A Study Assessing Arrhythmia Mapping With a Globe-Shaped, High-Density, Multi-Electrode Mapping Catheter

Clinical Evaluation of Arrhythmia Mapping With a Globe-Shaped, High-Density, Multi-Electrode Mapping Catheter

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05373862
Acronym
COSMOS
Enrollment
40
Registered
2022-05-13
Start date
2022-07-29
Completion date
2023-02-20
Last updated
2025-03-03

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Paroxysmal Atrial Fibrillation, Persistent Atrial Fibrillation, Premature Ventricular Complex, Scar-related Atrial Tachycardia, Ventricular Tachycardia

Brief summary

The purpose of this study is to assess the performance and safety for the use of the investigational catheter for intracardiac mapping in the atria and ventricles.

Interventions

DEVICEGlobe-Shaped, High-Density, Multi-Electrode Mapping Catheter

Participants scheduled to have a clinically-indicated catheter mapping and ablation procedure using globe-shaped, high-density, multi-electrode mapping catheter for management of atrial procedure (scar-related atrial tachycardia, persistent atrial fibrillation, paroxysmal atrial fibrillation) and Ventricular procedures (ventricular tachycardia, premature ventricular complex).

Sponsors

Biosense Webster, Inc.
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Diagnosed with and candidate for clinically-indicated catheter mapping and ablation procedure for the management of ventricular tachycardia, premature ventricular complex, atrial tachycardia or atrial fibrillation (participants having undergone a previous ablation procedure may be included) * At least one episode of the targeted arrhythmia (ventricular tachycardia, premature ventricular complex, atrial tachycardia or atrial fibrillation) must have been documented by electrocardiogram (ECG), Holter, loop recorder, telemetry, implanted device, or transtelephonic monitoring within 12 months of enrollment * Signed participants Informed Consent Form (ICF) * Able and willing to comply with all pre-, post-, and follow-up testing and requirements

Exclusion criteria

* Diagnosed with an arrhythmia requiring epicardial mapping * Study arrhythmia secondary to reversible cause, or secondary to electrolyte imbalance, thyroid disease, or non-cardiac cause * Atrial arrhythmias: participants with a left atrial size greater than (\>) 55 millimeters (mm) * Left Ventricular Ejection Fraction (LVEF) less than or equal to (\<=) 25 percent (%) for participants with ventricular arrhythmia * LVEF \<= 40% for participants with atrial arrhythmia * Documented intracardiac thrombus as detected on imaging within 24 hours prior to insertion of the investigational catheter * Contraindication to anticoagulation (that is heparin, warfarin, dabigatran) * History of blood clotting or bleeding abnormalities (example, hypercoagulable state) * Myocardial infarction within the past 2 months (60 days) * Documented thromboembolic event (including Transient Ischemic Attack \[TIA\]) within the past 12 months (365 days) * Uncontrolled heart failure or New York Heart Association (NYHA) function class IV * Implanted with a pacemaker or intracardiac cardiac defibrillator within the past 6 weeks (42 days) * Participants with known untreatable allergy to contrast media * Active illness or active systemic infection or sepsis * Diagnosed atrial or ventricular myxoma, interatrial baffle or patch, tumor or other abnormality that precludes catheter introduction or manipulation * Significant congenital anomaly or medical problem that in the opinion of the investigator would preclude enrollment in this study * Participants that have ever undergone a percutaneous or surgical valvular cardiac procedure (that is, ventriculotomy, atriotomy, and valve repair or replacement and presence of a prosthetic valve) * Any cardiac surgery within the past 60 days (2 months) (includes Percutaneous Coronary Intervention \[PCI\]) * Atrial septal closure within the past 6 weeks (42 days) * Presence of a condition that precludes vascular access * Women who are pregnant (as evidenced by pregnancy test if pre-menopausal), lactating, or who are of childbearing age and plan on becoming pregnant during the course of the clinical investigation * Categorized as vulnerable population and requires special treatment with respect to safeguards of well-being * Concurrent enrollment in an investigational study evaluating another device or drug

