Evaluation of BiCROS Fitting Benefits

Evaluate the Benefit of a BiCROS Hearing Aid Fitting

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05379231
Enrollment
22
Registered
2022-05-18
Start date
2022-07-11
Completion date
2023-03-21
Last updated
2025-06-11

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Unilateral Deafness

Brief summary

22 candidates for BiCROS hearing aids will be recruited and fitted with a hearing aid and a CROS device. Following a real-world trial, each participant will be tested on a range of outcome measures including speech recognition in noise, ratings of listening effort, and ratings of preference.

Detailed description

Unilateral Hearing Loss (UHL) can be defined as any degree of permanent hearing loss on one ear with normal hearing in the opposite ear. Unilateral hearing loss can be debilitating associated with audiological, psychosocial and educational challenges. Audiological challenges include the reduced ability to localize sounds, reduced awareness on the unaidable side and difficulties hearing in noise or at a distance. One solution for unaidable UHL is Contralateral Routing of Signals (CROS) and Bilateral Contralateral routing of signals (BiCROS) using a so-called CROS transmitter together with a hearing aid. The rationale for this clinical investigation is to collect clinical data with a rechargeable CROS transmitter to evaluate the benefits of a BiCROS fitting compared to the alternative treatment option, the monaural hearing aid fitting, and to no treatment in noisy listening situations. Hearing tests, ear impressions and hearing aid fitting visits are mainly conducted at the National Centre for Audiology in London, Ontario. Those visits are available at the Sonova Innovation Centre Toronto in Mississauga, Ontario, upon request. Outcome measurements are only conducted at the National Centre for Audiology in London, Ontario.

Interventions

A participant will wear a hearing aid with and without a CROS device on the unaidable ear.

Sponsors

Western University, Canada
CollaboratorOTHER
Sonova AG
Lead SponsorINDUSTRY

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
OTHER
Masking
NONE

Masking description

Participants were wearing the BiCROS fitting during home trial, where no masking was possible. For in-lab tests, the CROS device was deactivated for the monaural condition, allowing for some masking. The unaided condition could not be masked. Participants were not aware which beamformer configuration was active during in-lab tests.

Intervention model description

All participants were wearing the BiCROS fitting during home trial, and tested in-lab with all conditions, randomized and counterbalanced (BiCROS fitting, monaural fitting, unaided).

Eligibility

Sex/Gender
ALL
Age
18 Years to 99 Years
Healthy volunteers
Yes

Inclusion criteria

* unaidable hearing loss in worse ear (\>N6) * aidable hearing loss (N3 to N5) on the other ear * fluent English speaker

Exclusion criteria

\- aidable hearing loss in both ears

Design outcomes

Primary

MeasureTime frameDescription
Speech Intelligibility in Noise Measured With US Matrix Test (Noise From a Defined Loudspeaker)At the appointment following the 4 weeks home trial each participant went on wearing the BiCROS fittingMeasuring the Signal to Noise Ratio change in decibels (dB) for speech from the front loudspeaker and noise from the better ear's side with the US matrix Test. Results of the speech test will be the SRT (speech recognition threshold) in dB SNR (signal to noise ratio). The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A)). The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 65 dB (A).

Secondary

MeasureTime frameDescription
Reports of Subjective Quality of Hearing as Assessed by the SSQAt the appointment following the 4 weeks home trial each participant went on wearing the BiCROS fittingMeasured after home trial with the Speech, Spatial and Qualities of Hearing Scale-Benefit (Gatehouse, 2004). The scale ranges from -5 to 5, where negative values indicate that the condition under test is worse than the comparator, 0 indicates no change, and positive values indicate a benefit of the condition under test compared to the comparator.
Reports of Subjective Quality of Hearing as Assessed by the Bern Benefit QuestionnairesAt the appointment following the 4 weeks home trial each participant went on wearing the BiCROS fittingMeasured after home trial with the Bern Benefit in Single-Sided Deafness Questionnaire (Komps et al, 2011). The scale ranges from -5 to 5, where negative values indicate that the condition under test is better wihtout the hearing aid, 0 indicates no change, and positive values indicate a benefit with the hearing aid.
Speech Intelligibility in Noise Measured With US Matrix Test (Diffuse Noise)At the appointment following the 4 weeks home trial each participant went on wearing the BiCROS fittingMeasuring the Signal to Noise Ratio change in decibels (dB) for speech from the front loudspeaker in surrounding noise with the US Matrix Test. Results of the speech test will be the SRT (speech recognition threshold) in dB SNR (signal to noise ratio). The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A). The noise will be presented from a circle of 8 surrounding speakers from 45° to 315° at a level of 65 dB (A) with the participant seated in the centre of the circle.
Subjective Preference of Hearing Aid ConditionAt the appointment following the 4 weeks home trial each participant went on wearing the BiCROS fitting3 alternative forced choice (A, B, no preference or ABX) paradigm with three repetitions for speech in noise presented via loudsepakers
Listening Effort in Noise Using 7 Point Scale (Holmes et. al., 2018)At the appointment following the 4 weeks home trial each participant went on wearing the BiCROS fittingMeasuring listening effort for speech in multi-talker babble presented via loudspeaker. The scale ranges from 1 to 7, and a higher rating indicates less listening effort.
Sound Quality Ratings for Total Impression and Clarity Using Gabriellson et al (1990) Ratings ScalesAt the appointment following the 4 weeks home trial each participant went on wearing the BiCROS fittingRatings of speech in noise presented via loudspeakers using Gabriellson et al (1990) Ratings Scales. Scales include Clarity and Total Impression, and range from 1 to 10. The higher the rating, the better the Clarity and Total Impression.

