeat | 116 Participants | 230 Participants | 114 Participants |
| Race (NIH/OMB) American Indian or Alaska Native | 0 Participants | 0 Participants | 0 Participants |
| Race (NIH/OMB) Asian | 48 Participants | 89 Participants | 41 Participants |
| Race (NIH/OMB) Black or African American | 33 Participants | 60 Participants | 27 Participants |
| Race (NIH/OMB) More than one race | 0 Participants | 0 Participants | 0 Participants |
| Race (NIH/OMB) Native Hawaiian or Other Pacific Islander | 0 Participants | 0 Participants | 0 Participants |
| Race (NIH/OMB) Unknown or Not Reported | 17 Participants | 33 Participants | 16 Participants |
| Race (NIH/OMB) White | 62 Participants | 137 Participants | 75 Participants |
| Region of Enrollment United States | 160 Participants | 319 Participants | 159 Participants |
| Sex: Female, Male Female | 74 Participants | 130 Participants | 56 Participants |
| Sex: Female, Male Male | 86 Participants | 189 Participants | 103 Participants |
Adverse events
| Event type | EG000 affected / at risk | EG001 affected / at risk |
|---|---|---|
| deaths Total, all-cause mortality | 1 / 159 | 0 / 160 |
| other Total, other adverse events | 8 / 159 | 8 / 160 |
| serious Total, serious adverse events | 8 / 159 | 9 / 160 |
Outcome results
Number of Participants With Procedure-related Adverse Events
Patients undergoing Endoscopic retrograde cholangiopancreatography (ERCP) using either of the 2 duodenoscopes were assessed for superficial mucosal injury, perforation of the stomach or intestine, mucosal bleeding, pancreatitis, infection, death or any other symptom reported by the patient.
Time frame: 3 months
| Arm | Measure | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|
| Duodenoscope With Disposable Distal End Caps | Number of Participants With Procedure-related Adverse Events | 8 Participants |
| Conventional Reusable Duodenoscope | Number of Participants With Procedure-related Adverse Events | 9 Participants |
Duodenoscope Imaging Characteristics Profile: Image Quality
Providers assessed the 2 duodenoscopes for their imaging characteristics during the procedure in patients undergoing ERCP. Each characteristic was assessed using a published 5-point scoring system, with 1 being superior, 2 good, 3 satisfactory, 4 suboptimal and 5 unable to visualize or achieve stability of image resulting in termination of procedure.
Time frame: Imaging characteristics were assessed only during the procedure (approximately 15 minutes)
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Imaging Characteristics Profile: Image Quality | Rating = 1 | 154 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Imaging Characteristics Profile: Image Quality | Rating = 3 | 1 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Imaging Characteristics Profile: Image Quality | Rating = 5 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Imaging Characteristics Profile: Image Quality | Rating = 4 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Imaging Characteristics Profile: Image Quality | Rating = 2 | 4 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Imaging Characteristics Profile: Image Quality | Rating = 5 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Imaging Characteristics Profile: Image Quality | Rating = 1 | 154 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Imaging Characteristics Profile: Image Quality | Rating = 2 | 4 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Imaging Characteristics Profile: Image Quality | Rating = 3 | 2 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Imaging Characteristics Profile: Image Quality | Rating = 4 | 0 Participants |
Duodenoscope Imaging Characteristics Profile: Image Stability
Providers assessed the 2 duodenoscopes for their imaging characteristics during the procedure in patients undergoing ERCP. Each characteristic was assessed using a published 5-point scoring system, with 1 being superior, 2 good, 3 satisfactory, 4 suboptimal and 5 unable to visualize or achieve stability of image resulting in termination of procedure.
Time frame: Imaging characteristics were assessed only during the procedure (approximately 15 minutes)
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Imaging Characteristics Profile: Image Stability | Rating = 2 | 3 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Imaging Characteristics Profile: Image Stability | Rating = 4 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Imaging Characteristics Profile: Image Stability | Rating = 3 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Imaging Characteristics Profile: Image Stability | Rating = 5 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Imaging Characteristics Profile: Image Stability | Rating = 1 | 156 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Imaging Characteristics Profile: Image Stability | Rating = 5 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Imaging Characteristics Profile: Image Stability | Rating = 1 | 157 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Imaging Characteristics Profile: Image Stability | Rating = 2 | 3 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Imaging Characteristics Profile: Image Stability | Rating = 3 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Imaging Characteristics Profile: Image Stability | Rating = 4 | 0 Participants |
Duodenoscope Maneuverability: Achieving Optimal Papillary Position
Providers assessed the 2 duodenoscopes for maneuverability during the procedure in patients undergoing ERCP, using a previously defined scoring system which uses a 5-point scale ranging from 1 to 5, with 1 being easy to perform, 2 minimal difficulty, 3 moderate difficulty, 4 severe difficulty and 5 unable to complete the requisite maneuver.