Design outcomes

Primary

MeasureTime frameDescription
Number of Participants With Pre-ablation Mapping Requirements and Clinically Indicated Mapping Performed With the Investigational Catheter Without Resort to Non-study Mapping Catheter(s)Up to 7 days of index procedure on Day 1Number of participants with pre-ablation mapping requirements and clinically indicated mapping performed with the investigational catheter without resort to non-study mapping catheter(s) were reported.
Number of Participants With Serious Adverse Events (SAEs) Within 7 Days of Index Procedure Related to the Investigational CatheterUp to 7 days of index procedure on Day 1Number of participants with SAEs within 7 days of index procedure related to the investigational catheter were reported. An adverse event (AE) is any untoward medical occurrence in a participant whether or not related to the investigational device. SAE is any AE that results in: death, persistent or significant disability/incapacity, requires inpatient hospitalization or prolongation of existing hospitalization, is life-threatening experience, is a congenital anomaly/birth defect and may jeopardize participant and/or may require medical or surgical intervention to prevent one of the outcomes listed above.

Secondary

MeasureTime frameDescription
Number of Responders for Physician Assessment for PacingUp to 29 weeksNumber of responders for physician assessment for pacing (high output stimulation pacing and local pacing capture) were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for pacing.
Number of Responders for Physician Assessment for Catheter DesignUp to 29 weeksNumber of responders for physician assessment for catheter design were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for catheter design.
Number of Responders for Physician Assessment for WorkflowUp to 29 weeksNumber of responders for physician assessment for workflow were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for workflow.
Number of Responders for Physician Assessment for Catheter VisualizationUp to 29 weeksNumber of responders for physician assessment for catheter visualization were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for catheter visualization.
Number of Responders for Physician Assessment for Catheters InteractionsUp to 29 weeksNumber of responders for physician assessment for catheters interactions were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for catheters interactions.
Number of Responders for Physician Assessment for Maneuverability and HandlingUp to 29 weeksNumber of responders for physician assessment for maneuverability & handling were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for maneuverability and handling.
Number of Responders for Physician Assessment for Design and Coverage for Confirming Pulmonary Vein Isolation (PVI)Up to 29 weeksNumber of responders for physician assessment for design and coverage for confirming PVI were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for design and coverage for confirming PVI.
Number of Responders for Physician Assessment for Ability to Characterize the TissueUp to 29 weeksNumber of responders for physician assessment for ability to characterize the tissue were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for ability to characterize the tissue.
Number of Responders for Physician Assessment for Ability to Identify Arrhythmia Circuit or Source CorrectlyUp to 29 weeksNumber of responders for physician assessment for ability to identify arrhythmia circuit or source correctly were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for ability to identify arrhythmia circuit or source correctly.
Number of Participants With SAEs Excluding Investigational Catheter Related Within 7 Days of Index ProcedureUp to 7 days of index procedure on Day 1Number of participants with SAEs excluding investigational catheter related within 7 days of index procedure were reported. An AE is any untoward medical occurrence in a participant whether or not related to the investigational device. SAE is any AE that results in: death, persistent or significant disability/incapacity, requires inpatient hospitalization or prolongation of existing hospitalization, is life-threatening experience, is a congenital anomaly/birth defect and may jeopardize participant and/or may require medical or surgical intervention to prevent one of the outcomes listed above.
Number of Participants With Non-serious Adverse Events Within 7 Days of Index Procedure Related to the Investigational CatheterUp to 7 days of index procedure on Day 1Number of participants with non-serious AEs within 7 days of index procedure related to the investigational catheter were reported. An AE is any untoward medical occurrence in a participant whether or not related to the investigational device.
Number of Responders for Physician Assessment for ArrhythmogenicityUp to 29 weeksNumber of responders for physician assessment for arrhythmogenicity were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for arrhythmogenicity.
Number of Responders for Physician Assessment for Signal Collection and QualityUp to 29 weeksNumber of responders for physician assessment for signal collection and quality (UNIPOLAR signals in atria/ventricles, UNIPOLAR noise encountered, BIPOLAR signals in atria/ventricles, BASELINE noise encountered, and BIPOLAR noise encountered) were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for signal collection and quality.