Countries

Canada

Participant flow

Pre-assignment details

2 participants had to be excluded from continuing in the study before assignment because they did not meet the inclusion criteria.

Participants by arm

ArmCount
Complete Sample
All participants will perform the same tests in all conditions. Although each ear receives a different device configuration depending on the condition, the participant is assigned as a whole, and assignment is not by ears. BiCROS fitting: Each participant will be fitted with the experimental CROS device on the unaidable ear and with a compatible hearing aid on the other ear. Monaural hearing aid fitting: Each participant will be fitted with a hearing aid on the better hearing ear and no device on the unaidable ear. Unaided condition: No intervention, the participants are not fitted with a CROS device and/or hearing aid. Since the characteristics of each arm are unrelevant for the study outcome, there has been no separate analysis on each arm.
18
Total18

Withdrawals & dropouts

PeriodReasonFG000
Home VisitWithdrawal by Subject2
In-lab VisitWithdrawal by Subject2

Baseline characteristics

CharacteristicComplete Sample
Age, Continuous74.6 years
BiCROS fitting candidate18 Participants
Race and Ethnicity Not Collected— Participants
Region of Enrollment
Canada
18 participants
Sex: Female, Male
Female
8 Participants
Sex: Female, Male
Male
10 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 22
other
Total, other adverse events
2 / 22
serious
Total, serious adverse events
0 / 22

Outcome results

Primary

Speech Intelligibility in Noise Measured With US Matrix Test (Noise From a Defined Loudspeaker)

Measuring the Signal to Noise Ratio change in decibels (dB) for speech from the front loudspeaker and noise from the better ear's side with the US matrix Test. Results of the speech test will be the SRT (speech recognition threshold) in dB SNR (signal to noise ratio). The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A)). The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 65 dB (A).

Time frame: At the appointment following the 4 weeks home trial each participant went on wearing the BiCROS fitting

Population: Although each ear receives a different device configuration depending on the condition (except for the unaided condition), the participant is assigned as a whole, and assignment is not by ears.

ArmMeasureValue (MEAN)Dispersion
BiCROS With BeamformerSpeech Intelligibility in Noise Measured With US Matrix Test (Noise From a Defined Loudspeaker)-8.8 dBSRTStandard Deviation 3.4
BiCROS OmniSpeech Intelligibility in Noise Measured With US Matrix Test (Noise From a Defined Loudspeaker)-3.7 dBSRTStandard Deviation 2.9
Monaural With BeamformerSpeech Intelligibility in Noise Measured With US Matrix Test (Noise From a Defined Loudspeaker)-9.2 dBSRTStandard Deviation 2.7
Monaural OmniSpeech Intelligibility in Noise Measured With US Matrix Test (Noise From a Defined Loudspeaker)-2.8 dBSRTStandard Deviation 2.5
UnaidedSpeech Intelligibility in Noise Measured With US Matrix Test (Noise From a Defined Loudspeaker)-0.1 dBSRTStandard Deviation 6.8
Secondary

Listening Effort in Noise Using 7 Point Scale (Holmes et. al., 2018)

Measuring listening effort for speech in multi-talker babble presented via loudspeaker. The scale ranges from 1 to 7, and a higher rating indicates less listening effort.

Time frame: At the appointment following the 4 weeks home trial each participant went on wearing the BiCROS fitting

Population: Although each ear receives a different device configuration depending on the condition (except for the unaided condition), the participant is assigned as a whole, and assignment is not by ears.

