Skip to content

Comparison Of Effective Radiation Shielding Between Rampart IC, M1128 Shield vs Conventional Apron And Shields In Clinical Interventional Procedures

Comparison Of Effective Radiation Shielding Between A Portable, Configurable, Floor-Supported Radiation Protection System vs Conventional Apron And Shields In Clinical Interventional Procedures

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05347290
Acronym
RAMPART
Enrollment
63
Registered
2022-04-26
Start date
2022-03-09
Completion date
2022-12-09
Last updated
2023-02-03

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Radiation Exposure

Keywords

radiation, radiation exposure, interventional cardiology, cardiac catherization laboratory, cardiac cath lab, lead apron, RAMPART Shield, radiation protection

Brief summary

Previous studies have documented that orthopedic injuries and musculoskeletal pain are a likely result of wearing heavy leaded aprons. This single-center, randomized 1-3 month clinical study compares conventional lead aprons and ancillary shields to the Rampart IC, M1128 radiation protection system in order to improve radiation safety with minimal orthopedic strain by using the RAMPART device. There will be three study arms, each arm requiring 21 procedures for a total of 63 total procedures, Each procedure will be randomized, instead of individual subjects. Each enrolled subject may be involved in more than one procedure/case. Randomization stratification will be 1:1:1 to either Rampart shield only, lead apron and ancillary shield, or lead apron and Rampart Shield. Real-time dosimeters will be used in each procedure to measure radiation attenuation.

Detailed description

Previous studies have documented that orthopedic injuries and musculoskeletal pain are a likely result of wearing heavy leaded aprons that are designed to shield interventionalists and laboratory staff from radiation. A large, multicenter study found that interventional laboratory employees reported a 67% increase in the prevalence of musculoskeletal pain. Two Society of Cardiovascular Angiography and Intervention (SCAI) surveys were performed a decade apart in 2004 and 2014 which both showed similar results on the orthopedic complications on the operators who wear leaded apron protection. The surveys revealed a 60% incidence of orthopedic issues and operators who have been operating greater than 20 years in the Cath Lab compared to a 2.3% incidence of chronic spine problems in the general population. Interestingly, of the operators who had been in practice less than 5 years, 26% already had orthopedic complaints. The study also revealed that over a third of those surveyed had already had to take spine related periods of absence from the catheterization lab. Though in recent years focus on individual radiation attenuation technologies that support both attenuation and orthopedic safety have evolved, the emphasis has been on the operator. A solution has not existed to address radiation attenuation and orthopedic safety for the entire scrubbed interventional team with one system. To address the challenge of radiation protection in this environment without the burden of wearing a heavy lead apron, Rampart IC, M1128 was created. Rampart IC, M1128 is a fully adjustable and portable system that provides radiation protection for operators and their team without having to wear a lead apron. This study will have three randomization arms, each arm requiring 21 procedures for a total of 63 total procedures. Randomization stratification will be based on procedure, not individual subject. Each enrolled subject may be involved in more than one procedure/case. Computer software will be utilized to randomize procedures in a 1:1:1 ratio to either lead apron with ancillary shield, Rampart IC, M1128, or lead apron with the Rampart IC, M1128. Real-time dosimeters will be placed on interventional staff's left temple, shoulder, and hip in order to measure radiation attenuation.

Interventions

DEVICERampart IC, M1128

Rampart IC, M1128 aims to address the challenge of radiation protection in the cardiac catheterization laboratory without the burden of wearing a heavy lead apron. Rampart M1128 is a compactly designed, portable, fully adjustable system that provides radiation protection for interventionalists and their technicians. It has independently adjustable flexible panels, allowing interventionalists multiple vascular access points, including right radial, bi-lateral femoral, and bi-lateral pedal. Participants in this arm will conduct catherization procedures using only the Rampart System, and radiation exposure will be measured using real-time dosimeters.

DEVICELead Apron and Ancillary Shield

Participating providers in this arm will use conventional skirt and vest, thyroid collar, and 100% use of 2 types of ancillary shields. Most wrap-around skirts and vests were (0.25-0.5 mm) Pb. Thyroid shields were 0.5 mm Pb equivalent. Under the table lead x 2. Radiation exposure will be measured using real-time dosimeters.

DEVICERampart IC, M1128 and Lead Apron

Participating providers will use both the lead apron and the Rampart Shield. Radiation exposure will be measured using real-time dosimeters.

Sponsors

Northwell Health
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
OTHER
Masking
NONE

Intervention model description

This single center randomized 1-3 month clinical study compares conventional lead aprons and ancillary shields to the Rampart IC, M1128 fully configurable, portable, floor supported system that provides radiation protection and orthopedic safety for operators and the scrubbed interventional team members. There will be three study arms, each arm requiring 21 procedures for a total of 63 total procedures. Each enrolled subject may be involved in more than one procedure/case. For all arms, subjects will wear real-time dosimeters on their right temple, shoulder, and hip, and values will be collected post-procedure.

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* Participants \> 18 years of age. * Interventional cardiology staff at Lenox Hill Hospital that are scheduled to work in Room C of the cardiac catheterization laboratory.

Exclusion criteria

* Pregnancy * Unwilling to consent to the study protocol.

Design outcomes

Primary

MeasureTime frameDescription
Change in radiation attenuation90 daysChange in radiation attenuation as detected by dosimeter readings for staff assigned to each randomization stratification.

Countries

United States

Outcome results

None listed