Countries

Belgium, Croatia, Italy

Participant flow

Participants by arm

ArmCount
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)
Participants diagnosed with cardiac arrhythmias who were scheduled to have a clinically-indicated catheter mapping and ablation procedure of arrhythmia management for atrial and ventricular procedures, underwent pre-ablation mapping of the chambers of interest using the multi-electrode ECG mapping catheter (investigational catheter) prior to a standard of care ablation procedure.
40
Total40

Baseline characteristics

CharacteristicMulti-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)
Age, Continuous58.0 years
STANDARD_DEVIATION 15.73
Race and Ethnicity Not Collected— Participants
Sex: Female, Male
Female
15 Participants
Sex: Female, Male
Male
25 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 40
other
Total, other adverse events
1 / 40
serious
Total, serious adverse events
4 / 40

Outcome results

Primary

Number of Participants With Pre-ablation Mapping Requirements and Clinically Indicated Mapping Performed With the Investigational Catheter Without Resort to Non-study Mapping Catheter(s)

Number of participants with pre-ablation mapping requirements and clinically indicated mapping performed with the investigational catheter without resort to non-study mapping catheter(s) were reported.

Time frame: Up to 7 days of index procedure on Day 1

Population: Per-protocol analysis set included all participants who completed the protocol-required pre-ablation mapping with the investigational catheter.

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Participants With Pre-ablation Mapping Requirements and Clinically Indicated Mapping Performed With the Investigational Catheter Without Resort to Non-study Mapping Catheter(s)40 Participants
Primary

Number of Participants With Serious Adverse Events (SAEs) Within 7 Days of Index Procedure Related to the Investigational Catheter

Number of participants with SAEs within 7 days of index procedure related to the investigational catheter were reported. An adverse event (AE) is any untoward medical occurrence in a participant whether or not related to the investigational device. SAE is any AE that results in: death, persistent or significant disability/incapacity, requires inpatient hospitalization or prolongation of existing hospitalization, is life-threatening experience, is a congenital anomaly/birth defect and may jeopardize participant and/or may require medical or surgical intervention to prevent one of the outcomes listed above.

Time frame: Up to 7 days of index procedure on Day 1

Population: The safety analysis set consisted of all enrolled participants who underwent insertion of the study catheter.

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Participants With Serious Adverse Events (SAEs) Within 7 Days of Index Procedure Related to the Investigational Catheter1 Participants
Secondary

Number of Participants With Non-serious Adverse Events Within 7 Days of Index Procedure Related to the Investigational Catheter

Number of participants with non-serious AEs within 7 days of index procedure related to the investigational catheter were reported. An AE is any untoward medical occurrence in a participant whether or not related to the investigational device.

Time frame: Up to 7 days of index procedure on Day 1

Population: The safety analysis set consisted of all enrolled participants who underwent insertion of the study catheter.

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Participants With Non-serious Adverse Events Within 7 Days of Index Procedure Related to the Investigational Catheter1 Participants
Secondary

Number of Participants With SAEs Excluding Investigational Catheter Related Within 7 Days of Index Procedure

Number of participants with SAEs excluding investigational catheter related within 7 days of index procedure were reported. An AE is any untoward medical occurrence in a participant whether or not related to the investigational device. SAE is any AE that results in: death, persistent or significant disability/incapacity, requires inpatient hospitalization or prolongation of existing hospitalization, is life-threatening experience, is a congenital anomaly/birth defect and may jeopardize participant and/or may require medical or surgical intervention to prevent one of the outcomes listed above.

Time frame: Up to 7 days of index procedure on Day 1

Population: The safety analysis set consisted of all enrolled participants who underwent insertion of the study catheter.

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Participants With SAEs Excluding Investigational Catheter Related Within 7 Days of Index Procedure3 Participants
Secondary

Number of Responders for Physician Assessment for Ability to Characterize the Tissue

Number of responders for physician assessment for ability to characterize the tissue were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for ability to characterize the tissue.

Time frame: Up to 29 weeks

Population: The safety analysis set consisted of all enrolled participants who underwent insertion of the study catheter. Here, 'n' (number analyzed) is defined as participants analyzed for specified categories.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Characterize the TissueCompared to the PENTARAY catheter: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Characterize the TissueCompared to the PENTARAY catheter: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Characterize the TissueCompared to the PENTARAY catheter: Score 31 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Characterize the TissueCompared to the PENTARAY catheter: Score 45 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Characterize the TissueCompared to the PENTARAY catheter: Score 510 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Characterize the TissueCompared to the PENTARAY catheter: Score 617 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Characterize the TissueCompared to the PENTARAY catheter: Score 77 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Characterize the TissueCompared to other competitive multielectrode catheters: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Characterize the TissueCompared to other competitive multielectrode catheters: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Characterize the TissueCompared to other competitive multielectrode catheters: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Characterize the TissueCompared to other competitive multielectrode catheters: Score 48 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Characterize the TissueCompared to other competitive multielectrode catheters: Score 59 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Characterize the TissueCompared to other competitive multielectrode catheters: Score 610 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Characterize the TissueCompared to other competitive multielectrode catheters: Score 71 Participants
Secondary