ArmMeasureValue (MEAN)Dispersion
BiCROS With BeamformerListening Effort in Noise Using 7 Point Scale (Holmes et. al., 2018)3.75 score on a scaleStandard Deviation 1.8
BiCROS OmniListening Effort in Noise Using 7 Point Scale (Holmes et. al., 2018)4.9 score on a scaleStandard Deviation 1.9
Monaural With BeamformerListening Effort in Noise Using 7 Point Scale (Holmes et. al., 2018)3.9 score on a scaleStandard Deviation 1.6
Monaural OmniListening Effort in Noise Using 7 Point Scale (Holmes et. al., 2018)5 score on a scaleStandard Deviation 1.5
UnaidedListening Effort in Noise Using 7 Point Scale (Holmes et. al., 2018)5.7 score on a scaleStandard Deviation 1.8
Secondary

Reports of Subjective Quality of Hearing as Assessed by the Bern Benefit Questionnaires

Measured after home trial with the Bern Benefit in Single-Sided Deafness Questionnaire (Komps et al, 2011). The scale ranges from -5 to 5, where negative values indicate that the condition under test is better wihtout the hearing aid, 0 indicates no change, and positive values indicate a benefit with the hearing aid.

Time frame: At the appointment following the 4 weeks home trial each participant went on wearing the BiCROS fitting

Population: Although each ear receives a different device configuration, the participant is assigned as a whole, and assignment is not by ears.

ArmMeasureValue (MEAN)Dispersion
BiCROS With BeamformerReports of Subjective Quality of Hearing as Assessed by the Bern Benefit Questionnaires2.83 score on a scaleStandard Deviation 0.76
Secondary

Reports of Subjective Quality of Hearing as Assessed by the SSQ

Measured after home trial with the Speech, Spatial and Qualities of Hearing Scale-Benefit (Gatehouse, 2004). The scale ranges from -5 to 5, where negative values indicate that the condition under test is worse than the comparator, 0 indicates no change, and positive values indicate a benefit of the condition under test compared to the comparator.

Time frame: At the appointment following the 4 weeks home trial each participant went on wearing the BiCROS fitting

Population: Although each ear receives a different device configuration, the participant is assigned as a whole, and assignment is not by ears.

ArmMeasureGroupValue (MEAN)Dispersion
BiCROS With BeamformerReports of Subjective Quality of Hearing as Assessed by the SSQSpeech2.48 score on a scaleStandard Deviation 0.59
BiCROS With BeamformerReports of Subjective Quality of Hearing as Assessed by the SSQSpatial1.16 score on a scaleStandard Deviation 0.47
BiCROS With BeamformerReports of Subjective Quality of Hearing as Assessed by the SSQQualities2.55 score on a scaleStandard Deviation 0.35
Secondary

Sound Quality Ratings for Total Impression and Clarity Using Gabriellson et al (1990) Ratings Scales

Ratings of speech in noise presented via loudspeakers using Gabriellson et al (1990) Ratings Scales. Scales include Clarity and Total Impression, and range from 1 to 10. The higher the rating, the better the Clarity and Total Impression.

Time frame: At the appointment following the 4 weeks home trial each participant went on wearing the BiCROS fitting

Population: Although each ear receives a different device configuration depending on the condition (except for the unaided condition), the participant is assigned as a whole, and assignment is not by ears.

ArmMeasureGroupValue (MEAN)Dispersion
BiCROS With BeamformerSound Quality Ratings for Total Impression and Clarity Using Gabriellson et al (1990) Ratings ScalesClarity6.8 score on a scaleStandard Deviation 2.4
BiCROS With BeamformerSound Quality Ratings for Total Impression and Clarity Using Gabriellson et al (1990) Ratings ScalesTotal Impression6.1 score on a scaleStandard Deviation 2.7
BiCROS OmniSound Quality Ratings for Total Impression and Clarity Using Gabriellson et al (1990) Ratings ScalesClarity5.2 score on a scaleStandard Deviation 2.8
BiCROS OmniSound Quality Ratings for Total Impression and Clarity Using Gabriellson et al (1990) Ratings ScalesTotal Impression4.7 score on a scaleStandard Deviation 2.8
Monaural With BeamformerSound Quality Ratings for Total Impression and Clarity Using Gabriellson et al (1990) Ratings ScalesClarity6.5 score on a scaleStandard Deviation 2.3
Monaural With BeamformerSound Quality Ratings for Total Impression and Clarity Using Gabriellson et al (1990) Ratings ScalesTotal Impression6 score on a scaleStandard Deviation 2.3
Monaural OmniSound Quality Ratings for Total Impression and Clarity Using Gabriellson et al (1990) Ratings ScalesTotal Impression4.3 score on a scaleStandard Deviation 2.5
Monaural OmniSound Quality Ratings for Total Impression and Clarity Using Gabriellson et al (1990) Ratings ScalesClarity4.8 score on a scaleStandard Deviation 2.8
UnaidedSound Quality Ratings for Total Impression and Clarity Using Gabriellson et al (1990) Ratings ScalesClarity3.7 score on a scaleStandard Deviation 3
UnaidedSound Quality Ratings for Total Impression and Clarity Using Gabriellson et al (1990) Ratings ScalesTotal Impression3.7 score on a scaleStandard Deviation 2.6
Secondary