Time frame: Maneuverability will be assessed only during the procedure (approximately 10 minutes)
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Achieving Optimal Papillary Position | Rating = 2 | 4 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Achieving Optimal Papillary Position | Rating = 4 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Achieving Optimal Papillary Position | Rating = 3 | 1 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Achieving Optimal Papillary Position | Rating = 5 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Achieving Optimal Papillary Position | Rating = 1 | 154 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Achieving Optimal Papillary Position | Rating = 5 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Achieving Optimal Papillary Position | Rating = 1 | 154 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Achieving Optimal Papillary Position | Rating = 2 | 4 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Achieving Optimal Papillary Position | Rating = 3 | 2 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Achieving Optimal Papillary Position | Rating = 4 | 0 Participants |
Duodenoscope Maneuverability: Achieving Short Position of the Scope
Providers assessed the 2 duodenoscopes for maneuverability during the procedure in patients undergoing ERCP, using a previously defined scoring system which uses a 5-point scale ranging from 1 to 5, with 1 being easy to perform, 2 minimal difficulty, 3 moderate difficulty, 4 severe difficulty and 5 unable to complete the requisite maneuver.
Time frame: Maneuverability will be assessed only during the procedure (approximately 10 minutes)
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Achieving Short Position of the Scope | Rating = 2 | 3 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Achieving Short Position of the Scope | Rating = 4 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Achieving Short Position of the Scope | Rating = 3 | 2 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Achieving Short Position of the Scope | Rating = 5 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Achieving Short Position of the Scope | Rating = 1 | 154 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Achieving Short Position of the Scope | Rating = 5 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Achieving Short Position of the Scope | Rating = 1 | 155 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Achieving Short Position of the Scope | Rating = 2 | 4 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Achieving Short Position of the Scope | Rating = 3 | 1 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Achieving Short Position of the Scope | Rating = 4 | 0 Participants |
Duodenoscope Maneuverability: Intubation Into Esophagus
Providers assessed the 2 duodenoscopes for maneuverability during the procedure in patients undergoing ERCP, using a previously defined scoring system which uses a 5-point scale ranging from 1 to 5, with 1 being easy to perform, 2 minimal difficulty, 3 moderate difficulty, 4 severe difficulty and 5 unable to complete the requisite maneuver.
Time frame: Maneuverability will be assessed only during the procedure (approximately 10 minutes)
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Intubation Into Esophagus | Rating = 2 | 5 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Intubation Into Esophagus | Rating = 4 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Intubation Into Esophagus | Rating = 3 | 1 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Intubation Into Esophagus | Rating = 5 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Intubation Into Esophagus | Rating = 1 | 153 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Intubation Into Esophagus | Rating = 5 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Intubation Into Esophagus | Rating = 1 | 154 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Intubation Into Esophagus | Rating = 2 | 5 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Intubation Into Esophagus | Rating = 3 | 1 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Intubation Into Esophagus | Rating = 4 | 0 Participants |
Duodenoscope Maneuverability: Navigation Across Pylorus
Providers assessed the 2 duodenoscopes for maneuverability during the procedure in patients undergoing ERCP, using a previously defined scoring system which uses a 5-point scale ranging from 1 to 5, with 1 being easy to perform, 2 minimal difficulty, 3 moderate difficulty, 4 severe difficulty and 5 unable to complete the requisite maneuver.
Time frame: Maneuverability will be assessed only during the procedure (approximately 10 minutes)
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Navigation Across Pylorus | Rating = 2 | 3 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Navigation Across Pylorus | Rating = 4 | 2 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Navigation Across Pylorus | Rating = 3 | 1 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Navigation Across Pylorus | Rating = 5 | 2 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Navigation Across Pylorus | Rating = 1 | 151 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Navigation Across Pylorus | Rating = 5 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Navigation Across Pylorus | Rating = 1 | 155 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Navigation Across Pylorus | Rating = 2 | 4 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Navigation Across Pylorus | Rating = 3 | 1 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Navigation Across Pylorus | Rating = 4 | 0 Participants |
Duodenoscope Maneuverability: Passage Into Stomach
Providers assessed the 2 duodenoscopes for maneuverability during the procedure in patients undergoing ERCP, using a previously defined scoring system which uses a 5-point scale ranging from 1 to 5, with 1 being easy to perform, 2 minimal difficulty, 3 moderate difficulty, 4 severe difficulty and 5 unable to complete the requisite maneuver.
Time frame: Maneuverability will be assessed only during the procedure (approximately 10 minutes)
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Passage Into Stomach | Rating = 2 | 3 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Passage Into Stomach | Rating = 4 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Passage Into Stomach | Rating = 3 | 3 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Passage Into Stomach | Rating = 5 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Maneuverability: Passage Into Stomach | Rating = 1 | 153 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Passage Into Stomach | Rating = 5 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Passage Into Stomach | Rating = 1 | 155 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Passage Into Stomach | Rating = 2 | 2 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Passage Into Stomach | Rating = 3 | 2 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Maneuverability: Passage Into Stomach | Rating = 4 | 1 Participants |
Duodenoscope Mechanical Characteristics Profile: Air-water Button Functionality
Providers assessed the 2 duodenoscopes for their mechanical characteristics during the procedure in patients undergoing ERCP. Each characteristic was assessed using a published 5-point scoring system, with 1 being no water leakage or no difficulty in applying suction or inflating air, 2 minimal leakage or minimal difficulty with suction or inflation, 3 moderate leakage or difficulty, 4 severe leakage or difficulty and 5 unable to perform the requisite function.