Number of Responders for Physician Assessment for Ability to Identify Arrhythmia Circuit or Source Correctly

Number of responders for physician assessment for ability to identify arrhythmia circuit or source correctly were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for ability to identify arrhythmia circuit or source correctly.

Time frame: Up to 29 weeks

Population: The safety analysis set consisted of all enrolled participants who underwent insertion of the study catheter. Here, 'n' (number analyzed) is defined as participants analyzed for specified categories.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Identify Arrhythmia Circuit or Source CorrectlyCompared to the PENTARAY catheter: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Identify Arrhythmia Circuit or Source CorrectlyCompared to the PENTARAY catheter: Score 21 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Identify Arrhythmia Circuit or Source CorrectlyCompared to the PENTARAY catheter: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Identify Arrhythmia Circuit or Source CorrectlyCompared to the PENTARAY catheter: Score 43 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Identify Arrhythmia Circuit or Source CorrectlyCompared to the PENTARAY catheter: Score 514 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Identify Arrhythmia Circuit or Source CorrectlyCompared to the PENTARAY catheter: Score 618 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Identify Arrhythmia Circuit or Source CorrectlyCompared to the PENTARAY catheter: Score 74 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Identify Arrhythmia Circuit or Source CorrectlyCompared to other competitive multielectrode catheters: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Identify Arrhythmia Circuit or Source CorrectlyCompared to other competitive multielectrode catheters: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Identify Arrhythmia Circuit or Source CorrectlyCompared to other competitive multielectrode catheters: Score 31 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Identify Arrhythmia Circuit or Source CorrectlyCompared to other competitive multielectrode catheters: Score 45 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Identify Arrhythmia Circuit or Source CorrectlyCompared to other competitive multielectrode catheters: Score 514 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Identify Arrhythmia Circuit or Source CorrectlyCompared to other competitive multielectrode catheters: Score 67 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Ability to Identify Arrhythmia Circuit or Source CorrectlyCompared to other competitive multielectrode catheters: Score 71 Participants
Secondary

Number of Responders for Physician Assessment for Arrhythmogenicity

Number of responders for physician assessment for arrhythmogenicity were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for arrhythmogenicity.

Time frame: Up to 29 weeks

Population: The safety analysis set consisted of all enrolled participants who underwent insertion of the study catheter. Here, 'n' (number analyzed) is defined as participants analyzed for specified categories.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for ArrhythmogenicityCompared to the PENTARAY catheter: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for ArrhythmogenicityCompared to the PENTARAY catheter: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for ArrhythmogenicityCompared to the PENTARAY catheter: Score 31 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for ArrhythmogenicityCompared to the PENTARAY catheter: Score 41 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for ArrhythmogenicityCompared to the PENTARAY catheter: Score 510 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for ArrhythmogenicityCompared to the PENTARAY catheter: Score 620 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for ArrhythmogenicityCompared to the PENTARAY catheter: Score 78 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for ArrhythmogenicityCompared to other competitive multielectrode catheters: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for ArrhythmogenicityCompared to other competitive multielectrode catheters: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for ArrhythmogenicityCompared to other competitive multielectrode catheters: Score 31 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for ArrhythmogenicityCompared to other competitive multielectrode catheters: Score 45 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for ArrhythmogenicityCompared to other competitive multielectrode catheters: Score 57 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for ArrhythmogenicityCompared to other competitive multielectrode catheters: Score 612 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for ArrhythmogenicityCompared to other competitive multielectrode catheters: Score 74 Participants
Secondary

Number of Responders for Physician Assessment for Catheter Design

Number of responders for physician assessment for catheter design were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for catheter design.