Speech Intelligibility in Noise Measured With US Matrix Test (Diffuse Noise)

Measuring the Signal to Noise Ratio change in decibels (dB) for speech from the front loudspeaker in surrounding noise with the US Matrix Test. Results of the speech test will be the SRT (speech recognition threshold) in dB SNR (signal to noise ratio). The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A). The noise will be presented from a circle of 8 surrounding speakers from 45° to 315° at a level of 65 dB (A) with the participant seated in the centre of the circle.

Time frame: At the appointment following the 4 weeks home trial each participant went on wearing the BiCROS fitting

Population: Although each ear receives a different device configuration depending on the condition (except for the unaided condition), the participant is assigned as a whole, and assignment is not by ears.

ArmMeasureValue (MEAN)Dispersion
BiCROS With BeamformerSpeech Intelligibility in Noise Measured With US Matrix Test (Diffuse Noise)-14.2 dBSRTStandard Deviation 3.3
BiCROS OmniSpeech Intelligibility in Noise Measured With US Matrix Test (Diffuse Noise)-8 dBSRTStandard Deviation 4.2
Monaural With BeamformerSpeech Intelligibility in Noise Measured With US Matrix Test (Diffuse Noise)-13.9 dBSRTStandard Deviation 3.6
Monaural OmniSpeech Intelligibility in Noise Measured With US Matrix Test (Diffuse Noise)-9.3 dBSRTStandard Deviation 3
UnaidedSpeech Intelligibility in Noise Measured With US Matrix Test (Diffuse Noise)-3.4 dBSRTStandard Deviation 9.4
Secondary

Subjective Preference of Hearing Aid Condition

3 alternative forced choice (A, B, no preference or ABX) paradigm with three repetitions for speech in noise presented via loudsepakers

Time frame: At the appointment following the 4 weeks home trial each participant went on wearing the BiCROS fitting

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
BiCROS With BeamformerSubjective Preference of Hearing Aid ConditionDirectional0 Participants
BiCROS With BeamformerSubjective Preference of Hearing Aid ConditionNo preference7 Participants
BiCROS With BeamformerSubjective Preference of Hearing Aid ConditionOmnidirectional0 Participants
BiCROS With BeamformerSubjective Preference of Hearing Aid ConditionBiCROS5 Participants
BiCROS With BeamformerSubjective Preference of Hearing Aid ConditionMonaural6 Participants
BiCROS OmniSubjective Preference of Hearing Aid ConditionBiCROS4 Participants
BiCROS OmniSubjective Preference of Hearing Aid ConditionDirectional0 Participants
BiCROS OmniSubjective Preference of Hearing Aid ConditionOmnidirectional0 Participants
BiCROS OmniSubjective Preference of Hearing Aid ConditionMonaural3 Participants
BiCROS OmniSubjective Preference of Hearing Aid ConditionNo preference11 Participants
Monaural With BeamformerSubjective Preference of Hearing Aid ConditionNo preference2 Participants
Monaural With BeamformerSubjective Preference of Hearing Aid ConditionMonaural0 Participants
Monaural With BeamformerSubjective Preference of Hearing Aid ConditionBiCROS0 Participants
Monaural With BeamformerSubjective Preference of Hearing Aid ConditionDirectional16 Participants
Monaural With BeamformerSubjective Preference of Hearing Aid ConditionOmnidirectional0 Participants
Monaural OmniSubjective Preference of Hearing Aid ConditionDirectional14 Participants
Monaural OmniSubjective Preference of Hearing Aid ConditionMonaural0 Participants
Monaural OmniSubjective Preference of Hearing Aid ConditionNo preference4 Participants
Monaural OmniSubjective Preference of Hearing Aid ConditionBiCROS0 Participants
Monaural OmniSubjective Preference of Hearing Aid ConditionOmnidirectional0 Participants