Time frame: Mechanical characteristics assessed only during the procedure (approximately 15 minutes)
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Air-water Button Functionality | Rating = 2 | 3 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Air-water Button Functionality | Rating = 4 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Air-water Button Functionality | Rating = 3 | 1 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Air-water Button Functionality | Rating = 5 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Air-water Button Functionality | Rating = 1 | 155 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Air-water Button Functionality | Rating = 5 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Air-water Button Functionality | Rating = 1 | 156 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Air-water Button Functionality | Rating = 2 | 4 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Air-water Button Functionality | Rating = 3 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Air-water Button Functionality | Rating = 4 | 0 Participants |
Duodenoscope Mechanical Characteristics Profile: Elevator Efficiency
Providers assessed the 2 duodenoscopes for their mechanical characteristics during the procedure in patients undergoing ERCP. Each characteristic was assessed using a published 5-point scoring system, with 1 being easy to perform, 2 minimal difficulty, 3 moderate difficulty, 4 severe difficulty and 5 unable to complete the requisite maneuver.
Time frame: Mechanical characteristics assessed only during the procedure (approximately 15 minutes)
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Elevator Efficiency | Rating = 2 | 1 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Elevator Efficiency | Rating = 4 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Elevator Efficiency | Rating = 3 | 1 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Elevator Efficiency | Rating = 5 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Elevator Efficiency | Rating = 1 | 157 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Elevator Efficiency | Rating = 5 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Elevator Efficiency | Rating = 1 | 158 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Elevator Efficiency | Rating = 2 | 2 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Elevator Efficiency | Rating = 3 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Elevator Efficiency | Rating = 4 | 0 Participants |
Duodenoscope Mechanical Characteristics Profile: Hand Strain
Providers assessed the 2 duodenoscopes for their mechanical characteristics during the procedure in patients undergoing ERCP. Each characteristic was assessed using a published 5-point scoring system, with 1 being no strain, 2 minimal strain, 3 moderate strain, 4 severe strain and 5 unable to complete the procedure.
Time frame: Mechanical characteristics assessed only during the procedure (approximately 15 minutes)
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Hand Strain | Rating = 2 | 2 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Hand Strain | Rating = 4 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Hand Strain | Rating = 3 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Hand Strain | Rating = 5 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Hand Strain | Rating = 1 | 157 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Hand Strain | Rating = 5 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Hand Strain | Rating = 1 | 158 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Hand Strain | Rating = 2 | 2 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Hand Strain | Rating = 3 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Hand Strain | Rating = 4 | 0 Participants |
Duodenoscope Mechanical Characteristics Profile: Scope Stiffness
Providers assessed the 2 duodenoscopes for their mechanical characteristics during the procedure in patients undergoing ERCP. Each characteristic was assessed using a published 5-point scoring system, with 1 being easy to perform, 2 minimal difficulty, 3 moderate difficulty, 4 severe difficulty and 5 unable to complete the requisite maneuver.
Time frame: Mechanical characteristics assessed only during the procedure (approximately 15 minutes)
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Scope Stiffness | Rating = 5 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Scope Stiffness | Rating = 1 | 155 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Scope Stiffness | Rating = 2 | 4 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Scope Stiffness | Rating = 3 | 0 Participants |
| Duodenoscope With Disposable Distal End Caps | Duodenoscope Mechanical Characteristics Profile: Scope Stiffness | Rating = 4 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Scope Stiffness | Rating = 4 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Scope Stiffness | Rating = 3 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Scope Stiffness | Rating = 1 | 155 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Scope Stiffness | Rating = 5 | 0 Participants |
| Conventional Reusable Duodenoscope | Duodenoscope Mechanical Characteristics Profile: Scope Stiffness | Rating = 2 | 5 Participants |
Number of Participant With Procure Success (Ability to Perform Interventions)
Patients undergoing Endoscopic retrograde cholangiopancreatography (ERCP) using either of the 2 duodenoscopes will be assessed for their ability to perform endoscopic maneuvers like sphincterotomy, balloon sweep, use of basket/mechanical lithotriptor, stone clearance, stricture dilation using balloon/catheter, stent insertion, stent removal and use of single operator cholangioscopy/pancreatoscopy.
Time frame: 30-60 minutes
| Arm | Measure | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|
| Duodenoscope With Disposable Distal End Caps | Number of Participant With Procure Success (Ability to Perform Interventions) | 154 Participants |
| Conventional Reusable Duodenoscope | Number of Participant With Procure Success (Ability to Perform Interventions) | 155 Participants |