Time frame: Up to 29 weeks

Population: The safety analysis set consisted of all enrolled participants who underwent insertion of the study catheter.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignElectrodes spacing: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignElectrodes spacing: Score 44 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignAbility to achieve electrode contact with cardiac tissue: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignAbility to achieve electrode contact with cardiac tissue: Score 21 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignAbility to achieve electrode contact with cardiac tissue: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignAbility to achieve electrode contact with cardiac tissue: Score 44 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignAbility to achieve electrode contact with cardiac tissue: Score 511 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignAbility to achieve electrode contact with cardiac tissue: Score 622 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignAbility to achieve electrode contact with cardiac tissue: Score 72 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignElectrodes spacing: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignElectrodes spacing: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignElectrodes spacing: Score 510 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignElectrodes spacing: Score 613 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignElectrodes spacing: Score 713 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignElectrode array coverage: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignElectrode array coverage: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignElectrode array coverage: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignElectrode array coverage: Score 44 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignElectrode array coverage: Score 510 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignElectrode array coverage: Score 618 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter DesignElectrode array coverage: Score 78 Participants
Secondary

Number of Responders for Physician Assessment for Catheters Interactions

Number of responders for physician assessment for catheters interactions were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for catheters interactions.

Time frame: Up to 29 weeks

Population: The safety analysis set consisted of all enrolled participants who underwent insertion of the study catheter. Here, 'n' (number analyzed) is defined as participants analyzed for specified categories.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsCARTO 3 System: Score 11 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsCARTO 3 System: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsCARTO 3 System: Score 32 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsCARTO 3 System: Score 47 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsCARTO 3 System: Score 510 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsCARTO 3 System: Score 616 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsCARTO 3 System: Score 74 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsElectrophysiology (EP) Lab recording equipment: Score 11 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsEP Lab recording equipment: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsEP Lab recording equipment: Score 32 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsEP Lab recording equipment: Score 46 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsEP Lab recording equipment: Score 516 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsEP Lab recording equipment: Score 611 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsEP Lab recording equipment: Score 74 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsPacing equipment: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsPacing equipment: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsPacing equipment: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsPacing equipment: Score 44 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsPacing equipment: Score 512 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsPacing equipment: Score 63 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsPacing equipment: Score 78 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsIntroducer tool on catheter: Score 11 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsIntroducer tool on catheter: Score 21 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsIntroducer tool on catheter: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsIntroducer tool on catheter: Score 45 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsIntroducer tool on catheter: Score 58 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsIntroducer tool on catheter: Score 613 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsIntroducer tool on catheter: Score 712 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsSheath: Score 11 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsSheath: Score 21 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsSheath: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsSheath: Score 45 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsSheath: Score 59 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsSheath: Score 612 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsSheath: Score 712 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsIrrigation bag and tubing: Score 11 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsIrrigation bag and tubing: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsIrrigation bag and tubing: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsIrrigation bag and tubing: Score 46 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsIrrigation bag and tubing: Score 59 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsIrrigation bag and tubing: Score 612 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsIrrigation bag and tubing: Score 712 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsImplantable cardioverter-defibrillators (ICD) or other implantable devices: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsICD or other implantable devices: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsICD or other implantable devices: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsICD or other implantable devices: Score 40 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsICD or other implantable devices: Score 50 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsICD or other implantable devices: Score 61 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheters InteractionsICD or other implantable devices: Score 70 Participants
Secondary

Number of Responders for Physician Assessment for Catheter Visualization

Number of responders for physician assessment for catheter visualization were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for catheter visualization.

Time frame: Up to 29 weeks

Population: The safety analysis set consisted of all enrolled participants who underwent insertion of the study catheter. Here, 'n' (number analyzed) is defined as participants analyzed for specified categories.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter VisualizationVisualization under fluoroscopy: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter VisualizationVisualization under fluoroscopy: Score 22 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter VisualizationVisualization under fluoroscopy: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter VisualizationVisualization under fluoroscopy: Score 45 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter VisualizationVisualization under fluoroscopy: Score 510 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter VisualizationVisualization under fluoroscopy: Score 618 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter VisualizationVisualization under fluoroscopy: Score 73 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter VisualizationVisualization in CARTO 3: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter VisualizationVisualization in CARTO 3: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter VisualizationVisualization in CARTO 3: Score 34 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter VisualizationVisualization in CARTO 3: Score 46 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter VisualizationVisualization in CARTO 3: Score 59 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter VisualizationVisualization in CARTO 3: Score 69 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Catheter VisualizationVisualization in CARTO 3: Score 712 Participants
Secondary

Number of Responders for Physician Assessment for Design and Coverage for Confirming Pulmonary Vein Isolation (PVI)

Number of responders for physician assessment for design and coverage for confirming PVI were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for design and coverage for confirming PVI.

Time frame: Up to 29 weeks

Population: The safety analysis set consisted of all enrolled participants who underwent insertion of the study catheter. Here, 'N' (number of participants analyzed) signifies participants evaluated for this outcome measure and 'n' (number analyzed) is defined as participants analyzed for specified categories.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Design and Coverage for Confirming Pulmonary Vein Isolation (PVI)Compared to the PENTARAY catheter: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Design and Coverage for Confirming Pulmonary Vein Isolation (PVI)Compared to the PENTARAY catheter: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Design and Coverage for Confirming Pulmonary Vein Isolation (PVI)Compared to the PENTARAY catheter: Score 31 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Design and Coverage for Confirming Pulmonary Vein Isolation (PVI)Compared to the PENTARAY catheter: Score 41 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Design and Coverage for Confirming Pulmonary Vein Isolation (PVI)Compared to the PENTARAY catheter: Score 58 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Design and Coverage for Confirming Pulmonary Vein Isolation (PVI)Compared to the PENTARAY catheter: Score 612 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Design and Coverage for Confirming Pulmonary Vein Isolation (PVI)Compared to the PENTARAY catheter: Score 78 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Design and Coverage for Confirming Pulmonary Vein Isolation (PVI)Compared to other competitive multielectrode catheters: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Design and Coverage for Confirming Pulmonary Vein Isolation (PVI)Compared to other competitive multielectrode catheters: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Design and Coverage for Confirming Pulmonary Vein Isolation (PVI)Compared to other competitive multielectrode catheters: Score 31 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Design and Coverage for Confirming Pulmonary Vein Isolation (PVI)Compared to other competitive multielectrode catheters: Score 41 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Design and Coverage for Confirming Pulmonary Vein Isolation (PVI)Compared to other competitive multielectrode catheters: Score 510 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Design and Coverage for Confirming Pulmonary Vein Isolation (PVI)Compared to other competitive multielectrode catheters: Score 66 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Design and Coverage for Confirming Pulmonary Vein Isolation (PVI)Compared to other competitive multielectrode catheters: Score 73 Participants
Secondary

Number of Responders for Physician Assessment for Maneuverability and Handling

Number of responders for physician assessment for maneuverability & handling were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for maneuverability and handling.

Time frame: Up to 29 weeks

Population: The safety analysis set consisted of all enrolled participants who underwent insertion of the study catheter. Here, 'n' (number analyzed) is defined as participants analyzed for specified categories.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Left Atrium): Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Left Atrium): Score 21 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Left Atrium): Score 31 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Left Atrium): Score 46 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Left Atrium): Score 57 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Left Atrium): Score 614 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Left Atrium): Score 73 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Right Atrium): Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Right Atrium): Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Right Atrium): Score 31 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Right Atrium): Score 44 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Right Atrium): Score 54 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Right Atrium): Score 69 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Right Atrium): Score 72 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Left Ventricle): Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Left Ventricle): Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Left Ventricle): Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Left Ventricle): Score 40 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Left Ventricle): Score 50 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Left Ventricle): Score 61 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Left Ventricle): Score 71 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Right Ventricle): Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Right Ventricle): Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Right Ventricle): Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Right Ventricle): Score 42 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Right Ventricle): Score 52 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Right Ventricle): Score 64 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Deploy (Right Ventricle): Score 70 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Left Atrium): Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Left Atrium): Score 21 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Left Atrium): Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Left Atrium): Score 44 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Left Atrium): Score 58 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Left Atrium): Score 610 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Left Atrium): Score 79 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Right Atrium): Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Right Atrium): Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Right Atrium): Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Right Atrium): Score 41 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Right Atrium): Score 57 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Right Atrium): Score 67 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Right Atrium): Score 72 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Left Ventricle): Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Left Ventricle): Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Left Ventricle): Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Left Ventricle): Score 41 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Left Ventricle): Score 50 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Left Ventricle): Score 60 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Left Ventricle): Score 71 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Right Ventricle): Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Right Ventricle): Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Right Ventricle): Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Right Ventricle): Score 42 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Right Ventricle): Score 52 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Right Ventricle): Score 63 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Maneuver (Right Ventricle): Score 71 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Left Atrium): Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Left Atrium): Score 21 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Left Atrium): Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Left Atrium): Score 44 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Left Atrium): Score 57 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Left Atrium): Score 611 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Left Atrium): Score 78 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Right Atrium): Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Right Atrium): Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Right Atrium): Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Right Atrium): Score 42 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Right Atrium): Score 55 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Right Atrium): Score 68 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Right Atrium): Score 72 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Left Ventricle): Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Left Ventricle): Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Left Ventricle): Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Left Ventricle): Score 41 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Left Ventricle): Score 50 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Left Ventricle): Score 60 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Left Ventricle): Score 71 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Right Ventricle): Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Right Ventricle): Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Right Ventricle): Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Right Ventricle): Score 42 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Right Ventricle): Score 53 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Right Ventricle): Score 63 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to Reach Areas (Right Ventricle): Score 70 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Left Atrium): Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Left Atrium): Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Left Atrium): Score 31 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Left Atrium): Score 46 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Left Atrium): Score 55 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Left Atrium): Score 610 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Left Atrium): Score 710 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Right Atrium): Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Right Atrium): Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Right Atrium): Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Right Atrium): Score 44 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Right Atrium): Score 53 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Right Atrium): Score 69 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Right Atrium): Score 73 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Left Ventricles): Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Left Ventricles): Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Left Ventricles): Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Left Ventricles): Score 40 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Left Ventricles): Score 50 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Left Ventricles): Score 61 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Left Ventricles): Score 71 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Right Ventricles): Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Right Ventricles): Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Right Ventricles): Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Right Ventricles): Score 41 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Right Ventricles): Score 52 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Right Ventricles): Score 64 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingAbility to withdraw (Right Ventricles): Score 71 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingOverall ease of use of the catheter: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingOverall ease of use of the catheter: Score 21 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingOverall ease of use of the catheter: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingOverall ease of use of the catheter: Score 46 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingOverall ease of use of the catheter: Score 511 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingOverall ease of use of the catheter: Score 620 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingOverall ease of use of the catheter: Score 72 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingInsertion and withdrawal from introducer sheath: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingInsertion and withdrawal from introducer sheath: Score 21 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingInsertion and withdrawal from introducer sheath: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingInsertion and withdrawal from introducer sheath: Score 44 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingInsertion and withdrawal from introducer sheath: Score 510 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingInsertion and withdrawal from introducer sheath: Score 613 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingInsertion and withdrawal from introducer sheath: Score 712 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingLearning curve: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingLearning curve: Score 21 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingLearning curve: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingLearning curve: Score 46 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingLearning curve: Score 512 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingLearning curve: Score 617 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Maneuverability and HandlingLearning curve: Score 74 Participants
Secondary

Number of Responders for Physician Assessment for Pacing

Number of responders for physician assessment for pacing (high output stimulation pacing and local pacing capture) were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for pacing.

Time frame: Up to 29 weeks

Population: The safety analysis set consisted of all enrolled participants who underwent insertion of the study catheter. Here, 'N' (number of participants analyzed) signifies participants evaluated for this outcome measure and 'n' (number analyzed) is defined as participants analyzed for specified categories.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for PacingHigh output stimulation pacing: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for PacingHigh output stimulation pacing: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for PacingHigh output stimulation pacing: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for PacingHigh output stimulation pacing: Score 41 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for PacingHigh output stimulation pacing: Score 51 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for PacingHigh output stimulation pacing: Score 61 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for PacingHigh output stimulation pacing: Score 72 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for PacingLocal pacing capture: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for PacingLocal pacing capture: Score 21 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for PacingLocal pacing capture: Score 32 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for PacingLocal pacing capture: Score 44 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for PacingLocal pacing capture: Score 57 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for PacingLocal pacing capture: Score 64 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for PacingLocal pacing capture: Score 75 Participants
Secondary

Number of Responders for Physician Assessment for Signal Collection and Quality

Number of responders for physician assessment for signal collection and quality (UNIPOLAR signals in atria/ventricles, UNIPOLAR noise encountered, BIPOLAR signals in atria/ventricles, BASELINE noise encountered, and BIPOLAR noise encountered) were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for signal collection and quality.

Time frame: Up to 29 weeks

Population: The safety analysis set consisted of all enrolled participants who underwent insertion of the study catheter. Here, 'N' (number of participants analyzed) signifies participants evaluated for this outcome measure and 'n' (number analyzed) is defined as participants analyzed for specified categories.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR signals in Atria: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR signals in Atria: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR signals in Atria: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR signals in Atria: Score 43 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR signals in Atria: Score 55 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR signals in Atria: Score 65 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR signals in Atria: Score 76 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR signals in Ventricles: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR signals in Ventricles: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR signals in Ventricles: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR signals in Ventricles: Score 42 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR signals in Ventricles: Score 51 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR signals in Ventricles: Score 62 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR signals in Ventricles: Score 74 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR noise encountered: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR noise encountered: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR noise encountered: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR noise encountered: Score 45 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR noise encountered: Score 56 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR noise encountered: Score 611 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityUNIPOLAR noise encountered: Score 76 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR signals in Atria: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR signals in Atria: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR signals in Atria: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR signals in Atria: Score 44 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR signals in Atria: Score 56 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR signals in Atria: Score 67 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR signals in Atria: Score 715 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR signals in Ventricles: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR signals in Ventricles: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR signals in Ventricles: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR signals in Ventricles: Score 41 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR signals in Ventricles: Score 50 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR signals in Ventricles: Score 62 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR signals in Ventricles: Score 77 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBASELINE noise encountered: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBASELINE noise encountered: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBASELINE noise encountered: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBASELINE noise encountered: Score 45 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBASELINE noise encountered: Score 57 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBASELINE noise encountered: Score 612 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBASELINE noise encountered: Score 714 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR noise encountered: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR noise encountered: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR noise encountered: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR noise encountered: Score 45 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR noise encountered: Score 59 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR noise encountered: Score 69 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for Signal Collection and QualityBIPOLAR noise encountered: Score 715 Participants
Secondary

Number of Responders for Physician Assessment for Workflow

Number of responders for physician assessment for workflow were reported. A post-procedure survey of 11 questions, each with individual sub-questions were asked. Each question/sub-question was answered by the physician using a Likert scale of 1 to 7 (1=poor and 7=excellent). Responders refers to participants with physician assessment for workflow.

Time frame: Up to 29 weeks

Population: The safety analysis set consisted of all enrolled participants who underwent insertion of the study catheter. Here, 'n' (number analyzed) is defined as participants analyzed for specified categories.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowAbility to reduce fluoroscopy: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowAbility to reduce fluoroscopy: Score 21 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowAbility to reduce fluoroscopy: Score 34 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowAbility to reduce fluoroscopy: Score 45 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowAbility to reduce fluoroscopy: Score 514 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowAbility to reduce fluoroscopy: Score 64 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowAbility to reduce fluoroscopy: Score 710 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowTime for pre-procedure mapping: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowTime for pre-procedure mapping: Score 21 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowTime for pre-procedure mapping: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowTime for pre-procedure mapping: Score 44 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowTime for pre-procedure mapping: Score 514 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowTime for pre-procedure mapping: Score 619 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowTime for pre-procedure mapping: Score 72 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowTime for post-procedure mapping after standard of care ablation: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowTime for post-procedure mapping after standard of care ablation: Score 21 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowTime for post-procedure mapping after standard of care ablation: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowTime for post-procedure mapping after standard of care ablation: Score 44 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowTime for post-procedure mapping after standard of care ablation: Score 58 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowTime for post-procedure mapping after standard of care ablation: Score 611 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowTime for post-procedure mapping after standard of care ablation: Score 712 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowAbility to identify targets: Score 10 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowAbility to identify targets: Score 20 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowAbility to identify targets: Score 69 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowAbility to identify targets: Score 30 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowAbility to identify targets: Score 44 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowAbility to identify targets: Score 511 Participants
Multi-electrode Electrocardiogram (ECG) Mapping Catheter (Investigational Catheter)Number of Responders for Physician Assessment for WorkflowAbility to identify targets: Score 75 Participants